• Title/Summary/Keyword: rotational viscometer

Search Result 41, Processing Time 0.03 seconds

Comparison of the Viscosity of Ceramic Slurries using a Rotational Rheometer and a Vibrational Viscometer (회전형 레오미터와 진동형 점도계를 이용한 세라믹 슬러리의 점도 비교)

  • Ji, Hye;Lim, Hyung Mi;Chang, Young-Wook;Lee, Heesoo
    • Journal of the Korean Ceramic Society
    • /
    • v.49 no.6
    • /
    • pp.542-548
    • /
    • 2012
  • The viscosity of a ceramic slurry depends on the slurry concentration, particle shape and size, hydrodynamic interactions, temperature, shear rate, pre-treatment condition and the method of measurement with the selected equipment. Representative ceramic slurries with low to high viscosity levels are selected from colloidal silica, barium titanate slurry and glass frit paste. Rotational rheometers and vibrational viscometers are used to compare the measured viscosity for various ceramic slurries. The rotational rheometer measured the viscosity according to the change of the shear rate or the rotational speed. On the other hand, the vibrational viscometer measured one point of the viscosity in a fixed vibrational mode. The rotational rheometer allows the measurement of the viscosity of a ceramic paste with a viscosity higher than 100,000 cP, while the vibrational viscometer provides an easy and quick method to measure the viscosity without deformation of the ceramic slurry due to the measurement method. It is necessary to select suitable equipment with which to measure the viscosity depending on the purpose of the measurement.

Evaluation of Apparent Viscosity Properties for Electro-Rheological Fluid (ER유체의 겉보기 점도특성 평가에 관한 연구)

  • ;Morishita, Shin
    • Tribology and Lubricants
    • /
    • v.14 no.2
    • /
    • pp.42-48
    • /
    • 1998
  • Electro-Rheological (ER) fluid is a class of functional fluid whose apparent viscosity can be varied by the applied electric field strength. The ER fluid is classified into two types; one is a dispersive fluid and the other is a homogeneous. Dispersive ER fluid is a colloidal suspension of fine semiconducting particles in a dielectric liquid and liquid crystal (LC) is classed as homogeneous type ER fluid. LC has been originally developed for some electronic display devices. Various mechanical components applying ER fluid have been developed, and the their performance typically depends on the characteristics of ER fluid which have generally been evaluated by a rotational viscometer. However, the ER fluid introduced into various mechanical components undergoes not only simple shear flow but press flow or oscillating flow. For the evaluation of ER fluid, the authors developed an reciprocating type viscometer. The amplitude is controlled on 5 mm at the frequency from 50 to 1000 Hz. In the present paper, the performance of several types of ER fluid is evaluated by the reciprocating type viscometer and compared with those evaluated by a rotational viscometer.

Flow Properties of Traditional Kochujang : Effect of Fermentation Time (숙성기간에 따른 재래 고추장의 유동성 변화)

  • 유병승;최원석;류영기
    • Journal of the Korean Society of Food Science and Nutrition
    • /
    • v.28 no.3
    • /
    • pp.554-558
    • /
    • 1999
  • Flow properties of traditional kochujang at various fermentation times(0~12 weeks) were determined with rotational cylindrical(RC) and serrated plate plate(PP) viscometer. Magnitudes of consistency index(K) of power law model and Cassson parameters(yield stress and viscosity) measured by PP viscometer were higher than those using RC viscometer. All kochujang samples during fermentation were much shear thinning with values of flow behavior index(n) as low as 0.2~0.34. K value and Casson yield stress measured by PP viscometer had good correlations(r=0.94; r=0.91) with fermentation time. No significant changes in flow model parameters measured by RC viscometer were observed for kochu jang during fermentation. Magnitudes of flow model parameters measured by PP viscometer more closely correlated with fermantation times of kochujang than did RC viscometer.

