Wedge Factors in Various Field Sizes

조사야의 크기에 따른 쐐기 인자

  • 조철우 (연세대학교 원주의과대학 방사선종양학교실)
  • Published : 1995.12.01

Abstract

The wedge factor is defined as a ratio of the absorbed dose in a phantom at a depth of reference point on the central axis with the wedge in the place to the absorbed dose at the same point with the wedge removed. We attempted to show the wedge factors dependence on the field sizes. The wedge factors were measured at various field sizes on 6MV and 15MV x-ray of Varian Clinac 1800 and 5MV x-ray of Philips SL75/5. The single wedge factor measured for a reference field size(10cmx10cm) may not be valid for all field sizes. For the thick wedge, especially an autowedge on Philips SL75/5 for maximum field size width 30cm. the error can be significant(6.6%). Therefore, in the presence of a wedge filter in the beam, a field size dependent wedge factor may be necessary in the treatment dose calculations.

쐐기인자는 중심축선상의 어느 특정 깊이에서 측정한 흡수선량으로써 쐐기가 제거된 선량과 쐐기가 부착된 선량의 비로써 정의된다. 본 연구에서는 조사야의 크기별 쐐기인자의 의존성을 살펴보았다. 쐐기인자들은 Varian Clinac1800의 6MV와 15MV 광자선과 Philips SL75/5 의 5MV광자선의 쐐기에 대하여 여러 조사야에 따라 구하였다. 특정 조사야 (10cmx10cm) 에서 구한 단일 쐐기인자를 모든 조사야에 적용시키는 것은 타당치가 않다. 두꺼운 쐐기일 경우 특히 SL75/5 의 5MV 광자선에서 autowedge 의 30cmx30cm 조사야 일 때 오차가 중요하였다(6.6%). 따라서 이러한 쐐기의 치료선량을 계산할 때는 조사야 크기에 따른 쐐기인자를 적용하여야 한다.