DOI QR코드

DOI QR Code

Comparison of Radiation Doses between 64-slice Single Source and 128-slice Dual Source CT Coronary Angiography in patient

64-slice single source CT와 128-slice dual source CT를 이용한 관상동맥 조영 검사 시 환자선량 비교

  • Kang, Yeong-Han (Department of Diagnostic Radiology, Daegu Catholic University Hospital)
  • 강영한 (대구가톨릭대학교병원 영상의학과)
  • Received : 2012.02.12
  • Accepted : 2012.04.24
  • Published : 2012.04.30

Abstract

The purpose of this study was to estimate radiation doses from 64-slice single source Computed Tomography(SSCT) coronary angiography(CA) and 128-slice dual source Computed Tomography(DSCT). With SSCT CA, the effective dose averaged approximately 13.86 mSv when two dose modulation was not. The mean effective dose for DSCT CA with retrospectively gated helical(RGH) technique was 11.87 mSv, when prospective ECG gating transverse(PGT) without dose modulation technique was 5.61 mSv. The one with dose modulation in PGT technique and flash mode were 3.04 mSv and flash mode was 0.98 mSv respectively. The lifetime attributable risk(LAR) of cancer incidence from SSCT RGH mode averaged approximately 1 for 1,176, and DSCT averaged 1 for 1,960(RGH mode), 1 for 3,030(PGT without modulation), 1 for 5,882(PGT with modulation). Because of CTCA is associated with non-negligible risk of cancer. Doses can be reduced by application PGT, FLASH than RGH using DSCT.

본 연구는 64-절편 단선원 CT(SSCT)와 128-절편 이중선원(DSCT)을 이용한 관상동맥조영 CT 검사 시 방사선량에 대해 알아보고, 이 선량으로 인한 암 발생의 잠재적 위험(LAR)을 평가해 보았다. SSCT의 후향적동조화(RGH)스캔의 유효선량은 13.86 mSv 이었고, DSCT의 RGH 스캔의 유효선량은 11.87 mSv 이었다. DSCT의 전향적동조화 (PGT) 스캔 중 선량변조를 적용치 않은 모드의 유효선량은 5.61 mSv, 선량변조를 적용하였을 때는 3.04 mSv 이었다. 1회 스캔하는 FLASH 모드는 0.98 mSv 이었다. SSCT의 RGH 스캔의 LAR은 1,176명 중 1명, DSCT의 RGH에서는 1,960명 중 1명이었다. DSCT의 PGT는 3,030명 중 1명, 선량변조를 하면 5,882명 중 1명에서 발암 위험이 있었다. 따라서 SSCT와 DSCT를 이용한 관상동맥조영 CT 검사는 암발생 귀속위험과 관련성이 있음을 인지하여야 하고, 환자선량을 줄이기 위해 RGH보다 PGT, FLASH 등의 프로토콜을 적용하는 것이 필요하다.

