Diagnostic Reference Levels for Patient Radiation Doses in Pelvis and Lumbar spine Radiography in Korea

우리나라의 골반 및 요추 엑스선검사에서의 환자선량 권고량

  • Lee, Kwang-Yong (Radiation Safety Division, National Institute of Food and Drug Safety Evaluation) ;
  • Lee, Byung-Young (Radiation Safety Division, National Institute of Food and Drug Safety Evaluation) ;
  • Lee, Jung-Eun (Radiation Safety Division, National Institute of Food and Drug Safety Evaluation) ;
  • Lee, Hyun-Koo (Radiation Safety Division, National Institute of Food and Drug Safety Evaluation) ;
  • Jung, Seung-Hwan (Radiation Safety Division, National Institute of Food and Drug Safety Evaluation) ;
  • Kim, Byung-Woo (Radiation Safety Division, National Institute of Food and Drug Safety Evaluation) ;
  • Kim, Hyeog-Ju (Radiation Safety Division, National Institute of Food and Drug Safety Evaluation) ;
  • Kim, Dong-Sup (Radiation Safety Division, National Institute of Food and Drug Safety Evaluation)
  • 이광용 (식품의약품안전평가원 의료제품연구부 방사선안전과) ;
  • 이병영 (식품의약품안전평가원 의료제품연구부 방사선안전과) ;
  • 이정은 (식품의약품안전평가원 의료제품연구부 방사선안전과) ;
  • 이현구 (식품의약품안전평가원 의료제품연구부 방사선안전과) ;
  • 정승환 (식품의약품안전평가원 의료제품연구부 방사선안전과) ;
  • 김병우 (식품의약품안전평가원 의료제품연구부 방사선안전과) ;
  • 김혁주 (식품의약품안전평가원 의료제품연구부 방사선안전과) ;
  • 김동섭 (식품의약품안전평가원 의료제품연구부 방사선안전과)
  • Published : 2009.12.31

Abstract

Purpose : Pelvis and lumbar spine radiography, among various types of diagnostic radiography, include gonads of the human body and give patients high radiation dose. Nevertheless, diagnostic reference levels for patient radiation dose in pelvis and lumbar spine radiography has not yet been established in Korea. Therefore, the radiation dose that patients receive from pelvis and lumbar radiography is measured and the diagnostic reference level on patient radiation dose for the optimization of radiation protection of patients in pelvis and lumbar spine radiography was established. Methods : The conditions and diagnostic imaging information acquired during the time of the postero-anterior view of the pelvis and the postero-anterior and lateral view of the lumbar spine at 125 medical institutions throughout Korea are collected for analysis and the entrance surface dose received by patients is measured using a glass dosimeter. The diagnostic reference levels for patient radiation dose in pelvis and lumbar spine radiography to be recommended to the medical institutes is arranged by establishing the dose from the patient radiation dose that corresponds to the 3rd quartile values as the appropriate diagnostic reference level for patient radiation dose. Results : According to the results of the assessment of diagnostic imaging information acquired from pelvis and lumbar spine radiography and the measurement of patient entrance surface dose taken at the 125 medical institutes throughout Korea, the tube voltage ranged between 60~97 kVp, with the average use being 75 kVp, and the tube current ranged between 8~123 mAs, with the average use being 30 mAs. In the posteroanterior and lateral views of lumbar spine radiography, the tube voltage of each view ranged between 65~100 kVp (average use: 78 kVp) and 70~109 kVp (average use: 87 kVp), respectively, and the tube current of each view ranged between 10~100 mAs(average use: 35 mAs) and between 8.9~300 mAs(average use: 64 mAs), respectively. The measurements of entrance surface dose that patients receive during the pelvis and lumbar spine radiography show the following results: in the posteroanterior view of pelvis radiography, the minimum value is 0.59 mGy, the maximum value is 12.69 mGy and the average value is 2.88 mGy with the 1st quartile value being 1.91 mGy, the median being 0.59 mGy, and the 3rd quartile value being 3.43 mGy. Also, in the posteroanterior view of lumbar spine radiography, the minimum value is 0.64 mGy, the maximum value is 23.84 mGy, and the average value is 3.68 mGy with the 1st quartile value being 2.41 mGy, the median being 3.40 mGy, and the 3rd quartile value being 4.08 mGy. In the lateral view of lumbar spine radiography, the minimum value is 1.90 mGy, the maximum value is 45.42 mGy, and the average value is 10.08 mGy with the 1st quartile value being 6.03 mGy, the median being 9.09 mGy and the 3rd quartile value being 12.65 mGy. Conclusions : The diagnostic reference levels for patient radiation dose to be recommended to the medical institutes in Korea is 3.42 mGy for the posteroanterior view of pelvis radiography, 4.08 mGy for the posteroanterior view of lumbar spine radiography, and 12.65 mGy for the lateral view of lumbar spine radiography. Such values are all lower than the values recommended by 6 international organizations including World Health Organization, where the recommended values are 10 mGy for the posteroanterior view of pelvis radiography, 10 mGy for the posteroanterior view of lumbar spine radiography and 30 mGy for the lateral view of lumbar spine radiography.

