Abstract
The study is to produced a brain phantom simulating corpus striatum, which can evaluate the progression of parkinson's disease, to investigate possibility of reducing the brain exposure dose to CT while maintaining optimal image quality during PET-CT examinations. CT scans were performed by varying tube voltage (100, 120 kVp) and tube current (80, 140, 200 mAs) with $^{18}F$ FP-CIT injected into the phantom's hot sphere and background (radioactivity ratio 3:1)(reference condition; 120 kVp, 140 mAs). Estimated effective dose was calculated by using conversion factor according to each condition, and image quality was evaluated by setting SNR and CRChot image evaluation factors. Experimental results showed that the predicted effective dose below the CT imaging reference condition was reduced by at least 10% and by up to 60%, and the predicted effective dose beyond the reference condition was increased by 40%. In addition, there was no significant difference between SNR and CRChot of PET images, and it was confirmed that brain dose decreased with decrease of tube voltage and tube current. At the same time, there was no significant change in the quality of the image in terms of SNR and CRChot despite the change in scan conditions. This fact suggests that the quality of the images acquired under the existing dose conditions can be obtained even at low dose conditions and it is expected that it will be possible to use the brain PET-CT scan as a basic data for the research on reduction of dose and improvement of image quality.
본 연구는 파킨슨질환의 진행 정도를 평가할 수 있는 선조체를 모사한 brain 팬텀을 직접 제작하여, PET-CT검사 시 최적의 영상의 질을 유지하면서 CT 스캔에 의한 brain 선량 저감을 위한 가능성을 평가하였다. 팬텀의 hot sphere와 background (radi oacti vi ty rati o 3:1)에 $^{18}F$ FP-CIT를 주입하고, 관전압(100, 120 kVp)과 관전류(80, 140, 200 mAs)의 조건을 변화시키며 CT 스캔을 실험하였다(기준조건; 120 kVp, 140 mAs). 각 조건에 따라 예상유효선량을 conversion factor를 적용해 계산하였고, SNR과 CRChot의 영상평가 인자를 설정하여 영상의 질을 평가하였다. 실험결과 CT 촬영 기준조건 이하에서의 예상유효선량은 최소 10%에서 최고 60% 정도 감소하였고, 기준조건 이상에서의 예상유효선량은 40% 증가하였다. 또한 PET 영상의 SNR과 CRChot의 유의한 차이는 없었으므로, 관전압과 관전류의 감소에 따라 brain 선량이 감소함을 확인하였다. 이와 동시에 스캔 조건의 변화에도 불구하고 SNR과 CRChot 측면에서의 영상의 질에는 유의한 변화가 없었다. 이러한 사실은 낮은 선량 조건으로도 기존의 선량 조건으로 획득한 영상의 질 수준을 얻을 수 있었으므로, 추후 brain PET-CT 스캔의 선량감소와 동시에 영상의 질 향상에 관한 연구의 기초자료로 활용이 가능할 것으로 사료된다.