Positive Peaked Electrically Compound Action Potentials in Cochlear Implant Recipients (인공와우 이식자에서 Positive Peaked 청신경 복합활동전위)

  • Heo, Seung-Deok
    • Phonetics and Speech Sciences
    • /
    • v.1 no.2
    • /
    • pp.25-30
    • /
    • 2009
  • Animal experiments have shown that the positive peaked electrically compound action potentials (ECAPs) can be recorded in round window, intracochlear, and nerve trunk by stimulating a monopolar pulse. However, positive peaked ECAPs of cochlear implant recipients have never been reported because ECAPs are recorded from intracochlear electrodes after bipolar stimulation. In our experiment, the positive peaked ECAPs were recorded from 18 intracochlear electrodes in cochlear implant recipients with multiple cochlear anomalies. Thresholds in each channel were measured and the latency of P-, N-wave, and amplitude of P-N were analyzed. These results were identical with the electrically auditory brainstem response (EABR) on the input-output characteristics. In conclusion, the positive peaked ECAPs from the cochlear implant recipients are antidromic ECAPs recorded by perimodiolar electrodes stimulating cochlear implants with multiple anomalies. Therefore, positive peaked ECAPs can be used as useful audiological tools to evaluate the eighth nerve ending.

  • PDF

Development of Electrical Stimulator for Auditory Stimulation (청각 자극용 전기자극기 개발)

  • Heo, Seung-Deok;Jung, Dong-Keun;Kim, Lee-Suk;Kim, Gwang-Nyeon;Kang, Myung-Koo;Kim, Jae-Ryong;Kim, Gi-Ryon
    • Speech Sciences
    • /
    • v.11 no.3
    • /
    • pp.201-211
    • /
    • 2004
  • This paper introduces a development of an electrical stimulator for auditory stimulation. The electrical stimulator is useful in neurotological diagnosis, audiological evaluation, candidate selection for cochlear implantation, optimal device selection and decision making of MAP strategy for severe-to-profound hearing impaired persons. The development was based on sound parameters of auditory brainstem responses and auditory electrophysiological characteristic such as effective firing of auditory nerve and recording evoked potentials during refractory period of neuron. Besides pulse parameter could adjustable by programming for more varied electrical stimulation evoked response audiometry. Using the electrical stimulator, electrical square pulse was applied to promontory, and electrically evoked auditory brainstem response and electrically middle latency response were successfully recorded in cats.

  • PDF

Evaluation of the Device Failure Using Stimulus Artifact in the Cochlear Implantee (인공와우 이식자에서 자극 잡파를 이용한 고장 평가)

  • Heo, Seung-Deok;Kim, Sang-Ryeol;Ahn, Joong-Ki;Jung, Dong-Keun;Kang, Myung-Koo
    • Speech Sciences
    • /
    • v.14 no.2
    • /
    • pp.35-42
    • /
    • 2007
  • The aim of this study is to analyze the correlation between current intensity and amplitude of stimulus artifact on the cochlear implantee, and to find out basic information to check the device failure. Subjects were a prelingual child and 3 postlingual adults with more than severe hearing losses. The charge-balanced biphasic pulses were presented at stimulus rates of 11 pulses per second, each pulse width of $25{\mu}s$ with monopolar mode(MP1+2). Current intensities were delivered at 27.5, 33.7, 41.3, 50.5, 61.9, $75.8{\mu}A$. Stimulus artifacts were recorded by evoked potential system. This procedure was performed just before the initial stimulation, and then, the amplitude of stimulus artifacts were compared with each current intensity. The amplitude of stimulus artifacts was increased significantly according to the current intensity (p<0.01). The results suggest that the change of the amplitude of stimulus artifact can be used as a good cue to check the device failure in the cochlear implantee.