• Title/Summary/Keyword: 확장 불확도

Search Result 77, Processing Time 0.034 seconds

Calibration uncertainty of a spinning rotor gauge (스피닝 로터 게이지 교정 불확도)

  • 홍승수;임인태;신용현;정광화
    • Journal of the Korean Vacuum Society
    • /
    • v.12 no.1
    • /
    • pp.7-15
    • /
    • 2003
  • Spinning Rotor Gauge (SRG) has been used to transfer standard gauge for international comparison at the high vacuum standards. We calibrated a spinning rotor gauge by using dynamic calibration system (DCS) that was a national high vacuum standards system. And its uncertainties were evaluated with the International Organization for Standardization (ISO), they were recognized ai A type uncertainty, B type uncertainty, combined Standard uncertainty, and expanded uncertainty. The combined standard uncertainties were $1.8007\times10^{-5}$ Pa ~ ~$4.8422\times10^{-5}$ Pa for this spinning rotor gauge.

Uncertainty of Calibration using TVC for RF Multimeter (열전압변환기를 이용한 고주파전압계의 교정과 측정불확도)

  • 박정규;박석주;이환상;장경승;박명선
    • Proceedings of the Korea Electromagnetic Engineering Society Conference
    • /
    • 2001.11a
    • /
    • pp.310-314
    • /
    • 2001
  • 측정의 소급성과 신뢰성을 확보한다는 전제하에 오차의 개념을 버리고 측정시의 측정불확도를 분석하는 것이 측정과학에 있어서 세계적인 추세이다. 본 논문에서는 전달표준인 열전압변환기를 사용하여 고주파전압계의 정밀 교정에서 불확도를 산출하기 위하여 수학적 모델을 설정하고 측정에 관계되는 모든 불확도 요인을 분석하였다. 케이블의 영향과 열전대감지기와 직류전압계의 상관관계, 직류소스, RF소스, 감쇠기, 반사계수와 DUT 등 측정에 있어서 관계되는 모든 불확도요인을 분석하여 확장불확도를 산출하였다.

  • PDF

Measurement Uncertainty of Arsenic Concentration in Ambient PM2.5 Determined by Instrumental Neutron Activation Analysis (기기 중성자방사화분석을 이용한 대기 중 PM2.5 내 Arsenic 농도 분석의 측정 불확도)

  • Lim, Jong-Myoung;Lee, Jin-Hong;Moon, Jong-Wha;Chung, Yong-Sam
    • Journal of Korean Society of Environmental Engineers
    • /
    • v.30 no.11
    • /
    • pp.1123-1131
    • /
    • 2008
  • In this study, measurement uncertainty of instrumental neutron activation analysis was evaluated for ambient As concentration in PM2.5. Expanded uncertainties of the measurements were calculated by applying both ISO-GUM approximation and Monte Carlo Simulation(MCS). The estimate of As concentration on a specific day by the Monte Carlo Simulation differed from that of ISO-GUM approximation by less than 4%. Relative expanded uncertainties of As concentrations from a total number of 60 PM2.5 samples were also estimated to be more or less than 10% with 95% confidence level using the Monte Carlo Simulation. Sensitivity test of the measurement uncertainties showed that $\gamma$-ray counting error(62.3%), efficiency(18.5%), air volume(12.3%), neutron flux(2.3%), and absolute gamma-intensity(1.8%) are major factors of uncertainty variations.

CDN을 이용한 전자파 전도내성시험의 측정불확도

  • Mun Jun-Cheol;Im Yeong-Chae
    • The Proceeding of the Korean Institute of Electromagnetic Engineering and Science
    • /
    • v.16 no.4 s.56
    • /
    • pp.84-92
    • /
    • 2005
  • 본 논문은 CDN(Coupling/Decoupling Network)을 이용한 전자파 전도 내성 시험의 측정 불확도를 연구하였다. 측정 불확도의 영향 인자를 추출하여 인자별 불확도를 산출하였다. 측정 불확도 인자는 인가 레벨 교정시 반복 측정, $150\sim50\Omega$ 아답터의 삽입 손실, 신호 발생기의 정확도, 파워미터의 정확도, 전력 증폭기의 하모닉, 전력 증폭기와 CDN의 부정합, CDN과 파워미터 사이의 부정합에 의한 불확도를 고려하였다. 불확도 산출 결과 확장 불확도는 95% 신뢰 구간에서 1.17 dB(k=)를 가지며, 정보기기의 경우 시험 전압 3 V에서 ${\pm}0.42V$의 오차 범위를 가지는 것을 알 수 있었다. 또한 시험기기간의 부정합에 의한 불확도가 불확도 인자에서 가장 큰 영향을 미치는 것을 알 수 있었으며, 전자파 전도 내성 시험의 경우 임피던스 매칭이 가장 중요한 것으로 나타났다.

