Abstract
This study focused on the issue that when a diagnostic detector is found to have a defect, a patient would be exposed to radiation and image quality would be degraded. Though dose analysis, an experiment was conducted to evaluate detector performance as Signal to Noise Ratio (SNR) and Contrast to Noise Ratio (CNR). Absorbed dose, SNR and CNR were measured using a dosimeter and a tissue equivalent phantom. The experiment was conducted to compare whether the dose value shown after being attached to the back side of the phantom matches the dose value attached behind the detector, where in the conditions of skull, chest and abdomen were set at 75 kVp, 25 mAs, 110 kVp, 8 mAs, and 80 kVp, 20 mAs, respectively. As a result, there was a difference in that the dose values attached to the back side of the detector were 0.004 mGy, 0.006 mGy, 0.003 mGy, whereas those of the back side of the phantom were 0.006 mGy, 0.016 mGy, 0.017 mGy. In order to match both values, the condition was increased and SNR and CNR also increased from 88.32, 88.10, 4.09, 1.63, 87.94, 79.97 to 93.87, 93.75, 4.91, 4.03, 92.02, 84.92. Though this study, we found that when a detector is found to have a aging, it shortens the life of equipment and increases the dose of a patient, also the improvement effect of image quality is inadequate.
본 연구는 진단용 검출기가 결함이 발생하는 경우 환자의 방사선 피폭과 영상의 화질에 영향을 줄 수 있다는 점에 착안하여 실험을 진행하였고 선량분석을 통해 검출기 성능을 신호 대 잡음비와 대조도 대 잡음비로 평가하였다. 선량계, 조직등가팬텀을 이용해 흡수선량과 신호 대 잡음비, 대조도 대 잡음비를 측정하였다. 실험방법은 팬텀 후면에 부착하여 나온 선량 값과 검출기 뒤에 부착한 선량 값이 일치하는지를 비교하였고 두개부 흉부 복부의 조건을 각각 75 kVp 25 mAs, 110 kVp에 8 mAs, 80 kVp에 20mAs로 설정하고 검사하였다. 그 결과 검출기의 뒤쪽에 부착한 선량값은 0.004 mGy, 0.006 mGy, 0.003 mGy로 나타났고, 팬텀 후면은 0.006 mGy, 0.016 mGy, 0.017 mGy로 차이가 있었다. 두 값을 일치시키기 위해서 조건이 증가하였고 신호 대 잡음비와 대조도 대 잡음비는 88.32, 88.10, 4.09, 1.63, 87.94, 79.97에서 93.87, 93.75, 4.91, 4.03, 92.02, 84.92로 증가하였다. 본 연구를 통해 검출기가 노후화 되는 경우 기기의 수명단축과 환자의 선량증가의 원인이 되었으며 또한 영상화질의 개선효과도 미비하다는 것을 알 수 있었다.