Cardiac Magnetic Resonance Imaging Using Multi-physiological Intelligent Trigger System (멀티 생체신호 동기 시스템을 이용한 심장자기공명영상)

  • Park, Jinho;Yoon, Jong-Hyun;Yang, Young-Joong;Ahn, Chang-Beom
    • Investigative Magnetic Resonance Imaging
    • /
    • v.18 no.3
    • /
    • pp.244-252
    • /
    • 2014
  • Purpose : We proposed a multi-physiological signals based real-time intelligent triggering system(MITS) for Cardiac MRI. Induced noise of the system was analyzed. Materials and Methods: MITS makes cardiac MR imaging sequence synchronize to the cardiac motion using ECG, respiratory signal and second order derivative of $SPO_2$signal. Abnormal peaks due to arrhythmia or subject's motion are rejected using the average R-R intervals and R-peak values. Induced eddy currents by gradients switching in cardiac MR imaging are analyzed. The induced eddy currents were removed by hardware and software filters. Results: Cardiac MR images that synchronized to the cardiac and respiratory motion are acquired using MITS successfully without artifacts caused by induced eddy currents of gradient switching or subject's motion or arrhythmia. We showed that the second order derivative of the $SPO_2$ signal can be used as a complement to the ECG signals. Conclusion: The proposed system performs cardiac and respiratory gating with multi-physiological signals in real time. During the cardiac gating, induced noise caused by eddy currents is removed. False triggers due to subject's motion or arrhythmia are rejected. The cardiac MR imaging with free breathing is obtained using MITS.

Estimation of CyberKnife Respiratory Tracking System Using Moving Phantom (동적 팬톰을 이용한 사이버나이프 호흡동기 추적장치의 위치 정확성 평가)

  • Seo, Jae-Hyuk;Kang, Young-Nam;Jang, Ji-Sun;Shin, Hun-Joo;Jung, Ji-Young;Choi, Byong-Ock;Choi, Ihl-Bohng;Lee, Dong-Joon;Kwon, Soo-Il;Lim, Jong-Soo
    • Progress in Medical Physics
    • /
    • v.20 no.4
    • /
    • pp.324-330
    • /
    • 2009
  • In this study, we evaluated accuracy and usefulness of CyberKnife Respiratory Tracking System ($Synchrony^{TM}$, Accuray, USA) about a moving during stereotactic radiosurgery. For this study, we used moving phantom that can move the target. We also used Respiratory Tracking System called Synchrony of the Cyberknife in order to track the moving target. For treatment planning of the moving target, we obtained an image using 4D-CT. To measure dose distribution and point dose at the moving target, ion chamber (0.62 cc) and gafchromic EBT film were used. We compared dose distribution (80% isodose line of prescription dose) of static target to that of moving target in order to evaluate the accuracy of Respiratory Tracking System. We also measured the point dose at the target. The mean difference of synchronization for TLS (target localization system) and Synchrony were $11.5{\pm}3.09\;mm$ for desynchronization and $0.14{\pm}0.08\;mm$ for synchronization. The mean difference between static target plan and moving target plan using 4D CT images was $0.18{\pm}0.06\;mm$. And, the accuracy of Respiratory Tracking System was less 1 mm. Estimation of usefulness in Respiratory Tracking System was $17.39{\pm}0.14\;mm$ for inactivity and $1.37{\pm}0.11\;mm$ for activity. The mean difference of absolute dose was $0.68{\pm}0.38%$ in static target and $1.31{\pm}0.81%$ in moving target. As a conclusion, when we treat about the moving target, we consider that it is important to use 4D-CT and the Respiratory Tracking System. In this study, we confirmed the accuracy and usefulness of Respiratory Tracking System in the Cyberknife.

  • PDF

Evaluation of Respiration Reproducibility of Chest General X-ray Examination using Self-made Respiratory Synchronization Device (자체 제작한 호흡 동기화 장치를 통한 흉부 일반촬영 검사의 호흡 재현성 평가)

  • Kwon, Oh-Young;Lee, Chang-Hun;Yong, Keum-Ju;Jin, Seon-Hui;Jung, Da-Bin;Heo, Yeong-Cheol
    • Journal of the Korean Society of Radiology
    • /
    • v.15 no.7
    • /
    • pp.1049-1056
    • /
    • 2021
  • The purpose of this study was to develop a respiratory synchronization device for X-ray (X-RSD) to increase the reproducibility of inspiration when examining the Chest X-ray of a patient who difficulty in breathing coordination. The X-RSD was self-made using an air pressure sensor and air was injected by connecting a ventilator to the mannequin for CPR. At this time, the amount of injected air was quantified using the SkillReporting device. After placing the X-RSD on the chest of the mannequin, the amount of air was tested in 6 steps from 200 to 700 cc by 100 cc increased. For the accuracy evaluation, the sensitivity of X-RSD was measured by repeating a total of 80 measurements, and the sensitivity was 100%, and very precise results were obtained. After that, the images examined while viewing the X-RSD of the chest lateral examination and the images obtained by the blind examination were compared and evaluated. The lung volume of X-RSD was larger than that of the blind test, and the deviation was smaller. Overall, the use of X-RSD can help with chest X-ray examination of patients who have difficulty in cooperating, and it is thought that it will be possible to contribute to the reduction of exposure dose by reducing the repeat rate of general X-ray examinations.

