• Title/Summary/Keyword: radionuclide activity meters

Search Result 1, Processing Time 0.014 seconds

Quality Management of Radionuclide Activity Meter using Ge-68/Ga-68 Rod Sources (Ge-68/Ga-68 Rod Sources을 이용한 방사능측정기의 정도관리)

  • Jung, Seung Hwan;Jin, Gye Hwan
    • Journal of the Korean Society of Radiology
    • /
    • v.12 no.5
    • /
    • pp.575-582
    • /
    • 2018
  • This article compared accuracy of 5 types of radionuclide activity meters that are being used in medical institutions and proposed the correction factor for each radionuclide activity meters type, using Ge-68/Ga-68 radiation sources for scanner setting, regular scanner correction, attenuation correction, and normalization. The calibration constant between baseline values and measured values by CRC-15R, CRC-15 PET, CRC-712M, CRC-15 Beta, and CRC-25PET was 0.99999(P<0.0001), which showed very high linearity. In the accuracy test, CRC-15R, CRC-15 PET, CRC-712M, CRC-15 Beta, and CRC-25PET model showed -3.232%, -1.342%, -2.815%, -2.913%, and -3.089% respectively.