DOI QR코드

DOI QR Code

Development of Radiation Restrictor for Secondary Radiation Shielding of Mobile X-ray Generator

이동형 X선 발생장치의 2차 방사선 차폐를 위한 선속조절기 개발 연구

  • Received : 2018.08.28
  • Accepted : 2018.10.23
  • Published : 2018.10.31

Abstract

Mobile X-ray generators are used not in the radiation area but in open space, which causes the exposure of secondary radiation to the healthcare professionals, patients, guardians, etc., regardless of their intentions. This study aimed to investigate the shielding effect of the developed radiation restrictor to block the secondary radiation scattered during the use of mobile X-ray generator. Upon setting the condition of mobile X-ray generator with chest AP, spatial doses were measured by the existence of human equivalent phantom and radiation restrictor, and measured by the existences of phantom and radiation restrictor at the same length of 100 cm. Measurements were taken at intervals of 10 cm every $30^{\circ}$ from $-90^{\circ}$ (head direction) to $+90^{\circ}$ (body direction). Upon the study results, spatial doses in all direction were increased by 45% on average when using phantom in the same condition, however, they were decreased by 64% on average when using the developed radiation restrictor. The dose at 100 cm from the center of X-ray was $3.0{\pm}0.08{\mu}Gy$ without phantom and was increased by 40% with $4.2{\pm}0.08{\mu}Gy$ after phantom usage. The dose when using phantom and the developed radiation restrictor was $1.4{\pm}0.08{\mu}Gy$, which was decreased by 66% compared to the case without using them. Therefore, it is considered the scattered radiation can be shielded at 100-150 cm, the regulation of the distance between beds, effectively with the developed radiation restrictor when using mobile X-ray generators, which can lower the radiation exposure to the people nearby including healthcare professionals and patients.

Keywords

References

  1. Korea Centers for Disease Control and Prevention. Publication of Medical Radiation Occupational Radiation Statistics Statistical Yearbook[Internet], KCDC; 2017 [cited 2018 Jul 11] Available from: https://www.cdc.go.kr/CDC/intro/CdcKrIntro0201.jsp?menuIds=HOME001-MNU1154-MNU0005-MNU0011&cid=72953
  2. United Nations Scientific Committee on the Effects of Atomic Radiation. Annex A Medical Radiation Exposures. UNSCEAR REPORT 2008. 2010.
  3. National Council on Radiation Protection and Measurement. Ionizing Radiation Exposures of the Population of the United States. NCRP Publication 160. 2009.
  4. Kim JK, Kim JK. Evaluation on organ dose and image quality of lumbar spine radiography using glass dosimeter. Journal of Radiological Science and Technology. 2016;39(1):1-11. https://doi.org/10.17946/JRST.2016.39.1.01
  5. Park HS, Lim CH, Kang BS, You IG, Jung HR. A study on the evaluation of patient dose in interventional radiology. Journal of Radiological Science and Technology. 2012;35(4):299-308.
  6. Carlton, RR, Alder AM. Principles of Radiographic Imaging: an art and a science, 3thEdition. Delma, CA:DelmaThomsonLearning; 2001,240-7.
  7. Back CH. A study of scattered radiation effect on digital radiography imaging system. Journal of Radiological Science and Technology. 2017;40(1):71-8. https://doi.org/10.17946/JRST.2017.40.1.11
  8. International Commission on Radiological Protection. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 106. Elsevier; 2008.
  9. Ministry of Government Legislation. Rules for the Safety Management of Diagnostic Radiation Generators Article 4. 2017.
  10. Ministry of Government Legislation. Rules for the Safety Management of Diagnostic Radiation Generators Article 9. 2017.
  11. Ciraulo DL, Marini CP, Lloyd GT, Fisher J. Do surgical residents, emergency medicine physicians, and nurses experience significant radiation exposure during the resuscitation of trauma patients?. Journal of Trauma and Acute Care Surgery. 1994;36(5):703-5. https://doi.org/10.1097/00005373-199405000-00018
  12. Grazer RE, Meislin HW, Westerman BR, Criss EA. Exposure to ionizing radiation in the emergency department from commonly performed portable radiographs. Ann Emerg Med 1987;16(4):17-20.
  13. Ministry of Health and Welfare. Rules for Safety Management of Diagnostic Radiation Generators. Ministry of Health and Welfare No.528. 2017.
  14. Jang DG, Kim JH, Park ET. Analysis of the spatial dose rates according to the type of radiation source used in multi-bed hospital. Journal of Radiological Science and Technology. 2017;40(3):407-13. https://doi.org/10.17946/JRST.2017.40.3.08
  15. Jung HM, Jung JE, Hyun HJ, Won DY. The evaluation of space dose distribution for digital mammography equipment. Journal of the Korean Society of Radiology. 2015;9(1):59-63.
  16. Cho PG. Distribution of the scatter ray in fluoroscopy X-ray. The Korea contents society. 2011;11(10):349-54. https://doi.org/10.5392/JKCA.2011.11.10.349
  17. Korea Atomic Energy Research Institute. Radiation dose rate[Internet], KAERI; 2018 [cited 2018 Jul 19] Available from: https://www.kaeri.re.kr/sub/sub03/sub03_02_01.jsp
  18. Choi SK. Relationship between the distribution of space doses in X-ray rooms and the "Inverse square law of distance". The Korea Contents Society. 2013;13(8):301-7.
  19. Ministry of Health and Welfare. Medical law enforcement regulation No 00477. 2017.
  20. Choe DH, Ko SJ, Kang SS, Kim CS, Kim JH, Kim DH. Analysis of dose reduction of surrounding patients in Portable X-ray. Journal of the Korean Society of Radiology. 2013;7(2):113-20. https://doi.org/10.7742/jksr.2013.7.2.113
  21. Kim GJ, Koo BY. Apparatus for reducing scatter radiation in mobile X-ray equipment. Patent application No.10-2018-0078105. Korea Intellectual Property Office. 2018.
  22. ICRP. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP. 2007;37(2-4).