Uncertainty Assesment for Moving-boat ADCP Discharge Measurements by GUM Framework (GUM 표준안 기반 이동식 ADCP 유량 측정 불확도 평가)

  • Kim, Dongsu;Kim, Jong Min
    • Proceedings of the Korea Water Resources Association Conference
    • /
    • 2017.05a
    • /
    • pp.71-71
    • /
    • 2017
  • 하천에서 평수기 유량측정은 도섭법을 이용하기 위한 지점식 측정보다는 초음파 도플러 유속계(ADCP, Acoustic Doppler Current Profiler)를 보트에 탑재하여 운용하는 측정 방식이 점차 일반화되고 있다. ADCP는 초음파의 도플러효과를 이용하여 수심이나 횡방향의 유속 분포를 측정할 수 있는 측정 장비로 일반적으로 사용되는 down-looking ADCP는 수심방향의 유속분포와 수심을 측정하여 보트의 이동속도와의 벡터 내적을 이용하여 유량을 산정하게 된다. 그러나, 이동식 ADCP 유량 측정 성과의 불확도는 제공되지 않고 있는 상황인데, 이는 불확도 산정 표준안 미비, 유속 및 수심 등 측정 요소의 관측 환경 별 불확도 정보 부족, 불확도를 산정할 수 있는 툴의 부재 등에 기인한다. 본 연구에서는 이동식 ADCP 불확도 산정 표준안을 개발하고 현장 실험을 통해 불확도 요인에 대한 규명, 불확도를 편리하게 산정할 수 있는 툴을 개발하고자 하였다. 불확도 산정 표준안으로 최근 WMO를 위시한 국제적으로 하천 유량 측정 불확도 표준안으로 채택되고 있는 GUM(Guide to the Expression of Uncertainty Measurement)을 기반으로 이동식 ADCP 유량 산정 알고리즘을 적용하여 불확도 적용 기법을 개발하였다. GUM 표준안을 기반으로 유량 측정불확도를 산정하기 위한 불확도 요인분석은 실규모 하천의 특성을 대부분 모의할 수 있는 한국건설기술연구원의 안동하천실험센터에서 수행된 실험자료를 기반으로 다양한 인자들에 대한 요소 별 불확도 분석을 수행하였다. GUM 표준안에 의하면 불확도 요인들은 오차전파의 법칙에 기반하여 전체 불확도에 전파되며, 이렇게 합성된 불확도는 t-분포의 신뢰수준 95%일 경우의 보정계수 2를 곱하여 최종적으로 확장불확도를 산정하게 된다. 이동측정방식의 ADCP의 경우 GUM 표준안에 적용하여 불확도를 평가하기 위해서 사용되는 수식이 방대하고, 매우 복잡하기 때문에 이를 실무자가 평가하기에는 한계가 있다. 이에 따라 본 연구에서는 ADCP의 유량 측정불확도를 보다 편리하게 평가하기 위하여 ADCP 유량 측정불확도 평가 소프트웨어인 AQUA(ADCP Discharge Uncertainty Assesment)를 개발하였으며, 이를 통해 실무자나 연구자들이 ADCP의 불확도 평가에 보다 편리하게 접근할 수 있을 것이라 판단된다.

  • PDF

A Study on Comparison between the Propagation of Uncertainty by GUM and Monte-Carlo Simulation (측정 불확도 표현 지침서(GUM)와 Monte-Carlo Simulation에 의한 불확도 전파 결과의 비교 연구)

  • Jungkee Shu;Hyungsik Min;Minsu Park;Jin-Chun Woo;Jongsang Kim
    • Journal of the Korean Chemical Society
    • /
    • v.47 no.1
    • /
    • pp.31-37
    • /
    • 2003
  • The expanded uncertainties calculated by the application of GUM -approximation and Monte-Carlo simulation were compared about the model equation of one-point calibration which is widely used for the measurements and chemical analysis. For the comparisons, we assumed a set of artificial data at the various level of concentration and dispersion of t or normal distribution. Mistakes of more then 50 % was revealed at the values calculated by GUM-approximation in comparison with those of Monte-Carlo simulation because of the excess dispersion from t-distribution and non-linearity by division in the equation. In contrary, the mistake of calculation due to non-linearity of the equation was not observed in the level of detection limits with the equation of one-point calibration, because of the relatively large values of uncertainty in response.