  • PDF

The Comparison of Properties of ER and MR Fluids Using a Rotational Viscometer (회전식 점도측정기를 이용한 ER 및 MR유체의 특성 비교)

  • 이영록;전도영
    • The Korean Journal of Rheology
    • /
    • v.11 no.2
    • /
    • pp.73-81
    • /
    • 1999
  • The intention of this research is to give criteria of designing devices using ER fluids and MR fluids. The Properties of commercial ER fluids and MR fluids are compared using a rotational viscometer. The yield strength is compared upon changes of shear rate, temperature and applied fields. MR fluids seem less sensitive to temperature change than ER fluids. In cases of MR and ER fluid dampers, the time delay and damping force are measured in tension and compression mode when the applied field changes.

  • PDF

Permanent Deformation Properties of Asphalt Binder Modified by Pyrolysis Carbon Black of Waste Tires (열분해 카본블랙을 이용한 아스팔트 바인더의 소성변형 특성)

  • Lee, Dong-Hang;Kim, Jung-Ku;Lee, Kwan-Ho
    • Journal of the Korea Academia-Industrial cooperation Society
    • /
    • v.14 no.8
    • /
    • pp.4028-4032
    • /
    • 2013
  • Recycling method for pyrolyzed carbon black from pyrolysis process of waste tires is needed. Carbon black from pyrolysis of waste tires was used to modify and improve the permanent deformation properties of asphalt binder. 0%, 5%, 10%, 15% and 20% of pyrolyzed carbon black was mixed. Couple of laboratory tests, such as softening point, flash point test, rotational viscometer test and dynamic shear rheometer test, were carried out. The use of pyrolyzed carbon black incresed the softening point, rotational viscosity at 135oC, and resistance of permanent deformation.

Identification of Dynamic property of MR Fluid (MR 유체의 동특성)

  • Ahn, Young-Kong;Ha, Jong-Yong;Ahn, Kyoung-Kwan;Yang, Bo-Suk;Shiraishi, Toshihiko;Morishita, Shin
    • Proceedings of the Korean Society for Noise and Vibration Engineering Conference
    • /
    • 2005.05a
    • /
    • pp.576-579
    • /
    • 2005
  • MR (Magneto Rheological) fluids are well known as a smart fluid and their application researches to control vibration have been conducted by many researchers. However, their dynamic properties have not been identified clearly yet. Therefore, the MR effect is investigated by using a rotational viscometer and a single degree of freedom system with an MR damper. The results obtained from the experimental study show that stiffness and viscous damping coefficients of the system with an MR damper are changed according to the variation of the applied current.

  • PDF

Effect of Zeta-Potential on the Viscosity of Clay-Water Suspension

  • Lee, Young-Seek;Ree, Jong-Baik;Ree, Taik-Yue
    • Bulletin of the Korean Chemical Society
    • /
    • v.3 no.3
    • /
    • pp.83-88
    • /
    • 1982
  • Viscosity and zeta-potential of 11.0 wt. % aqueous bentonite suspension containing various electrolytes and hydrogen-ion concentration were measured by using a Couette type automatic rotational viscometer and Zeta Meter, respectively. The effects of pH and elcctrolytes on the rheological properties of the suspension were investigated. A system, which has a large zeta-potcntial, has a small intrinsic relaxation time ${\beta}$ and a small intrinsic shear modulus $1/{\alpha}$ in the Ree-Eyring generalized viscosity equation, i.e., such a system has a small viscosity value, since ${\eta}={\beta}/{\alpha}$. In general, a stable suspension system has large zeta-potential. The stability condition of clay-water suspension can be estimated by viscometric method since stable suspension generally has small viscosity. The correlation between the stability, viscosity and zeta-potential has been explained by the Ree-Eyring theory of viscous flow.