Keywords

References

  1. Raff GL, Gallagher MJ, O'Neill WW and Goldstein JA, "Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography", J Am Coll Cardiol, Vol. 46, No. 3, pp.552-557, 2005. https://doi.org/10.1016/j.jacc.2005.05.056
  2. Mollet NR, Cademartiri F, Carlos AG, and et al. "High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional angiography", Circulation, Vol. 112, pp.2318-2323, 2005. https://doi.org/10.1161/CIRCULATIONAHA.105.533471
  3. Brenner DJ, Hall EJ, "Computed tomography-an increasing source of radiation exposure", N Engl J Med, Vol. 357, pp.2277-2284, 2007. https://doi.org/10.1056/NEJMra072149
  4. Achenbach S, Ropers D, Kuettner A, and et al. "Contrast-enhanced coronary artery visualization by dual-source computed tomography-initial experience", Eur J Radiol, Vol. 57, pp.331-335, 2006. https://doi.org/10.1016/j.ejrad.2005.12.017
  5. Johnson TR, Konstantin N, Wintersperger BJ, et al. "Dual-source CT cardiac imaging: initial experience", Eur Radiol, Vol. 16, pp1409-1415, 2006. https://doi.org/10.1007/s00330-006-0298-y
  6. H Scheffel, H Alkadhi, A Plass, and et al. "Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control", Eur Radiol Vol. 16, pp.2739-2747, 2006. https://doi.org/10.1007/s00330-006-0474-0
  7. Francone M, Napoli A, Carbone I, and et al. "Noninvasive imaging of the coronary arteries using a 64-row multidetector CT scanner: initial clinical experience and radiation dose concerns", Radiol Med, Vol. 112, No. 1, pp31-46, 2007. https://doi.org/10.1007/s11547-007-0118-8
  8. McCollough CH, Primak AN, Saba O, and et al. "Dose performance of a 64-channel dual-source CT scanner", Radiology Vol. 243, No .3, pp.775-784, 2007. https://doi.org/10.1148/radiol.2433061165
  9. S Baumuller, S Leschka, L Desbiolles, and et al. "Dual-source versus 64-section CT coronary angiography at lower heart rates: Comparison of accuracy and radiation dose", RSNA, Radiology, Vol. 253, No. 1, pp.56-64, 2009. https://doi.org/10.1148/radiol.2531090065
  10. Hausleiter J, Meyer T, Hadamitzky M, and et al. "Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates", Circulation, Vol. 113, pp.1305-1310, 2006. https://doi.org/10.1161/CIRCULATIONAHA.105.602490
  11. Stolzmann P, Scheffel H, Schertler T, and et al. "Radiation dose estimates in dual-source computed tomography coronary angiography", Eur Radiol, Vol. 18, No. 3, pp.592-599, 2008. https://doi.org/10.1007/s00330-007-0786-8
  12. JP Earls, EL Berman, BA Urban, and et al. "Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: Improved image quality and reduce radiation dose", Radiology, Vol. 246, No. 3, pp.1081-1086, 2008.
  13. US Food and Drug Administration, Center for Devices and Radiological Health. "Whole body scanning using computed tomography(CT): what are the radiation risks from CT?", http://www.fda.gov/ cdrh/ct/risks.html.
  14. Gham Hur, Seong Wan Hong, Su Young Kim, and et al. "Uniform image achieved by tube current modulation using SD of attenuation in coronary CT angiography", AJR, Vol. 189, pp.188-196, 2007. https://doi.org/10.2214/AJR.06.1201
  15. ICRP, "Managing Patient Dose in Multi-detector Computed Tomography", ICRP publication 102, 2007.
  16. Bongartz, SJ Golding, AG Jurik, M Leonardi, EvP van Meerten, J Geleijns, and et al. "European guidelines on quality criteria for computed tomography", EUR 16262, 2000. http://www.drs.dk/guidelines /ct/quality
  17. Committee to Assess Health Risks from Exposure to low levels of Ionizing Radiation, National Research Council of the National Academies. Health risks from Exposure to Low Levels of Ionizing Radiation, BEIR VII Phase 2. Washington, DC: The National Academies Press, 2006
  18. 이대규, 김문찬, 방성호, 한동균. Dual source CT에서 High-pitch Cardiac scan mode를 이용한 관상동맥 검사의 유용성. 대한전산화단층기술학회지. Vol. 12, No. 1, p25-33, 2010.
  19. JF Paul, and HT Abada, "Strategies for reduction of radiation dose in cardiac multislice CT", EUR Radiol, Vol. 17, pp.2028-2037, 2007. https://doi.org/10.1007/s00330-007-0584-3
  20. Poll LW, Cohnen M, Brachten S, Ewen K, and Modder U, "Dose reduction in multi-slice CT of the heart by use of ECG-controlled tube current modulation ("ECG pulsing"): phantom measurements", Rofo, Vol. 174, No. 12, pp.1500-1505, 2002. https://doi.org/10.1055/s-2002-35945
  21. Leschka S, Wildermuth S, Boehm T, and et al. "Noninvasive coronary angiography with 64-section CT: effect of average heart rate and heart rate variability on image quality", Radiology, Vol. 241, No. 2, pp.378-385, 2006. https://doi.org/10.1148/radiol.2412051384
  22. Leschka S, Husmann L, Desbiolles LM, and et al. "Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT", Eur Radiol, Vol. 16, No. 9, pp.1964-1972, 2006. https://doi.org/10.1007/s00330-006-0262-x
  23. S Achenbach, T Goroll, M Seltmann, and et al. "Detection of coronary artery stenosis by low-dose, prospectively ECG-triggered, high-pitch spiral coronary CT angiography", JACC, Cardiovascular imaging, Vol. 4, No. 4, pp.211, 2011.
  24. Jung B, Mahnken AH, Stargardt A, and et al. "Individually weight-adapted examination protocol in retrospectively ECG-gated MSCT of the heart", Eur Radiol Vol. 13, pp.2560-2566, 2003 https://doi.org/10.1007/s00330-003-2111-5
  25. Einstein AJ, Henzlova MJ, and Rajagopalan S, "Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography", JAMA, Vol. 298, No. 3, pp.317- 323, 2007. https://doi.org/10.1001/jama.298.3.317
  26. AC Weustink, NR mollet, LA Neefjes, and et al. "Preserved Diagnostic Performance of Dual-Source CT coronary Angiography with Reduced Radiation Exposure and Cancer Risk", Radiology, Vol. 252, No. 1, pp.53-60, 2009. https://doi.org/10.1148/radiol.2521082023

Cited by

  1. The study of MDCT of Radiation dose in the department of Radiology of general hospitals in the local area vol.6, pp.4, 2012, https://doi.org/10.7742/jksr.2012.6.4.281
  2. CT 검사별 노출되는 유효선량과 생애 암 귀속 위험도 평가 vol.14, pp.2, 2012, https://doi.org/10.7742/jksr.2020.14.2.169