A study on quality assurance and evaluation of uncertainty for the analysis of natural gas (천연가스 분석의 불확도 평가 및 품질 보증을 위한 연구)

  • Woo, Jin-Chun;Kim, Young-Doo;Bae, Hyun-Kil;Lee, Kang-Jin;Her, Jae-Young
    • Analytical Science and Technology
    • /
    • v.19 no.6
    • /
    • pp.490-497
    • /
    • 2006
  • The sources of uncertainty in the analysis of liquified natural gas (LNG) process are evaluated. The uncertainty sources evaluated are the repeatability of measurement, non-linearity of GC, the uncertainty of standard gas used for calibration, difference of gas sampling and deviation after GC calibration and major revealed sources are the non-linearity of GC, the uncertainty of standard gas and the deviation after GC calibration. The determined values and uncertainties of methane and ethane as the major components are $90.0%mol/mol{\pm}1.9%$ (relative and 95% level of confidence) and $6.26%mol/mol{\pm}0.08$ (relative and 95% level of confidence), respectively. The contribution of uncertainty varies depending on the source of uncertainty and gas component. In the case of methane, non-linearity of GC, the uncertainty of standard gas and deviation after GC calibration contribute 0.28%, 0.25% and 0.24% of relative expanded uncertainty, respectively.

Uncertainties in Pressure Calibration of Laboratory Standard Microphones by Reciprocity Technique (가역방법에 의한 표준 마이크로폰 음압교정의 불확도)

  • 서상준;권휴상;이용봉;서재갑
    • The Journal of the Acoustical Society of Korea
    • /
    • v.23 no.2
    • /
    • pp.90-102
    • /
    • 2004
  • According to the Mutual Recognition Arrangement (MRA), the calibration and/or test laboratories should satisfy the management and technical requirements ISO 17025 or equivalent. Chapter 5, Section 5.10.4 of the technical requirement of ISO 17025 suggests the required informations for calibration certificates, one of them is to state the uncertainty of measurement. The uncertainties of measurement in reciprocity calibration of standard laboratory microphone were calculated. The expanded uncertainties for 1 and 1/2 inch microphones were 0.03 dB in the middle frequency range and they increased up to 0.10 dB and 0.11 dB at 20 Hz, 0.07 dB and 0.08 dB at high frequency, respectively.

Calibration uncertainty of a capacitance diaphragm gauge (용량형진공계 교정 불확도)

  • 홍승수;신용현;정광화;임인태;우삼용;김정형;최상철
    • Journal of the Korean Vacuum Society
    • /
    • v.10 no.2
    • /
    • pp.173-181
    • /
    • 2001
  • We calibrated a capacitance diaphragm gauge(CDG) of 1,333 Pa range by using ultrasonic interferometer manometer (UIM) that was a national low vacuum standards system. And its uncertainties were evaluated according to International Organization for Standardization(ISO), they were named to A type uncertainty, B type uncertainty, combined standard uncertainty, and expanded uncertainty, We obtained that the combined standard uncertainties were $1.38 \times10^{-2}\; Pa\sim3.03 \times10^{-1} $Pa and the relative uncertainties(combined standard uncertainty/standard pressure) were $2.3 \times 10^{-4}\;Pa\sim7.9 X\times10^{-3} $Pa for this 1,333 Pa CDG.

  • PDF

Measurement System of Bidirectional Reflectance-distribution Function (양방향 반사율 분포함수 측정시스템)

  • Hwang, Ji-Soo
    • Korean Journal of Optics and Photonics
    • /
    • v.21 no.2
    • /
    • pp.46-52
    • /
    • 2010
  • A theory of bidirectional reflectance-distribution function (BRDF), a newly developed BRDF measurement system, and a method for evaluating the uncertainty of BRDF measurements are presented. The BRDF measurement system which measures BRDF in a wavelength range of (380~1500) nm with an angle range of $(-75{\sim}75)^{\circ}$ was installed. The measurement uncertainties, consisting of correlated terms and uncorrelated terms, were evaluated for the BRDF measurement system, resulting in the relative expanded uncertainty less than 3% (k=2).