Flow Mechanism of Dilatant Systems. (Ⅰ) Starch Suspension in Water

  • Bang, Jeong-Hwang;Kim, Eung-Ryul;Hahn Sang-Joon;Ree, Tai-kyue
    • Bulletin of the Korean Chemical Society
    • /
    • v.4 no.5
    • /
    • pp.212-217
    • /
    • 1983
  • Depending on the range of shear rates, temperatures and concentrations, the potato starch suspension in water behaves as a typical dilatant system. The flow curves of the suspension at various concentrations and temperatures were obtained by using a Couette type rotational viscometer. The flow mechanism of the suspension is explained by a structure model of starch granules in the suspension. Based on the experimental results, a general flow equation for the dilatant system is proposed. By analyzing the temperature dependency of the relaxation time, the activation enthalpy and activation entropy for flow in the starch-water suspension were calculated, the former being about 10 kcal/mol.

Comparison of Vane-shear Strength Measured by Different Methods in Deep-sea Sediments from KODOS area, NE Equatorial Pacific (북동태평양 KODOS지역 심해 퇴적물의 베인 전단강도 측정 방법에 따른 결과 비교)

  • Chi, Sang-Bum;Jung, Hoi-Soo;Kim, Hyun-Sub;Moon, Jai-Woon
    • The Sea:JOURNAL OF THE KOREAN SOCIETY OF OCEANOGRAPHY
    • /
    • v.4 no.4
    • /
    • pp.390-399
    • /
    • 1999
  • Siliceous and calcareous deep-sea core sediments were collected by a multiple corer from the KODOS (Korea Deep Ocean Study) area, northeast equatorial Pacific, to compare vane shear strengths measured by two different apparatuses and in different places of on-board and on-land laboratories. The apparatuses were 1) a hand-held vane with four blades of $2.0{\times}2.0$ cm, and 2) a motorized shear vane system with four blades of $1.0{\times}0.88$ attached on a rotational viscometer. Depth profiles of shear strengths of core samples determined by the apparatuses do not show any consistent difference. Also, there is no consistent difference between shear strength values measured on-board and on-land laboratories after storing the core samples for three months in a cold room by a motorized shear vane system. However, there are considerable differences between depth profiles of shear strengths measured at four different points (holes) of a core sample. Moreover, significant differences among the profiles of different tube samples from a multiple corer within a sampling station were observed. Heterogeneity in physical properties of each depth and sediment column, possibly due to bioturbation and bottom current flows, is likely responsible for the differences in the geotechnical properties.

  • PDF

Electrorheological Properties of ER Fluid under High Shear Flow (고속 전단유동에서 ER유체의 전기유변 특성)

  • Kim Y. C.;Kim K. W.
    • Proceedings of the Korean Society of Tribologists and Lubrication Engineers Conference
    • /
    • 2003.11a
    • /
    • pp.229-234
    • /
    • 2003
  • As electrorheological fluid(ER fluid) has a characteristic that apparent viscosity varies when electric field applied, so rheological characteristic(yield stress & viscosity) changes in proportion to the electric field applied and the response time is very short within a few miliseconds . In case of using ER fluid for journal bearing as lubricant, it is estimated that it's possible to realize very effective journal bearing system that is not complicate and has a very quick response time. It is necessary to examine the influence of rheological characteristic that varies with electric field applied on bearing characteristic to apply ER fluid to journal bearing, however there are few studies for about that. As for the journal bearing, it comes under high shear flow mode that has shear rate range of $10^3\~10^4s^{-1}$ because rotational speed is very high and clearance is small. But most of the studies for about ER fluid issued until now is about the range of $10\~10^2s^{-1}$. So, there are a lot of difficulties to understand the characteristic offish shear flow mode and furthermore it is restricted to make an experiment for about the characteristic of ER fluid because of the limitation of experimental equipment. The equipment was prepared to make an experiment lot high shear flow mode that has the range of $10^3\~10^4s^{-1}$ using ER fluid that is composed of silicon oil with dispersed particle of starch. Using the above system, the fluid characteristic of ER fluid was studied.

  • PDF