Analysis of Resonance Based Micromechanical Bio-Chemical Sensing Structures

공진 기반 마이크로기계 생화학 센싱 구조물의 해석

  • 여민구 (국민대학교 자동차전문대학원) ;
  • 신윤혁 (국민대학교 기계설계대학원) ;
  • 임홍재 (국민대학교 기계자동차공학부) ;
  • 임시형 (국민대학교 기계자동차공학부)
  • Published : 2008.11.05

Abstract

A microcantilever is a well-known MEMS structure for sensing bio-chemical molecules. When bio-chemical molecules are adsorbed on the microcantilever's surface, resonance frequency shift is generated. There are two issues in this phenomena. The first one is which one between mass change and surface stress change effects is more dominant on the resonance frequency shift. The second one is what will be the performance change when the boundary condition is changed from cantilevers to double clamped beams. We have studied the effect of surface stress change and compared it with that of mass change by using FEM analysis. Furthermore, for microstructures having different boundary conditions, we have studied Q-factor, which determines the detection limit of micro/nano mechanical sensors.

Keywords