A Thermal Blood Flow Sensor with Contact Force Compensation

접촉력 보정이 가능한 열적 방식의 혈류량 측정기

  • Received : 2012.02.02
  • Accepted : 2012.12.20
  • Published : 2013.03.01

Abstract

This paper proposes a thermal peripheral blood flowmeter integrated with a force sensor that is capable of contact force compensation. We fabricate this blood flowmeter using a nickel RTD (resistance temperature detector) and piezoresistive force sensor by using microfabrication technology. In an experiment, we obtained a decreasing trend for the blood flow under an increasing contact force with a linear tendency of 31.7%/N. We then performed a compensation process based on this obtained trend. As a result, the maximum variance in the blood flow at 1-3N was 9.8%. Thus we achieved consistent blood flow measurement independent of the contact force. In this work, we verified that the thermal blood flowmeter integrated with a force sensor has the ability to accurately measure the blood flow independent of the contact force.

본 논문은 접촉력 보정이 가능한 혈류량 측정기를 제안한다. 기존 혈류량 측정기는 접촉력의 영향을 최소화하기 위해 혈류량 센서를 피부에 고정했으나 이런 점은 전자기기에 적용이 용이하지 않다는 단점이 있다. 이에 본 연구에서는 혈류량 센서에 힘 센서를 집적하여 혈류량과 접촉력의 동시 측정을 통해 접촉력에 따른 혈류량 변화 오차를 보정하여 접촉력에 무관한 혈류량 측정이 가능한 소자 및 방법을 제안한다. 제작된 혈류량 측정기의 성능 분석 결과, 접촉력에 따른 혈류량 변화는 31.7%/N 의 선형적 감소경향임을 확인하였고, 이를 이용하여 접촉력에 따른 혈류량 오차를 보정한 결과, 접촉력 1~3N에서 최대 편차가 9.8%로 나타나 다른 접촉력 조건에서도 일정한 혈류량 측정이 가능함을 보였다. 제안한 소자는 접촉력에 무관하게 정확한 혈류량을 측정할 수 있어, 접촉 가능한 전자기기에의 적용이 용이하다.

Keywords

References

  1. Fromy, B., Abraham, P. and Saumet, J.-L., 2000, "Progressive Calibrated Pressure Device to Measure Cutaneous Blood Flow Changes to External Pressure Strain," Brain Res Protocol, Vol. 5, pp. 198-203. https://doi.org/10.1016/S1385-299X(00)00013-1
  2. Midttun, M., Sejrsen, P. and Colding-Jorgensen, M., 1996, "Heat-washout: a New Method for Measuring Cutaneous Blood Flow Rate in Areas with and without Arteriovenous Anastomoses," Clin Physiol, Vol.16, pp.275-289. https://doi.org/10.1111/j.1475-097X.1996.tb00574.x
  3. Nagata, K., Hattori, H., Sato, N., Ichige, Y. and Kiguchi, M., 2009, "Heat Trasfer Analysis for Peripheral Blood Flow Measurement System," Rev. Sci. Instrum, Vol. 80, 064902. https://doi.org/10.1063/1.3155458
  4. Omata, M., Matsumura, K. and Imamiya, A., 2007, "A Pressure Sensing Mouse Button for Multilevel Click and Drag," INTERACT 2007, LNCS 4662, Part I, pp. 434-446.
  5. Nitzan, M., Fairs, S. L. E. and Roberts, V. C., 1988, "Simultaneous Measurement of Skin Blood Flow by the Transient Thermal-Clearance Method and Laser Doppler Flowmetry," Med. & Biol. Eng & Comput, Vol. 26, pp. 407-410. https://doi.org/10.1007/BF02442300
  6. Holloway, G. A. and Watkins, D. W., 1977, "Laser Doppler Measurement of Cutaneous Blood Flow," J. Invest. Derma, Vol. 69, No. 3, pp. 306-309. https://doi.org/10.1111/1523-1747.ep12507665