DOI QR코드

DOI QR Code

Usefulness of volumetric BMD measurement by using low dose CT image acquired on L-spine Bone SPECT/CT

L-spine Bone SPECT/CT에서 획득된 저선량 CT 영상을 이용한 용적 골밀도 결과의 유용성

  • Hyunsoo Ko (Department of Nuclear Medicine, Asan Medical Center) ;
  • Soonki Park (Department of Nuclear Medicine, Asan Medical Center) ;
  • Eunhye Kim (Department of Nuclear Medicine, Asan Medical Center) ;
  • Jongsook Choi (Department of Nuclear Medicine, Asan Medical Center) ;
  • Wooyoung Jung (Department of Nuclear Medicine, Asan Medical Center) ;
  • Dongyun Lee (Department of Nuclear Medicine, Asan Medical Center)
  • 고현수 (서울아산병원 핵의학과) ;
  • 박순기 (서울아산병원 핵의학과) ;
  • 김은혜 (서울아산병원 핵의학과) ;
  • 최종숙 (서울아산병원 핵의학과) ;
  • 정우영 (서울아산병원 핵의학과) ;
  • 이동윤 (서울아산병원 핵의학과)
  • Received : 2023.08.01
  • Accepted : 2023.09.12
  • Published : 2023.11.30

Abstract

Purpose: CT scan makes up for the weak point of the nuclear medicine image having a low resolution and also were used for attenuation correction on image reconstruction. Recently, many studies try to make use of CT images additionally, one of them is to measure the bone mineral density(BMD) using Quantitative CT(QCT) software. BMD exams are performed to scan lumbar and femur with DXA(Dual-Energy X-Ray Absorptiometry) in order to diagnose bone disease such as osteopenia, osteoporosis. The purpose of this study is to identify the usefulness of QCT_BMD analyzed with low dose CT images on L-spine Bone SPECT/CT comparing with DXA_BMD. Materials and Methods: Fifty five women over 50 years old (mean 66.4 ± 9.1) who took the both examinations(L-spine Bone SPECT/CT with SIEMENS Intevo 16 and DXA scan with GE Lunar prodigy advance) within 90 days from April 2017 to July 2022, BMD, T-score and disease classification were analyzed. Three-dimensional BMD was analyzed with low dose CT images acquired on L-spine Bone SPECT/CT scan on Mindways QCT PROTM software and two-dimensional BMD was analyzed on DXA scan. Basically, Lumbar 1-4 were analyzed and the patients who has lesion or spine implants on L-spine were excluded for this study. Pearson's correlation analysis was performed in BMD and T-score, chi-square test was performed in disease classification between QCT and DXA. Results: On 55 patients, the minimum of QCT_BMD was 18.10, maximum was 166.50, average was 82.71 ± 31.5 mg/cm3. And the minimum of DXA-BMD was 0.540, maximum was 1.302, average was 0.902 ± 0.201 g/cm2, respectively. The result shows a strong statistical correlation between QCT_BMD and DXA_BMD(p<0.001, r=0.76). The minimum of QCT_T-score was -5.7, maximum was -0.1, average was -3.2 ± 1.3 and the minimum of DXA_T-score was -5.0, maximum was 1.7, average was -2.0 ± 1.3, respectively. The result shows a statistical correlation between QCT T-score and DXA T-score (p<0.001, r=0.66). On the disease classification, normal was 5, osteopenia was 25, osteoporosis was 25 in QCT and normal was 10, osteopenia was 25, osteoporosis was 20 in DXA. There was under-estimation of bone decrease relatively on DXA than QCT, but there was no significant differences statistically by chi-square test between QCT and DXA. Conclusion: Through this study, we could identify that the QCT measurement with low dose CT images QCT from L-Spine Bone SPECT/CT was reliable because of a strong statistical correlation between QCT_BMD and DXA_BMD. Bone SPECT/CT scan can provide three-dimensional information also BMD measurement with CT images. In the future, rather than various exams such as CT, BMD, Bone scan are performed, it will be possible to provide multipurpose information via only SPECT/CT scan. In addition, it will be very helpful clinically in the sense that we can provide a diagnosis of potential osteoporosis, especially in middle-aged patients.