A Development of Physio-Module for Echocardiography (심초음파용 생체신호측정모듈의 개발)

  • Jang, Won-Seuk;Kim, Nam-Hyun;Jeon, Dae-Keun
    • Journal of the Institute of Electronics Engineers of Korea SC
    • /
    • v.47 no.4
    • /
    • pp.21-29
    • /
    • 2010
  • In this study, we aimed to develope the physio-module for echocardiography. This physio-module includes multi-functions such as ECG, respiration, PCG, heart sound, and this is used to diagnose a cardiac disease in using ultrasound images synchronized with biosignals of physio-module. In this paper, the developed physio-module was verified by applying various test patterns considering each biosignal's characteristics and we could get the performance of QRS trigger delay time within international standard, EC-13 criteria. And ECG's change in physio-module and blood flow in M-mode was synchronized.

Emotion Recognition Method Using Heart-Respiration Connectivity (심장과 호흡의 연결성을 이용한 감성인식 방법)

  • Lee, Dong Won;Park, Sangin;Whang, Mincheol
    • Science of Emotion and Sensibility
    • /
    • v.20 no.3
    • /
    • pp.61-70
    • /
    • 2017
  • Physiological responses have been measured to recognize emotion. Although physiological responses have been interrelated between organs, their connectivities have been less considered for emotion recognizing. The connectivities have been assumed to enhance emotion recognition. Specially, autonomic nervous system is physiologically modulated by the interrelated functioning. Therefore, this study has been tried to analyze connectivities between heart and respiration and to find the significantly connected variables for emotion recognition. The eighteen subjects(10 male, age $24.72{\pm}2.47$) participated in the experiment. The participants were asked to listen to predetermined sound stimuli (arousal, relaxation, negative, positive) for evoking emotion. The bio-signals of heart and respiration were measured according to sound stimuli. HRV (heart rate variability) and BRV (breathing rate variability) spectrum were obtained from spectrum analysis of ECG (electrocardiogram) and RSP (respiration). The synchronization of HRV and BRV spectrum was analyzed according to each emotion. Statistical significance of relationship between them was tested by one-way ANOVA. There were significant relation of synchronization between HRV and BRV spectrum (synchronization of HF: F(3, 68) = 3.605, p = 0.018, ${\eta}^2_p=0.1372$, synchronization of LF: F(3, 68) = 5.075, p = 0.003, ${\eta}^2_p=0.1823$). HF difference of synchronization between ECG and RSP has been able to classify arousal from relaxation (p = 0.008, d = 1.4274) and LF's has negative from positive (p = 0.002, d = 1.7377). Therefore, it was confirmed that the heart and respiration to recognize the dimensional emotion by connectivity.

The Optimum of Respiratory Phase Using the Motion Range of the Diaphragm: Focus on Respiratory Gated Radiotherapy of Lung Cancer (횡격막의 움직임을 이용한 최적화된 호흡 위상의 선택: 폐암의 호흡 동기 방사선치료 중심)

  • Kim, Myoungju;Im, Inchul;Lee, Jaeseung;Kang, Suman
    • Journal of the Korean Society of Radiology
    • /
    • v.7 no.2
    • /
    • pp.157-163
    • /
    • 2013
  • This study was to analyze quantitatively movement of planning target volume (PTV) and change of PTV volume through movement of diaphragm according to breathing phase. The purpose of present study was to investigate optimized respiration phase for radiation therapy of lung cancer. Simulated breathing training was performed in order to minimize systematic errors which is caused non-specific or irregular breathing. We performed 4-dimensional computed tomography (4DCTi) in accordance with each respiratory phase in the normalized respiratory gated radiation therapy procedures, then not only defined PTVi in 0 ~ 90%, 30 ~ 70% and 40 ~ 60% in the reconstructed 4DCTi images but analyzed quantitatively movement and changes of volume in PTVi. As a results, average respiratory cycle was $3.4{\pm}0.5$ seconds by simulated breathing training. R2-value which is expressed as concordance between clinically induced expected value and actual measured value, was almost 1. There was a statistically significant. And also movement of PTVi according to each respiration phase 0 ~ 90%, 30 ~ 70% and 40 ~ 60% were $13.4{\pm}6.4mm$, $6.1{\pm}2.9mm$ and $4.0{\pm}2.1mm$ respectively. Change of volume in PTVi of respiration phase 30 ~ 70% was decreased by $32.6{\pm}8.7%$ and 40 ~ 60% was decreased by $41.6{\pm}6.2%$. In conclusion, PTVi movement and volume change was reduced, when we apply a short breathing phase (40 ~ 60%: 30% duty cycle) range. Furthermore, PTVi margin considered respiration was not only within 4mm but able to get uniformity of dose.

