A Comparison Study of Volumetric Modulated Arc Therapy Quality Assurances Using Portal Dosimetry and MapCHECK 2

  • Jin, Hosang (Department of Radiation Oncology, Stephenson Cancer Center, University of Oklahoma Health Sciences Center) ;
  • Jesseph, Fredrick B. (Department of Radiation Oncology, Stephenson Cancer Center, University of Oklahoma Health Sciences Center) ;
  • Ahmad, Salahuddin (Department of Radiation Oncology, Stephenson Cancer Center, University of Oklahoma Health Sciences Center)
  • Received : 2014.04.18
  • Accepted : 2014.06.03
  • Published : 2014.06.30

Abstract

A Varian Portal Dosimetry system was compared to an isocentrically mounted MapCHECK 2 diode array for volumetric modulated arc therapy (VMAT) QA. A Varian TrueBeam STx with an aS-1000 digital imaging panel was used to acquire VMAT QA images for 13 plans using four photon energies (6, 8, 10 and 15 MV). The EPID-based QA images were compared to the Portal Dose Image Prediction calculated in the Varian Eclipse treatment planning system (TPS). An isocentrically mounted Sun Nuclear MapCHECK 2 diode array with 5 cm water-equivalent buildup was also used for the VMAT QAs and the measurements were compared to a composite dose plane from the Eclipse TPS. A ${\gamma}$ test was implemented in the Sun Nuclear Patient software with 10% threshold and absolute comparison at 1%/1 mm (dose difference/distance-to-agreement), 2%/2 mm, and 3%/3 mm criteria for both QA methods. The two-tailed paired Student's t-test was employed to analyze the statistical significance at 95% confidence level. The average ${\gamma}$ passing rates were greater than 95% at 3%/3 mm using both methods for all four energies. The differences in the average passing rates between the two methods were within 1.7% and 1.6% of each other when analyzed at 2%/2 mm and 3%/3 mm, respectively. The EPID passing rates were somewhat better than the MapCHECK 2 when analyzed at 1%/1 mm; the difference was lower for 8 MV and 10 MV. However, the differences were not statistically significant for all criteria and energies (p-values >0.05). The EPID-based QA showed large off-axis over-response and dependence of ${\gamma}$ passing rate on energy, while the MapCHECK 2 was susceptible to the MLC tongue-and-groove effect. The two fluence-based QA techniques can be an alternative tool of VMAT QA to each other, if the limitations of each QA method (mechanical sag, detector response, and detector alignment) are carefully considered.

