A Development of Physio-Module for Echocardiography

심초음파용 생체신호측정모듈의 개발

  • Jang, Won-Seuk (Graduate Prograrn of Biomedica1 Engineering, Graduate School, Yonsei University) ;
  • Kim, Nam-Hyun (Department of Medical Engineering, College of Medicine, Yonsei University) ;
  • Jeon, Dae-Keun (R&D Center, MEDlANA Co. Ltd.)
  • 장원석 (연세대학교 대학원 생체공학협동과정) ;
  • 김남현 (연세대학교 의과대학 의학공학교실) ;
  • 전대근 ((주)메디아나 기엽부설연구소)
  • Received : 2009.05.31
  • Accepted : 2010.07.02
  • Published : 2010.07.25

Abstract

In this study, we aimed to develope the physio-module for echocardiography. This physio-module includes multi-functions such as ECG, respiration, PCG, heart sound, and this is used to diagnose a cardiac disease in using ultrasound images synchronized with biosignals of physio-module. In this paper, the developed physio-module was verified by applying various test patterns considering each biosignal's characteristics and we could get the performance of QRS trigger delay time within international standard, EC-13 criteria. And ECG's change in physio-module and blood flow in M-mode was synchronized.

본 연구는 심초음파에 사용되는 생체신호측정모듈을 개발하고자 하였다. 생체신호측정모듈은 심전도, 호흡, 맥파, 심음 측정 기능을 포함하며, 생체신호측정모듈에서 측정된 4가지의 생체신호와 초음파 영상을 동기화시켜 심장의 수축과 이완에 따른 심장질환을 진단하는데 사용된다. 생체신호의 특성을 고려한 테스트 패턴을 적용하여 개발된 생체신호측정 회로부를 검증하였으며, 특히 심전도의 QRS trigger 지연시간은 국제규격인 EC-13에 만족하였다. 또한, 개발된 심전도, 호흡, 맥압, 심음이 동기화 되는 것을 확인하였으며, 심초음파의 M-mode 영상에서 혈류의 흐름과 심전도의 변화가 일치함을 확인하였다.

Keywords

References

  1. J. G. Webster(ed.), "Medical Instrumentation application and design", Houghton Mifflin Company, second edition, 1992.
  2. B. B. Winter and J. G. Webster, "Driven-Right-Leg Circuit Design", IEEE Trans. Biomedical Engineering, Vol. BME-30, no. 1, pp. 62-66, January 1983. https://doi.org/10.1109/TBME.1983.325168
  3. J. E. Yount and M. R. Newman, "Development of a physiologically appropriate circuit for testing thoracic impedance respiration monitors", Proceedings of the 14th Annual Meeting of the International Federation for Medical and Biological Engineering, Espoo, Finland; 11-16 August, 1985.
  4. Association for the Advancement of Medical Instrumentation. "Apnea monitoring by means of thoracic impedance pneumography",(AAMI TIR4:1989). Arlington, VA: AAMI; 2000.
  5. M. E. Travel, "Clinical Phonocardiography and External Pulse Recoding", 2nd ed, Chicago: Year Book, 1972.
  6. H. Vermarien, "Phonocardiography", in J. G, Webster(ed.), Encyclopedia of Medical Devices and Instrumentation, New York: Wiley, pp. 2265-2277, 1988.
  7. Association for the Advancement of Medical Instrumentation. "Cardiac monitors, heart rate meters, and alarms",(ANSI/AAMI EC13:2002). Arlington, VA: AAMI; 2002.
  8. Jae K. Oh, James B. Seward, A. Jamil Tajik, "The Echo Manual", Third Edition, 2008, Wolters Kluwer