• Title/Summary/Keyword: 조사야크기

Search Result 43, Processing Time 0.022 seconds

Wedge Factors in Various Field Sizes (조사야의 크기에 따른 쐐기 인자)

  • 조철우
    • Progress in Medical Physics
    • /
    • v.6 no.2
    • /
    • pp.29-34
    • /
    • 1995
  • The wedge factor is defined as a ratio of the absorbed dose in a phantom at a depth of reference point on the central axis with the wedge in the place to the absorbed dose at the same point with the wedge removed. We attempted to show the wedge factors dependence on the field sizes. The wedge factors were measured at various field sizes on 6MV and 15MV x-ray of Varian Clinac 1800 and 5MV x-ray of Philips SL75/5. The single wedge factor measured for a reference field size(10cmx10cm) may not be valid for all field sizes. For the thick wedge, especially an autowedge on Philips SL75/5 for maximum field size width 30cm. the error can be significant(6.6%). Therefore, in the presence of a wedge filter in the beam, a field size dependent wedge factor may be necessary in the treatment dose calculations.

  • PDF

Variation in Depth Dose Data between Open and Wedge Fields for 6 MV X-Rays (6MV X선에 있어서 쇄기형 조사야와 개방 조사야 사이의 깊이 선량률의 차이)

  • U, Hong;Ryu, Sam-Uel;Park, In-Kyu
    • Radiation Oncology Journal
    • /
    • v.7 no.2
    • /
    • pp.279-285
    • /
    • 1989
  • Central axis depth dose data for 6 MV X-rays, including tissue maximum ratios, were measured for wedge fields according to Tatcher's equation. In wedge fields, the differences in magnitude which increased with depth, field size, and wedge thickness increased when compared with the corresponding open field data. However, phantom scatter correction factors for wedge fields differed less than $1\%$ from the corresponding open field factors. The differences in central axis percent depth dose between two types of fields indicated beam hardening by the wedge filter The deviation of percent depth doses and scatter correction factors between the effective wedge field and the nominal wedge field at same angle was negligible. The differences were less than $3.20\%$ between the nominal or effective wedge fields and the open fields for percent depth doses to the depth 7cm in $6cm{\times}6cm$ field. For larger $(10cm{\times}10cm)$ field size, however, the deviation of percnet depth doses between the nominal or effective wedge fields and the open fields were greater-dosimetric errors were $3.56\%$ at depth 7cm and nearly $5.30\%$ at 12cm. We suggest that the percent depth doses of individual wedge and wedge transmission factors should be considered for the dose calculation or monitor setting in the treatment of deep seated tumor.

  • PDF

Nationwise Survey of the X-ray Beam Collimator Utilization in General Diagnostic Radiograph (진단방사선 일반촬영에서의 X-ray Beam Collimator 사용 전국 실태조사)

  • Kim, Jee Hye;Sung, Dong-Wook;Kim, Jeong Wook;Shin, Jin Ho;Lee, Soon Keun;Jung, Kyung Il;Uhm, Jong Kwan;Lee, Ki Nam;Seong, Ho Jin;Kim, Youn Hyun;Kim, Hyeog Ju
    • Progress in Medical Physics
    • /
    • v.24 no.2
    • /
    • pp.119-126
    • /
    • 2013
  • Due to the introduction of CR and DR, it has been neglected the use of the X-ray beam collimator and field size. This study examines nationwide survey of the proper use of collimator and field size by area in a specific field of plain radiography and the current status. Authors emphasized the need for the field size criteria, and propose a standard reference field size in each specific radiologic examination. Total 333 medical institutions (included in Seoul, Gyeonggi-do, Jeolla, Chungcheong, Gangwon-do, Busan area), were investigated in relation to the status of the X-ray beam collimation field size, type specific inspection areas, medical facilities, and image analyses by type to figure out whether they use the adjustment of image field to the specific examination. To assess the awareness and the impact of radiation exposure to the collimation adjustable, 168 radiographers who was working in 10 general hospitals, 10 hospitals, and 10 clinics, were surveyed how they haver adjusted the actual field size. We examine that 61.3% of medical institutions used the "Proper collimation" and only 49.9% of them employed proper one in lumbar spine densely crowded by major organs. 69% among general hospitals, and 65% among hospitals using DR system were using proper collimation. Radiographers recognized that proper adjustment of collimation could reduce the harmful radiation dose on patients. In the survey, 97.6% of respondents were aware of this fact, but only 83.3% of respondents did the adjustment of the size of the collimation field. The using of proper collimation field was low in the nationwide survey, so the effort to reduce the radiation dose on the patients is urgently needed. A unified standard for the field accompanied by thorough education should be needed.