The Relationship and Mechanism Underlying the Effect of Conscious Breathing on the Autonomic Nervous System and Brain Waves (의식적 호흡이 자율신경과 뇌파에 영향을 미치는 기전에 관하여)

  • Kang, Seung Wan
    • Perspectives in Nursing Science
    • /
    • v.14 no.2
    • /
    • pp.64-69
    • /
    • 2017
  • Purpose: Breathing can be controlled either unconsciously or consciously. In Asian countries, various conscious breathing-control techniques have been practiced for many years to promote health and wellbeing. However, the exact mechanism underlying these techniques has not yet been established. The purpose of this study is to explore the physiological mechanism explaining how conscious breathing control could affect the autonomic nervous system, brain activity, and mental changes. Methods: The coupling phenomenon among breathing rhythm, heart rate variability, and brain waves was explored theoretically based on the research hypothesis and a review of the literature. Results: Respiratory sinus arrhythmia is a well-known phenomenon in which heart rate changes to become synchronized with breathing: inhalation increases heart rate and exhalation decreases it. HRV BFB training depends on conscious breathing control. During coherent sinusoidal heart rate changes, brain ${\alpha}$ waves could be enhanced. An increase in ${\alpha}$ waves was also found and the synchronicity between heart beat rhythm and brain wave became strengthened during meditation. Conclusion: In addition to the effect of emotion on breathing patterns, conscious breathing could change heart beat rhythms and brainwaves, and subsequently affect emotional status.

Integration of a Radiotherapy Robot System (방사선 로봇 시스템 통합)

  • Choi, You-Rak;Jeong, Gyoeng-Min;Choi, Young-Su
    • Proceedings of the Korea Information Processing Society Conference
    • /
    • 2017.04a
    • /
    • pp.1060-1061
    • /
    • 2017
  • 방사선 로봇 시스템은 X-Ray를 이용하는 로봇 기반 자동 치료 시스템으로 방사선 치료계획 시스템, 방사선가속기, 방사선 치료 로봇, 호홉 추적 시스템, 스마트베드로 구성된다. X-Ray를 이용하는 치료 시스템인 관계로 안정적인 제어가 요구되며, 환자의 호흡에 의한 병소 위치 변위 발생에도 X-Ray를 병소에 정확하게 정량 조사해야한다. 본 논문에서는 방사선 로봇 시스템을 구성하는 서브시스템 간의 데이터 송수신 동기화와 시스템 안정성 확보, 그리고 시스템 통합을 위한 문서 작업을 획기적으로 줄이면서 시스템 통합을 단시간에 수행한 과정에 대하여 기술한다.

An accuracy analysis of Cyberknife tumor tracking radiotherapy according to unpredictable change of respiration (예측 불가능한 호흡 변화에 따른 사이버나이프 종양 추적 방사선 치료의 정확도 분석)