목 적 : 골반 및 요추 엑스선검사는 진단 엑스선검사 중 생식선을 포함하고 있고 환자가 받는 방사선량이 많은 검사로서 우리나라에는 골반 및 요추 엑스선검사에서의 환자선량 권고량이 마련되어 있지 않다. 따라서 국내 의료기관에서의 골반 및 요추 엑스선검사 시 환자가 받는 방사선량을 측정 평가하고 골반 및 요추 엑스선검사에서 환자의 방사선 방어 최적화를 위한 환자선량 권고량을 확립한다. 방 법 : 전국 125개 의료기관에서 골반 전후면 촬영, 요추 전후면 및 측면 촬영시 촬영조건과 진단영상정보를 조사 분석하고 환자가 받는 입사표면선량을 유리선량계를 사용하여 측정 평가한다. 환자가 받는 방사선량 중 제3사분위값에 해당하는 선량값을 의료기관에 권고할 골반 및 요추 엑스선검사에서의 환자선량 권고량으로 확립한다. 결 과 : 전국 125개 의료기관에서 골반 및 요추 엑스선검사 시 진단영상정보를 조사하고 환자가 받는 입사 표면선량을 측정한 결과 골반 전후면 엑스선검사에서는 관전압이 60~97 kVp, 평균 75 kVp를 사용하였고, 관전류-시간곱(mAs)는 8~123 mAs, 평균 29.7 mAs를 사용하였다. 요추 전후면 및 측면 엑스선검사에서는 관전압을 각각 65~100 kVp, 평균 78 kVp와 70~109 kVp, 평균 87 kVp를 사용하였고, mAs도 각각 10~100 mAs, 평균 35.2 mAs와 8.9~300 mAs, 평균 64.1 mAs를 사용하였다. 골반 및 요추 엑스선검사에서의 환자가 받는 입사표면선량을 측정한 결과, 골반 전후면 엑스선검사에서는 최소값 0.59 mGy, 최대값 12.69 mGy, 평균값 2.88mGy이었으며, 제1사분위값은 1.91 mGy, 중앙값은 2.67 mGy, 제3사분위값은 3.42 mGy이었다. 요추 전후면 엑스선검사에서는 최소값 0.64 mGy, 최대값 23.84 mGy, 평균값 3.68 mGy이었으며, 제1사분위값은 2.41 mGy, 중앙값은 3.40 mGy, 제3사분위값은 4.08 mGy이었다. 요추 측면 엑스선검사에서는 최소값 1.90 mGy, 최대값 45.42 mGy, 평균값 10.08 mGy이었으며, 제1사분위값은 6.03 mGy, 중앙값은 9.09 mGy, 제3사분위값은 12.65 mGy이었다. 결 론 : 우리나라의 의료기관에서 골반 전후면 엑스선검사에서 의료기관에 권고할 환자선량 권고량은 3.42 mGy, 요추 전후면 엑스선검사에서는 4.08 mGy, 요추 측면촬영에서는 12.65 mGy로 세계보건기구 등 6개 국제기구가 공동으로 권고한 골반 전후면 검사 10 mGy, 요추 전후면 검사 10 mGy 및 요추 측면 검사 30 mGy 보다는 낮았다.

Keywords

References

  1. 과학기술부 연구보고서 : 국민 피폭선량 종합 DB구축, 2005
  2. ICRP Publication 60 : 1990 Recommendations of the International Commission on Radiological Protection, Oxford, 1991
  3. UNSCEAR 2000 Report : Vol. I Sources and Effects of Ionization Radiation, Annex D Medical radiation exposure, NewYork, 2000
  4. ICRP Publication 73: Radiological protection and safety in medicine, Oxford, 1996
  5. IAEA Safety Series No.115 : International Basic Safety Standards for Protection against Ionizing Radiation and for the Safety of Radiation Sources, Vienna, 1996
  6. ICRP Publication 103: 2007 Recommendations of the International Commission on Radiological Protection, Oxford, 2007
  7. ICRP Supporting Guidance 2: Radiation and Your Patient, A Guide for Medical Practioners, Oxford, 2001
  8. Nationwide Evaluation of X-ray Trends(NEXT): http://www.fda.gov/cdrh/radhlth/next.html
  9. NRPB Report: Doses to Patients from Medical X-ray Examinations in the UK-2000 Review, Chilton, 2002
  10. 독일연방방사선방어청 보고서 : Bundesamt fur Strahlenschutz Bekanntmachung der diagnostischen Referenzwerte fur radiologische und nuklearmedizinische Untersuchungen (방사선검사 및 핵의학검사에 적용되는 진단참고준위 공시), Salzgitter, 2003
  11. Dose Ace 유리선량계 판독기 매뉴얼, 아사히테크노글라스주식회사, 2004
  12. 식품의약품안전청연구보고서: 골반 요추 엑스선검사에서의 환자선량 평가, 서울, 2008
  13. 의료피폭의 가이드라인, 사단법인 일본방사선기사회 의료피폭가이드라인위원회, 2000
  14. Patient Exposure and Dose Guide-2003, Conference of Radiation Control Program Directors, Inc. 2003