Variation of Image Quality and Dose by Applying Multi-Leaf Collimator for Digital Mammography (디지털 유방촬영장치에서 다엽 조리개 적용으로 인한 화질과 선량의 변화)

  • Kwon, Soon Mu;Kim, Boo Soon;Park, Hyung Jun;Kang, Yeong Han
    • Journal of the Korean Society of Radiology
    • /
    • v.9 no.7
    • /
    • pp.535-540
    • /
    • 2015
  • Collimator has important functions with control primary X-ray that decrease radiation exposure dose for patients and reduce scatter ray and make better quality of image. But there are no regulations for X-ray mammography device of collimator, so widely used device adopt rectangularly controlled collimator. Though digital X-ray mammography device expand supply recently, rectangularly controlled collimator of film/screen mode still used. After searching for real condition of beam field with digital mammography, we made a multi-leaf collimator which is able to adjust the beam field in accordance with size and shape of breast, and we measuring up the transitions of image quality, average glandular dose(AGD) and, Dose area product(DAP). There are no significant differences between rectangularly controlled collimator and multi-leaf collimator, and DAP value decreased by 50.72%. As conclusion, there needs to expand the use of multi-leaf collimator for optimum adoption of beam field in digital mammography, and also need to develop an automatic regulation of beam field for reduce of exposure dose to patients.

A Comparison of coincidence between the Light field & the Radiation field using film and BIS (필름과 BIS 영상장치를 이용한 광/방사선조사야 일치성 비교평가)

  • Bang, Dong-Wan;Seok, Jin-Yong;Jeong, Yun-Ju;Choi, Byeong-Don;Park, Jin-Hong
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.16 no.2
    • /
    • pp.33-41
    • /
    • 2004
  • Purpose : Film has been the primary tool in coincidence testing between the light field and the radiation field, which constitutes the quality assurance list of a linear accelerator. But there is a great chance of errors being different among the observer when using film. Thus this study set out to use the BIS(Beam Image System) in addition to film in comparing and evaluating coincidence results between the two fields and in searching for the improvement measures. Materials & Methods : Photon beam of 6 and 15MV was exposed to film and the BIS using a linear accelerator. The light and radiation fields were each $50{\times}50,\;100{\times}100,\;and\;200{\times}200mm^2$. The gantry angle was $0^{\circ}$ when using film and $0^{\circ}\;and\;270^{\circ}$ when using the BIS. The devices adopted to test coincidence between the two fields were a ruler and film scanner when using film. With the BIS, the width of the scanned light and radiation fields was measured for errors with setting the X and Y axis. Results : The visual measurements of the observer with film resulted that the radiation field was bigger than the light field and that their maximum error was 1.9mm. The results were the same with the measurements using the film scanner except for the average error, which was less than 1.9mm. On the contrary, the measurements using the BIS showed that the light field was bigger than the radiation field at the gantry angle of $0^{\circ}\;and\;270^{\circ}$. The maximum error was 0.96mm, and the error range was $<{\pm}2mm$ both in the X and Y axis. The average error of ${\Delta}X$, Y was the smallest in the order of the visual film measurements, film scanner measurements, and BIS measurements Conclusion . This requires a careful measurement for accurate quality assurance since errors are much different according to each observer that tests coincidence between visual fields with film. And an observer needs to use another image device or develop a measuring device of his own if it seems necessary for accurate measurements.

  • PDF

Evaluation on Organ Dose and Image Quality of Lumbar Spine Radiography Using Glass Dosimeter (유리선량계를 이용한 요추검사의 장기선량 및 영상의 평가)

  • Kim, Jae-Kyeom;Kim, Jeong-Koo
    • Journal of radiological science and technology
    • /
    • v.39 no.1
    • /
    • pp.1-11
    • /
    • 2016
  • The purpose of this study was to provide resources for medical exposure reduction through evaluation of organ dose and image resolution for lumbar spine around according to the size of the collimator in DR system. The size of the collimator were varied from $8^{\prime\prime}{\times}17^{\prime\prime}$ to $14^{\prime\prime}{\times}17^{\prime\prime}$ by 1" in AP and lateral projection for the lumbar spine radiography with RANDO phantom. The organ dose measured for liver, stomach, pancreas, kidney and gonad by the glass dosimeter. The image resolution was analyzed using the Image J program. The organ dose of around lumbar spine were reduced as the size of the collimator is decreased in AP projection. There were no significant changes decreasing rate whenever the size of the collimator were reduced 1" in the gonad. The organ dose showed higher on liver and kidney near the surface in lateral projection. There were decreasing rate of less than 5% in liver and kidney, but decreasing rate was 24.34% in the gonad whenever the size of the collimator were reduced 1". Organ dose difference for internal and external of collimator measured $549.8{\mu}Gy$ in the liver and $264.6{\mu}Gy$ in the stomach. There were no significant changes organ dose difference that measured $1,135.1{\mu}Gy$ in the gonad. Image Quality made no difference because SNR and PSNR were over than 30 dB when the collimator size is less than $9^{\prime\prime}{\times}17^{\prime\prime}$ on AP projection and $10^{\prime\prime}{\times}17^{\prime\prime}$ on lateral projection. Therefore, we are considered that the recommendations criterion for control of collimator were suggested in order to reduce unnecessary X-ray exposure and to obtain good image quality because lumbar spine radiography contains a lot of peripheral organs rather than other area radiography.