Keywords

References

  1. Jaszczak RJ. The early years of single photon emission computed tomography (SPECT): an anthology of selected reminiscences. Phys Med Biol. 2006;51:R99-115  https://doi.org/10.1088/0031-9155/51/13/R07
  2. Terry Jones. History and future technical innovation in positron emission tomography. J Med Imaging (Bellingham). 2017 Jan;4(1):011013  https://doi.org/10.1117/1.JMI.4.1.011013
  3. Brian F. Hutton. The origins of SPECT and SPECT/CT. Eur J Nucl Med Mol Imaging (2014) 41 (Suppl 1):S3-S16. DOI 10.1007/s00259-013-2606-5 
  4. T. Beyer et al. A combined PET/CT scanner for clinical oncology. J. Nucl. Med. 41(8), 1369-1379 (2000) 
  5. Khulood Al-Riyami. The role of bone SPECT/CT in patients with persistent or recurrent lumbar pain following lumbar spine stabilization surgery.Eur J Nucl Med Mol Imaging. 2019 Apr;46(4):989-998. doi: 10.1007/s00259-018-4141-x. Epub 2018 Sep 6 
  6. Guezennec C, Keromnes N, Robin P, et al. Incremental diagnostic utility of systematic double-bed SPECT/CT for bone scintigraphy in initial staging of cancer patients. Cancer Imaging 2017; 17(1):16  https://doi.org/10.1186/s40644-017-0118-4
  7. Kazunari Ishii. Impact of CT attenuation correction by SPECT/CT in brain perfusion images. Annals of Nuclear Medicine volume 26, pages241-247 (2012)  https://doi.org/10.1007/s12149-011-0567-y
  8. Nikolas K. Knowles. Quantitative Computed Tomography (QCT) derived Bone Mineral Density (BMD) in finite element studies: a review of the literature. Journal of Experimental Orthopaedics volume 3, Article number: 36 (2016) 
  9. Engelke K et al. Advanced CT based in vivo methods for the assessment of bone density, structure, and strength. Curr Osteoporos Rep 11:246-255. doi:10.1007/s11914-013-0147-2 (2013) 
  10. O. Museyko. A low-radiation exposure protocol for 3D QCTof the spine. Osteoporos Int (2014) 25:983-992. DOI 10.1007/s00198-013-2544-x 
  11. Adams JE: Dual-energy x-ray absorptiometry, in Grampp S (ed): Radiology of Osteoporosis. Berlin, Springer Verlag, 2008, pp 105-12 
  12. Kwang J. Bone Densitometry. emin Nucl Med 41:220-228 (2011)  https://doi.org/10.1053/j.semnuclmed.2010.12.002
  13. Clinical use of quantitative QCT of the hip in the management of osteoporosis in adults - the 2015 ISCD ocfficial position part1. https://iscd.org/learn/official-positions/adult-positions/
  14. American College of Radiology. ACR Practice Guideline for the Performance of Quantitative Computed Tomography (QCT) Bone Densitometry. 2008 
  15. Lenka Vermeeren. Value of SPECT/CT for detection and anatomic localization of sentinel lymph nodes before laparoscopic sentinel node lymphadenectomy in prostate carcinoma. J Nucl Med. 2009 Jun;50(6):865-70. doi: 10.2967/jnumed.108.060673 
  16. Zeintl J, Vija AH, Yahil A et al .Quantitative accuracy of clinical Tc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med2010; 51: 921-8 
  17. Bailey DL, Willowson KP. An evidence-based review of quantitative SPECT imaging and potential clinical applications. J Nucl Med 2013; 54: 83-9.99m  https://doi.org/10.2967/jnumed.112.111476
  18. Gary F. Gates. SPECT bone scanning of the spine. Seminars in Nuclear Medicine. Volume 28, Issue 1, January 1998, Pages 78-94  https://doi.org/10.1016/S0001-2998(98)80020-2
  19. Arum Parthipun. Nuclear Medicine Blog. 2020. at https://www.nuclearmedicineblog.com/2020/11/15/spect-ct-of-the-spine/
  20. S Zgliczynski. Measurement of bone mineral density (BMD) with quantitative computed tomography (QCT) in postmenopausal osteoporosis: effect of estrogen. Endokrynol Pol. 1992;43(3):350-7 
  21. G. Guglielmi et al. Osteoporosis: diagnosis with lateral and posteroanterior dual x-ray absorptiometry compared with quantitative CT. Radiology, vol. 192, no. 3, pp. 845-50, Sep. 1994  https://doi.org/10.1148/radiology.192.3.8058958
  22. K. G. Faulkner et al. Effect of bone distribution on vertebral strength: assessment with patient-specific nonlinear finite element analysis. Radiology, vol. 179, no. 3, pp. 669-74, Jun. 1991  https://doi.org/10.1148/radiology.179.3.2027972
  23. S. D. Rockoff, et al. The relative contribution of trabecular and cortical bone to the strength of human lumbar vertebrae. Calcified Tissue Research, vol. 3, no. 1, pp. 163-175, Dec. 1969  https://doi.org/10.1007/BF02058659
  24. QCT vs. DXA: What's the Score? Mindways CT white paper. https://www.qct.com/Downloads/Whitepaper-20130402_online.pdf 
  25. G. M. Blake, et al. A longitudinal study of supine lateral DXA of the lumbar spine: a comparison with posteroanterior spine, hip and total-body DXA. Osteoporosis International, vol. 6, no. 6, pp. 462-70, Jan. 1996  https://doi.org/10.1007/BF01629579
  26. B. C. C. Khoo. Differences in structural geometrical outcomes at the neck of the proximal femur using two-dimensional DXA-derived projection (APEX) and three-dimensional QCT-derived (BIT QCT) techniques. Osteoporos Int (2012) 23:1393-1398. DOI 10.1007/s00198-011-1727-6 
  27. Hans Lundin. Laser-Supported Dual Energy X-Ray Absorptiometry (DXL) Compared to Conventional Absorptiometry (DXA) and to FRAX as Tools for Fracture Risk Assessments. PLoS One. 2015 Sep 28;10(9):e0137535. doi: 10.1371/journal.pone.0137535. eCollection 2015 
  28. B C C Khoo. Differences in structural geometrical outcomes at the neck of the proximal femur using two-dimensional DXA-derived projection (APEX) and three-dimensional QCT-derived (BIT QCT) techniques. Osteoporos Int. 2012 Apr;23(4):1393-8. doi: 10.1007/s00198-011-1727-6. Epub 2011 Jul 23 
  29. Joel Greffier. CT dose reduction using Automatic Exposure Control and iterative reconstruction: A chest paediatric phantoms study. Physica Medica Volume 32, Issue 4, April 2016, Pages 582-589  https://doi.org/10.1016/j.ejmp.2016.03.007
  30. John Damilakis. Radiation exposure in X-ray-based imaging techniques used in osteoporosis. Eur Radiol. 2010; 20(11): 2707-2714. Published online 2010 Jun 18. doi: 10.1007/s00330-010-1845-0