  • Seo, jung min;Lee, chang yeol;Huh, hyun do;Kim, wan sun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.27 no.2
    • /
    • pp.157-166
    • /
    • 2015
  • Purpose : Cyber-Knife tumor tracking system, based on the correlation relationship between the position of a tumor which moves in response to the real time respiratory cycle signal and respiration was obtained by the LED marker attached to the outside of the patient, the location of the tumor to predict in advance, the movement of the tumor in synchronization with the therapeutic device to track real-time tumor, is a system for treating. The purpose of this study, in the cyber knife tumor tracking radiation therapy, trying to evaluate the accuracy of tumor tracking radiation therapy system due to the change in the form of unpredictable sudden breathing due to cough and sleep. Materials and Methods : Breathing Log files that were used in the study, based on the Respiratory gating radiotherapy and Cyber-knife tracking radiosurgery breathing Log files of patients who received herein, measured using the Log files in the form of a Sinusoidal pattern and Sudden change pattern. it has been reconstituted as possible. Enter the reconstructed respiratory Log file cyber knife dynamic chest Phantom, so that it is possible to implement a motion due to respiration, add manufacturing the driving apparatus of the existing dynamic chest Phantom, Phantom the form of respiration we have developed a program that can be applied to. Movement of the phantom inside the target (Ball cube target) was driven by the displacement of three sizes of according to the size of the respiratory vertical (Superior-Inferior) direction to the 5 mm, 10 mm, 20 mm. Insert crosses two EBT3 films in phantom inside the target in response to changes in the target movement, the End-to-End (E2E) test provided in Cyber-Knife manufacturer depending on the form of the breathing five times each. It was determined by carrying. Accuracy of tumor tracking system is indicated by the target error by analyzing the inserted film, additional E2E test is analyzed by measuring the correlation error while being advanced. Results : If the target error is a sine curve breathing form, the size of the target of the movement is in response to the 5 mm, 10 mm, 20 mm, respectively, of the average $1.14{\pm}0.13mm$, $1.05{\pm}0.20mm$, with $2.37{\pm}0.17mm$, suddenly for it is variations in breathing, respective average $1.87{\pm}0.19mm$, $2.15{\pm}0.21mm$, and analyzed with $2.44{\pm}0.26mm$. If the correlation error can be defined by the length of the displacement vector in the target track is a sinusoidal breathing mode, the size of the target of the movement in response to 5 mm, 10 mm, 20 mm, respective average $0.84{\pm}0.01mm$, $0.70{\pm}0.13mm$, with $1.63{\pm}0.10mm$, if it is a variant of sudden breathing respective average $0.97{\pm}0.06mm$, $1.44{\pm}0.11mm$, and analyzed with $1.98{\pm}0.10mm$. The larger the correlation error values in both the both the respiratory form, the target error value is large. If the motion size of the target of the sine curve breathing form is greater than or equal to 20 mm, was measured at 1.5 mm or more is a recommendation value of both cyber knife manufacturer of both error value. Conclusion : There is a tendency that the correlation error value between about target error value magnitude of the target motion is large is increased, the error value becomes large in variation of rapid respiration than breathing the form of a sine curve. The more the shape of the breathing large movements regular shape of sine curves target accuracy of the tumor tracking system can be judged to be reduced. Using the algorithm of Cyber-Knife tumor tracking system, when there is a change in the sudden unpredictable respiratory due patient coughing during treatment enforcement is to stop the treatment, it is assumed to carry out the internal target validation process again, it is necessary to readjust the form of respiration. Patients under treatment is determined to be able to improve the treatment of accuracy to induce the observed form of regular breathing and put like to see the goggles monitor capable of the respiratory form of the person.

  • PDF

The Assessment of the Breath Hold and the Free Breath Methods about the Blood Flow Evaluation by Using Phase Contrast MRI (위상대조도 자기공명영상을 이용한 심장 혈류평가에 있어서 호흡정지 기법과 비 호흡정지 기법의 비교평가)

  • Kim, Seong-Ho
    • Journal of radiological science and technology
    • /
    • v.39 no.2
    • /
    • pp.149-156
    • /
    • 2016
  • Measurement of cardiac blood flow using the magnetic resonance imaging has been limited due to breathing and involuntary movements of the heart. The present study attempted to improve the accuracy of cardiac blood flow testing through phase contrast magnetic resonance imaging by presenting the adequate breathing method and imaging variables by comparing the measurement values of cardiac blood flow. Each was evaluated by comparing the breath hold retrospective 1NEX and non breath hold retrospective 1-3NEX in the ascending aorta and descending aorta. As a result, the average blood flow amount/velocity of the breath hold retrosepctive 1NEX method in the ascending aorta were $96.17{\pm}19.12ml/sec$, $17.04{\pm}4.12cm/sec$ respectively, which demonstrates a statistically significant difference(p<0.05) with the non-breath hold retrospective method 1NEX of $72.31{\pm}13.27ml$ and $12.32{\pm}3.85$. On the other hand, the average 2NEX blood flow and mean flow velocity is $101.90{\pm}24.09$, $16.84{\pm}4.32$, 3NEX $103.06{\pm}25.49$, $16.88{\pm}4.19$ did not show statistically significant differences(p>0.05).The average blood flow amount/ velocity of the breath hold retrospective 1NEX method in the descending aorta were $76.68{\pm}19.72ml/s$, and $22.23{\pm}4.8$, which did not demonstrate a significant difference in comparison to non-breath hold retrospective method 1-3 NEX. Therefore, the non breath hold retrospective method does not significantly differ in terms of cardiac blood flow in comparison with the breath hold retrospective method in accordance with the increase of NEX, so pediatric patients or patients who are not able to breathe well must have the diagnostic value of their cardiac blood flow tests improved.