Surface and Percentage Depth Doses for Multileaf Collimator Conjunction with Conventional Block (다엽 콜리메이터와 제작차폐물의 동시 사용시 표면선량 변화)

  • 양광모;서현숙
    • Progress in Medical Physics
    • /
    • v.13 no.2
    • /
    • pp.62-68
    • /
    • 2002
  • A muiltileaf collimator (MLC) is used as a replacement for conventional blocks. The MLC, however may not be appropriate for a fine field shaping. For the fine field shaping, conventional block can be added under the MLC. But it may significantly affect on the dosimetric characteristics such as surface dose of skin, buildup region and percent depth doses. We performed the study to evaluate the surface dose and the maximum depth dose using MLC conjunction with conventional blocks for various field sizes and energies. We confirmed the surface dose was increased by using the additional conventional block under the MLC ranging from 10 to 35.6% according to various field sizes and radiation beam energies. The surface dose was effectively reduced by application of 2 or 3 m thickness of lead plate as electron filter.

  • PDF

Treatment Planning Correction Using MRI in the Radiotherapy of Cervical Cancer (자궁 경부암의 방사선 치료계획에서 자기공명 영상을 이용한 조사야 교정)

  • Shin, Sei-One;Cho, Kil-Ho;Park, Chan-Won
    • Journal of Yeungnam Medical Science
    • /
    • v.12 no.2
    • /
    • pp.203-209
    • /
    • 1995
  • Purpose: To evaluate the role of MRI in the management of cervical cancer treated by conventional four-field whole pelvic irradiation. Method and material: Between 1993-march and 1994-february, 20 patients(4 Stage I B, 3 Stage II A, 13 Stage II B) with invasive cervical cancer were eligible for evaluation of accuracy of conventionally designed lateral treatment field without MRI determination. Results: 5 out of 20 Patients had inadequate margin without MRI. The position of uterine fundus was more important than cervix in correction of field size and the center of treatment field. Conclusion: This Preliminary data show MRI determination of uterine position prior to radiotherapy planning is essential in the case of four-field whole pelvic irradiation technique.

  • PDF

Evaluation of the Usefulness of Tungsten Nanoparticles as an Alternative to Lead Shielding Materials in Electron Beam Therapy (전자선 치료시 납 차폐체 대체물질로서의 텅스텐 나노입자의 유용성 평가)

  • Kim, Ji-Hyang;Kim, Na-Kyoung;Lee, Gyu-Yeong;Jung, Da-Bin;Heo, Yeong-Cheol
    • Journal of the Korean Society of Radiology
    • /
    • v.15 no.7
    • /
    • pp.949-956
    • /
    • 2021
  • The purpose of this paper is to evaluate whether tungsten nanoparticles have a shielding effect on scattered light generated at high doses as an alternative material to lead used to shield scattered light in electron beam therapy. A plate was manufactured to set the position of the dosimeter and the size of the radiation field to be constant. The glass dosimeter was placed at 12 points, which were 1, 2, and 4 cm apart from the center of the field of 10 × 10 cm2 in the cross direction. A total of 12 types of tungsten nanoparticle shields were developed with a thickness of 0.75 mm to 4.00 mm and a size of 10 × 10 cm2 using 0.4, 0.75, and 1 mm materials. Using a linear accelerator, measurements were made four times at 6 MeV and four times at 12 MeV, and the dose intensity was investigated at 100 MU. The 4 mm shielding plate showed the highest shielding effect at 1 cm from the irradiation field. The 1 mm shielding plate at 2 cm from the irradiation field had the lowest shielding effect. As the thickness of the tungsten shielding plate increased, the electron beam's shielding effect increased sharply. It was confirmed that tungsten nanoparticles can reduce the amount of scattered light generated by electron beam therapy. Therefore, this study will provide basic data when follow-up studies are conducted on the shielding ability of tungsten nanoparticles.

Analysis of Transport Characteristics for Double Gate MOSFET using Analytical Current-Voltage Model (해석학적 전류-전압모델을 이용한 이중게이트 MOSFET의 전송특성분석)

  • Jung Hak-Kee
    • Journal of the Korea Institute of Information and Communication Engineering
    • /
    • v.10 no.9
    • /
    • pp.1648-1653
    • /
    • 2006
  • In this paper, transport characteristics have been investigated using analytical current-voltage model for double gate MOSFET(DGMOSFET). Scaling down to 100nm of gate length for MOSFET can bring about various problems such as a threshold voltage roll-off and increasing off current by tunneling since thickness of oxide is down by 1.fnm and doping concentration is increased. A current-voltage characteristics have been calculated according to changing of channel length,using analytical current-voltage relation. The analytical model has been verified by calculating I-V relation according to changing of oxide thickness and channel thickness as well as channel length. A current-voltage characteristics also have been compared and analyzed for operating temperature. When gate voltage is 2V, it is shown that a current-voltage characteristic in 77K is superior to in room temperature.