Keywords

References

  1. Bakhtiari M, Kumaraswamy L, Bailey DW, de Boer S, Malhotra HK, Podgorsak MB: Using an EPID for patient- specific VMAT quality assurance. Med Phys 38:1366-1373 (2011) https://doi.org/10.1118/1.3552925
  2. Sharma DS, Mhatre V, Heigrujam M, Talapatra K, Mallik S: Portal dosimetry for pretreatment verification of IMRT plan: a comparison with 2D ion chamber array. J Appl Clin Med Phys 11:238-248 (2010)
  3. Clemente S, Caivano R, Cozzolino M, et al: To evaluate the accuracy of dynamic versus static IMRT delivery using portal dosimetry. Clin Transl Oncol 16:208-212 (2014) https://doi.org/10.1007/s12094-013-1065-6
  4. Bailey DW, Kumaraswamy L, Bakhtiari M, Malhotra HK, Podgorsak MB: EPID dosimetry for pretreatment quality assurance with two commercial systems. J Appl Clin Med Phys 13:82-99 (2012)
  5. Bailey DW, Kumaraswamy L, Bakhtiari M, Podgorsak MB: A two-dimensional matrix correction for off-axis portal dose prediction errors. Med Phys 40:051704 (2013) https://doi.org/10.1118/1.4800493
  6. Bailey DW, Kumaraswamy L, Podgorsak MB: An effective correction algorithm for off-axis portal dosimetry errors. Med Phys 36:4089-4094 (2009) https://doi.org/10.1118/1.3187785
  7. Huang Y-C, Yeh C-Y, Yeh J-H, et al: Clinical practice and evaluation of electronic portal imaging device for VMAT quality assurance. Med Dosim 38:35-41 (2013) https://doi.org/10.1016/j.meddos.2012.05.004
  8. Iftimia I, Cirino ET, Xiong L, Mower HW: Quality assurance methodology for Varian RapidArc treatment plans. J Appl Clin Med Phys 11:130-143 (2010) https://doi.org/10.1120/jacmp.v11i4.3164
  9. Nicolini G, Vanetti E, Clivio A, et al: The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy. Radiat Oncol 3:24 (2008) https://doi.org/10.1186/1748-717X-3-24
  10. Rowshanfarzad P, Sabet M, O'Connor DJ, McCowan PM, McCurdy BM, Greer PB: Detection and correction for EPID and gantry sag during arc delivery using cine EPID imaging. Med Phys 39:623-635 (2012) https://doi.org/10.1118/1.3673958
  11. Gloi AM, Buchanan RE, Zuge CL, Goettler AM: RapidArc quality assurance through MapCHECK. J Appl Clin Med Phys 12:39-47 (2011) https://doi.org/10.1120/jacmp.v12i2.3251
  12. Jursinic PA, Sharma R, Reuter J: MapCHECK used for rotational IMRT measurements: step-and-shoot, TomoTherapy, RapidArc. Med Phys 37:2837-2846 (2010) https://doi.org/10.1118/1.3431994
  13. Rinaldin G, Perna L, Agnello G, et al: Quality assurance of RapidArc treatments: Performances and pre-clinical verifications of a planar detector (MapCHECK2). Phys Medica 30:184-190 (2014) https://doi.org/10.1016/j.ejmp.2013.05.004
  14. Shim SJ, Shim JB, Lee SH, et al: Quality assurance of volumetric modulated arc therapy for elekta synergy. Korean J Med Phys 23:33-41 (2012)
  15. Varian Medical Systems: TrueBeamTM TrueBeam STxTM Technical reference guide (Version 1.6 Volume 2: Imaging; Docu ID: B501671R01, Revision D). Varian Medical Systems, Inc., Switzerland, (2011), pp.205-207
  16. Varian Medical Systems: Eclipse algorithms reference guide (P/N B503486R01B, document version 1.0). Varian Medical Systems, Inc., Finland, (2011), pp.393-423
  17. Low DA, Harms WB, Mutic S, Purdy JA: A technique for the quantitative evaluation of dose distributions. Med Phys 25:656-661 (1998) https://doi.org/10.1118/1.598248
  18. Stock M, Kroupa B, Georg D: Interpretation and evaluation of the gamma index and the gamma index angle for the verification of IMRT hybrid plans. Phys Med Biol 50:399-411 (2005) https://doi.org/10.1088/0031-9155/50/3/001
  19. Sweet J, Nicolaou N: Arc splitting for VMAT patient QA [abstract]. Med Phys 39:3746 (2012)
  20. Jin H, Keeling VP, Johnson DA, Ahmad S: Interplay effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc quality assurance. J Appl Clin Med Phys 15:1-13 (2014)
  21. Greer PB: Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry. Med Phys 32:3558-3568 (2005) https://doi.org/10.1118/1.2128498
  22. Greer PB: Off-axis dose response characteristics of an amorphous silicon electronic portal imaging device. Med Phys 34: 3815-3824 (2007) https://doi.org/10.1118/1.2779944
  23. Deng J, Pawlicki T, Chen Y, Li J, Jiang SB, Ma C-M: The MLC tongue-and-groove effect on IMRT dose distributions. Phys Med Biol 46:1039-1060 (2001) https://doi.org/10.1088/0031-9155/46/4/310
  24. Olch AJ: Evaluation of the accuracy of 3DVH software estimates of dose to virtual ion chamber and film in composite IMRT QA. Med Phys 39:81-86 (2012) https://doi.org/10.1118/1.3666771

Cited by

  1. Verification of light & radiation field coincidence quality assurance for radiation therapy by using a-Se based DR system vol.66, pp.8, 2014, https://doi.org/10.3938/jkps.66.1303
  2. Detection of IMRT delivery errors based on a simple constancy check of transit dose by using an EPID vol.67, pp.10, 2014, https://doi.org/10.3938/jkps.67.1876
  3. Modulation index for VMAT considering both mechanical and dose calculation uncertainties vol.60, pp.18, 2014, https://doi.org/10.1088/0031-9155/60/18/7101
  4. EPID‐based dosimetry to verify IMRT planar dose distribution for the aS1200 EPID and FFF beams vol.17, pp.6, 2014, https://doi.org/10.1120/jacmp.v17i6.6336
  5. The effect of extremely narrow MLC leaf width on the plan quality of VMAT for prostate cancer vol.11, pp.None, 2016, https://doi.org/10.1186/s13014-016-0664-0
  6. Optimal Density Assignment to 2D Diode Array Detector for Different Dose Calculation Algorithms in Patient Specific VMAT QA vol.42, pp.1, 2014, https://doi.org/10.14407/jrpr.2017.42.1.9
  7. Evaluation of a 2D Diode Array Corresponding to Flattening Filter Free X-ray Beams vol.74, pp.5, 2018, https://doi.org/10.6009/jjrt.2018_jsrt_74.5.473
  8. Modeling scattered radiation from multi-leaf collimators (MLCs) to improve calculation accuracy of in-air output ratio vol.42, pp.3, 2014, https://doi.org/10.1007/s13246-019-00781-2
  9. Prediction of VMAT delivery accuracy with textural features calculated from fluence maps vol.14, pp.1, 2014, https://doi.org/10.1186/s13014-019-1441-7