A Study of Performance Characteristics for Electronic Personal Dosimeters in Photon and Electron Radiation Field (광자 및 베타 방사선에 대한 전자개인선량계의 성능특성연구)

  • Kim, Hyun-Ki;Kim, Bong-Hwan;Lee, Jai-Ki
    • Journal of Radiation Protection and Research
    • /
    • v.22 no.2
    • /
    • pp.85-95
    • /
    • 1997
  • TLD and film badges have been traditionally used as formal dosimeters in personal monitoring and are still most widely used. Recently, electronic personal dosimeters based upon Si diode or miniature G-M tube were developed and are getting attractions due to their merits of active nature ; indication of dose rates and the commutative dose, and facilitation of record keeping and radiological control. Response characteristics of the electronic dosimeters including reproducibility, accuracy, linearity, energy and angular dependencies, detection threshold, and response time were examined for three commercial types ; EPD2, STEPHEN6000, and PD-3i. The results were compared with the relevant requirements of IEC standards and Ontario Hydro standards to conclude that their general performances were good. Some specific deficiencies, e.g. incapability of shallow dose measurement of STEPHEN6000, and PD-3i, however, should be corrected to be used as a formal dosimeter.

  • PDF

The Study of Energy Compensation Filter Thickness for Each Energy Area of Low Energy X-ray Beam Optimization on Active Electronic Personal Dosimeter (능동형 전자식 개인피폭선량계의 저에너지 X선 영역별 최적화를 위한 에너지보상 필터 두께에 대한 연구)

  • Kim, Jung-Su;Park, Youn-Hyun;Chae, Hyun-Sic
    • Journal of the Korean Society of Radiology
    • /
    • v.16 no.5
    • /
    • pp.519-526
    • /
    • 2022
  • Electronic personal dosimeter (EPD) provide real time monitoring and a direct indication of the accumulated dose or dose rate in terms of personal dose. Most EPD do not perform well in low energy photon radiation fields present in medical radiation environments. It has poor responsibility and large error rate for low energy photon radiation of medical radiation environments. This study evaluated to optimal additional filtration for EPD using silicon PIN photodiode detector form 40 to 120 kVp range in medical radiation environments. From 40 to 80 kVp energy range, Al 0.2 mm and Sn 1.0 mm overlapped filtration showed good responsibility to dose rate and from 80 kVp to 120 kVp energy range, Al 0.2 mm and Sn 1.6 mm overlapped filtration showed good responsibility to dose rate.

The comparison of angular dependence for optical stimulated luminescence dosimeter(OSLD) and electronic personal dosimeter(EPD) used in Diagnostic Radiology (영상의학과에서 사용되는 광자극 형광선량계와 전자식 개인선량계의 방향 의존성 비교)

  • Kwon, Soon-mu;Park, Jeong-kyu;Kim, Boo-soon
    • Journal of Digital Contents Society
    • /
    • v.16 no.3
    • /
    • pp.463-470
    • /
    • 2015
  • The angular dependence of active dosimeters, EPD, is analysed and compared with that of passive dosimeters, OSLD, after evaluating their relative response and uncertainty of measurement, where it is known that the personal use of them has been increased recently. There appeared a minor variation for average relative response of OSLD in the horizontal and vertical directions within the range $0^{\circ}{\sim}{\pm}90^{\circ}$, which are 0.97 and 0.95 respectively. The variations of angular dependence in the same situations with OSLD are 0.65 and 0.62, respectively, which also reveals a negligible effect on the overall uncertainty. EPDs within the interval $0^{\circ}{\sim}{\pm}60^{\circ}$ for horizontal and vertical directions are 0.94 and 0.97, respectively. These satisfy the requirements of IEC 61526. Uncertainties about the dependence of direction from horizontal and vertical directions are 0.44, 0.40, respectively. The impact of these uncertainties on the overall uncertainty was negligible. However, we observed a significant change in reactivity: the relative reactivities for $+90^{\circ}$ and $-90^{\circ}$ from the horizontal direction are 0.60, 0.37, while that form vertical direction is 0.06. The direction dependence of OSLD was superior to EPD in the range of $0^{\circ}{\sim}{\pm}90^{\circ}$. There appeared a rapidly changing structural features in EPD response for a certain direction. Therefore, we conclude that concurrent use of passive dosimeters and auxiliary dosimeter provides accurate data for personal dose measurements.

Development of Electronic Personal Dosimeter with Hybrid Preamplifier using Semiconductor Detector (반도체 검출기를 이용한 Hybrid 전치증폭기형 전자식 개인선량계 개발)

  • Lee, B.J.;Kim, B.H.;Chang, S.Y.;Kim, J.S.;Rho, S.R.
    • Journal of Radiation Protection and Research
    • /
    • v.27 no.1
    • /
    • pp.51-57
    • /
    • 2002
  • An electronic personal dosimeter(EPD) with hybrid type preamplifier adopting a semiconductor detector as a radiation detector has been developed, manufactured and test-evaluated. The radiation detection characteristics of this EPD has been performance-tested by using a reference photon radiation field. After several test-irradiations to a $^{137}Cs$ gamma radiation source the radiation detection sensitivity of this EPD appeared to be $3.8\;cps/Gy{\cdot}h^{-1}$. The linearity of radiation response was kept within 8% of the dose equivalent ranges of $10{\mu}Sv{\sim}4Sv$ and the angular dependence was under less than 4% in angles of ${\pm}60^{\circ}$. It was confirmed that the energy response range was in $60{\sim}1,250keV$ given in the ISO standard. This EPD satisfied the international criteria for the EPD in the mechanical and the environmental performance test for 9 test categories according to IEC 61526.

Performance Analysis of Electronic Personal Dosimeter(EPD) for External Radiation Dosimetry (전자개인선량계(EPD)의 외부피폭방사선량 평가 성능분석)

  • Lee, Byoung-Il;Kim, Taejin;Lim, Young-Khi
    • Journal of Radiation Protection and Research
    • /
    • v.40 no.4
    • /
    • pp.261-266
    • /
    • 2015
  • As performance of electronic personal dosimeter (EPD) used for auxiliary personal dosimeter in nuclear power plants (NPPs) has been being continuously improved, we investigated application cases in Korea and other countries and also tested it in NPPs to assess the performance of EPD for external radiation dosimetry. Result of performance tests done in domestic NPPs was similar to those obtained by IAEA in cooperation with EURADOS (IAEA-TECDOC-1564). In addition, EPD/TLD dose ratio has shown similar tendency of EPD/Film-badge dose ratio from the research by the Japan Atomic Power Company (JAPC) and EPD provided more conservative value than TLD or Film-badge. Although some EPD's failures have been discussed, EPD has shown continuous improvement according to the report of Institute of Nuclear Power Operation (INPO) and data from domestic NPPs. In conclusion, It is considered that the general performance of EPD is adequate for external radiation dosimetry compared with that of TLD, providing appropriate performance checking procedure and alternative measures for functional failure.