An Assessment of Entrance Surface Dose Using the nanoDot Dosimeter

나노도트선량계를 이용한 입사표면선량의 평가

  • Kim, Jong-Eon (Department of Radiological Science, Kaya University) ;
  • Im, In-Chul (Department of Radiological Science, Dongeui University) ;
  • Park, Cheol-Woo (Department of Radiological Science, Dongeui Science College)
  • Received : 2011.10.29
  • Accepted : 2011.12.22
  • Published : 2011.12.30

Abstract

The purpose of this study is an assessment between the measured value of the nanoDot dosimeter and the calculated value of Non Dosimeter Dosimetry-Method(NDD-M) for entrance surface dose in general radiography. Measurement and calculation of the entrance surface doses were performed for head(AP), abdomen(AP), pelvis(AP), thoracic spine(AP) and lumbar spine(AP). As a result, the relative ratios of the measured value to the calculated value were acquired 1.5-2.1 for each region. Reproducibility acquired 0.035 as a coefficient of variation.

이 연구의 목적은 일반방사선촬영에서 입사표면선량에 대하여 나노도트선량계의 측정값과 Non Dosimeter Dosimetry-Method(NDD-M)의 계산값과 비교 평가하는 데 있다. 입사표면선량들의 측정과 계산은 두부(AP), 복부(AP), 골반(AP), 흉추(AP)와 요추(AP)에 대하여 수행하였다. 결과로서, 계산값에 대한 측정값의 상대비들은 각 부위에 대하여 1.5-2.1을 얻었다. 재현성은 변동계수로서 0.035를 얻었다.

Keywords

References

  1. P. Engel-Hills, "Radiation protection in medical imaging", Radiography, Vol. 12, pp.153-160, 2006. https://doi.org/10.1016/j.radi.2005.04.008
  2. I.O. Olarinoye and I. Sharifat, "A protocol for setting dose reference level for medical radiography in nigeria: a review", Bajopas, Vol. 3, No. 1, pp.138-141, 2010.
  3. 방사선안전관리 시리즈 No. 14, 환자선량 측정 가이드라인, 식품의약품안전청, 2007.
  4. 방사선안전관리 시리즈 No. 17, 흉부엑스선검사에서의 환자선량 권고량 가이드라인, 식품의약품안전청, 2008.
  5. 방사선안전관리 시리즈 No. 19, CT 엑스선검사에서의 환자선량 권고량 가이드라인, 식품의약품안전평가원, 2009.
  6. IAEA Safety series No.115, International basic safety standards for protection against ionizing radiation and for the safety of radiation sources, Vienna, 1996.
  7. REX-525R(F) user's manual, Listem Corporation.
  8. microStar user manual, Landauer Inc., 2009.
  9. PIXY the anthropomorpic training/teaching phantom, Radiology Support Devices Inc.
  10. C.S. Reft, "The energy dependence and dose reponse of a commercial optically stimulated luminescent detector for kilovolatge photon, megavoltage photon, electron, proton, and carbon beams", Med. Phys., Vol. 36, No. 5, pp.1690-1699, 2009. https://doi.org/10.1118/1.3097283
  11. 김종언, 임인철, 이효영, "진단방사선촬영에서 광자극형광선량계의 에너지의존성에 대한 보정인자", 한국방사선학회지, Vol. 5, No. 5, pp. 261-265, 2011.

Cited by

  1. Assessment of Radiation Dose Due to X-Ray Simple Series Examinations vol.8, pp.2, 2014, https://doi.org/10.7742/jksr.2014.8.2.81
  2. Dose Evaluation Using Mathematical Simulation of Radiation Exposure Body in Mammography System vol.8, pp.4, 2014, https://doi.org/10.7742/jksr.2014.8.4.155
  3. 진단방사선의 노출 조건에 따른 열형광선량계와 광자극형광 선량계의 선량값 상대비 vol.6, pp.6, 2012, https://doi.org/10.7742/jksr.2012.6.6.499
  4. 소아 5세 표준촬영 가이드라인과 임상 촬영조건의 입사표면선량과 주요 장기흡수선량 비교 vol.40, pp.3, 2011, https://doi.org/10.17946/jrst.2017.40.3.01
  5. X-선장치에서 출력선량 측정에 관한 연구 vol.14, pp.3, 2011, https://doi.org/10.7742/jksr.2020.14.3.289