Abstract
The purpose of this study is to evaluate and compare the quality of digital X-ray imaging system. The image quality evaluation was conducted By using Modulation transfer function indicating the quantitative resolution of the image and the noise power spectrum showing the noise characteristics. Using a IEC61267 radiation quality was applied to the geometry to be used in clinical and geometry presented in IEC62220-1 and Additional filter, grid, the clinical dose and the MTF value of edge phantom was measured. Result of the MTF corresponding to each item(Grid, Filter, SID, kVp, mAs), the clinical condition 100cm, 180cm, measurements of the spatial frequency of the MTF IEC62220-1Geometry 150cm became similarly apparent, rather spatial frequency was also the case high in clinical conditions 100cm. NPS results, as the dose(mAs) is increased, NPS showed that reduced. The image quality evaluation using IEC61267 the Radiation quality, Image quality of the video using the clinical conditions Geometry than image quality evaluation using IEC62220-1Geometry was better. It shows that MTF and NPC in IEC and clinical condition were not significantly different. In order to apply the evaluation method of image quality applied with clinical conditions rather than the future method, to be presented evaluation of IEC standard, based on the results of the image characterization studies in this paper, the methods that shows good quality of spatial resolution and decrease NPS value as the least dose, used suitable parameters for whether or not using added filter, grid, change SID and clinical quality(kVp), dose(mAs) etc should be found. then It is believed to be able to properly maintain the actual quality of the image of the digital radiographic imaging system in clinical.
본 연구의 목적은 디지털방사선시스템의 영상 품질을 비교하고 평가함에 있다. 영상의 정량적인 분해능을 나타내는 변조전달함수, 노이즈 특성을 나타내는 잡음력 스펙트럼을 이용하여 영상 품질평가를 하였다. IEC61267 선질을 사용하여 IEC62220-1에서 제시하는 기하학적인 조건과 실제 임상에서 사용되어지는 기하학적인 조건을 사용하여 부가필터 및 그리드, 임상선량을 이용하여 edge 팬텀을 사용하여 MTF값을 측정하였다. Grid, Filter, SID, kVp, mAs에 따른 MTF 결과는 임상조건 100, 180cm 와 IEC62220-1 기하학적인 조건 150cm에서 MTF 공간주파수 측정값은 비슷하게 나타났으며, 오히려 임상조건 100, 180cm에서 공간주파수가 높은 경우도 있었다. NPS 결과는 선량이 증가함에 따라 감소함을 나타내었다. IEC61267 선질을 이용한 영상품질평가에서는 IEC62220-1기하학적인 조건을 이용한 품질평가보다 임상조건 기하학적인 조건을 사용한 영상의 품질이 좋았다. IEC 와 임상조건에서의 MTF와 NPS는 크게 차이가 나지 않았음을 알 수 있었다. 그러므로 향후 IEC 표준에서 제시하는 영상품질 평가보다는 임상 조건을 적용한 영상품질 평가방법을 적용하기 위해서 부가필터사용 유 무, 그리드사용 유 무, SID변화, 선질, 선량 등의 파라메터를 적절히 이용하여 가장 적은 선량으로 공간분해능이 좋고, NPS가 감소하는 방법들을 찾는다면 향후 실제 임상에서 사용되고 있는 디지털방사선시스템을 최적의 품질로 유지할 수 있는 참고 자료로 활용될 수 있을 것으로 사료된다.