Evaluation of the Device Failure Using Stimulus Artifact in the Cochlear Implantee

인공와우 이식자에서 자극 잡파를 이용한 고장 평가

  • 허승덕 (동아대학교 의과대학 이비인후과학교실) ;
  • 김상렬 (메디컬 이비인후과) ;
  • 안중기 (동아대학교 의과대학 이비인후과학교실) ;
  • 정동근 (동아대학교 의과대학 의공학교실) ;
  • 강명구 (동아대학교 의과대학 이비인후과학교실)
  • Published : 2007.06.30

Abstract

The aim of this study is to analyze the correlation between current intensity and amplitude of stimulus artifact on the cochlear implantee, and to find out basic information to check the device failure. Subjects were a prelingual child and 3 postlingual adults with more than severe hearing losses. The charge-balanced biphasic pulses were presented at stimulus rates of 11 pulses per second, each pulse width of $25{\mu}s$ with monopolar mode(MP1+2). Current intensities were delivered at 27.5, 33.7, 41.3, 50.5, 61.9, $75.8{\mu}A$. Stimulus artifacts were recorded by evoked potential system. This procedure was performed just before the initial stimulation, and then, the amplitude of stimulus artifacts were compared with each current intensity. The amplitude of stimulus artifacts was increased significantly according to the current intensity (p<0.01). The results suggest that the change of the amplitude of stimulus artifact can be used as a good cue to check the device failure in the cochlear implantee.