DOI QR코드

DOI QR Code

Reducing Radiation Exposure Dose on Operator by Using Lateral Protection in Neuro-Intervention

뇌혈관 중재적시술에 있어 측방향 차폐체를 이용한 시술자 피폭 선량 저감화 방법 연구

  • 김종덕 (전남대학교병원 영상의학과) ;
  • 안병주 (남부대학교 방사선학과) ;
  • 이준행 (남부대학교 방사선학과)
  • Received : 2013.08.05
  • Accepted : 2014.01.25
  • Published : 2014.01.30

Abstract

The bi-plane cerebrovascular angiography radiation is done the radiation exposure at the forward and lateral direction as opposed to the one of the source. So, the exposure dose of radiation workers increases further. Therefore, the medical diagnostic radiation workers as well as patients is interested to ways to reduce the dose. The exposure dose of cerebral angiography and interventional radiology must be considered the primary radiation of X-ray tube directly, scattered primary radiation between lateral tube and lateral detector and relatively small secondary scatter radiation in the walls of room. The aim of study is that the exposure dose of primary and scatter radiation reduce as much as possible to install protection device of lateral protection than common shielding of table and ceiling. As a result, the dose of fluoroscopy was reduced approximately 3.64 times the gonads, thyroid approximately 3.13 times, 4.42 times around eyes. And the dose of DSA was reduced approximately 4.98 times the gonads, thyroid approximately 3.00 times, 1.67 times around eyes. Consequently, medical practitioners can be helpful for radiation dose-exposure for the lateral protection of bi-plane cerebrovascular angiography more than the common shield method in cerebrovascular angiography and interventional radiological procedures.

양방향 뇌혈관촬영기는 하나의 선원에서 방사선이 나오는 것이 아니라 정방향과 측방향에서 방사선 피폭이 이루어지기 때문에 방사선 관계종사자의 피폭이 더욱 많아질 수밖에 없다. 따라서 환자가 받는 선량도 중요하겠지만 시술을 시행하고 있는 방사선 관계종사자 역시 피폭선량을 줄일 수 있는 방법에 대해 많은 관심을 보이고 있다. 본 연구의 목적은 양방향 뇌혈관촬영기의 혈관검사 및 중재적방사선시술에 있어 X-선 관구에서 직접 조사되는 1차 방사선, 정방향관구와 측방향 검출기사이에서 발생하는 1차 산란방사선, 상대적으로 적지만 촬영실 벽이나 바닥에서 반사되는 2차 산란방사선 발생을 간과하지 않을 수 없기에 기존의 일반적인 차폐방법인 천정형 차폐, 테이블형 차폐방법에서 보다 더 적극적인 차폐방법인 측방향 차폐체의 방어용구를 설치하여 시술자가 받는 직접방사선 및 산란방사선에 의한 피폭선량을 최대한 줄이고자 노력하였다. 그 결과 투시측정에서는 생식샘 약 3.64배, 갑상샘 약 3.13배, 눈 약 4.42배 정도 더 감소하였고, 디지털 감산 혈관조영측정에서는 생식샘 약 4.98배, 갑상샘 약 3.00배, 눈 약1.67배의 피폭선량 감소효과를 얻어내었다. 결론적으로 양방향 뇌혈관촬영기의 혈관검사 및 중재적방사선시술시 설치하였던 측방향 차폐체의 방어용구는 일반적인 차폐방법보다 시술자의 피폭선량을 감소시키는데 많은 효과를 주었다고 사료된다.

Keywords

References

  1. Suzuki, S., et al. "Patient skin dose during neuroembolization by multiple-point measurement using a radiosensitive indicator", American Journal of Neuroradiology, Vol. 29, No. 6, pp.1076-1081, 2008. https://doi.org/10.3174/ajnr.A1045
  2. http://kans.re.kr/nabi/edu/edu01_1_02.html
  3. J.L Kim, "New concepts and principles of Optical Stimulation Luminescence Dosimeter", Radioisotope journal, Vol. 18, No. 4, pp. 75-80, 2003.
  4. http://www.osldosimetry.com/documents/MicroStar_Specifications.p
  5. Hanyang Uiversity Nuclear Engineering. "The 2007 Recommendations of the International Commission on Radiological Protection", ICRP Publication 103.
  6. Vano, E., Gonzalez, L., Beneytez, F., et al., "Lens injuries induced by occupational exposure in non-optimised interventional radiology laboratories". Brit. J. Radiol., Vol. 71, pp. 728-733, 1998. https://doi.org/10.1259/bjr.71.847.9771383
  7. Vano, E., Gonzalez, L., Guibelalde, E., et al., "Radiation exposure to medical staff in interventional and cardiac radiology", Brit. J. Radiol., Vol. 71, pp. 954-960, 1998. https://doi.org/10.1259/bjr.71.849.10195011
  8. Y.I. Jang, J.N. Song, Y.j. Kim, "Efficacy of a Protective Grass Shield in Reduction of Radiation Exposure Dose During Interventional Radiology", The Korean Society of Radiology, Vol. 5, No. 5, pp. 303-308, 2011. https://doi.org/10.7742/jksr.2011.5.5.303
  9. National Institute of Food and drug Safety Evaluation, "Performance Evaluation Technical implicit radiation treatment device", The korean Radiological Society, Radiation Safety Management Series, Vol. 26, 2011.
  10. Y.J. Sung, H.M. Park, J.D. Kim, K.M. Kim, N.G. Choi, "A study on the usage of Whole body shields for reduction of radiation exposure", The korean Society of Cardio-Vascular Interventional Technology, Vol. 13, No. 1, pp. 167-171, 2010.
  11. H.S. Byun, "Recent Trends in the Treatment of Cerebral Aneurysms: Comparison between Endovascular Coil Embolization and Surgical Clipping", Neurointervention, Vol. 4, pp. 1-5, 2009.
  12. http://sizekorea.kats.go.kr/

Cited by

  1. Design for integrated pyroprocessing plant level simulator vol.60, 2013, https://doi.org/10.1016/j.anucene.2013.05.004