The survey of the surface doses of the dental x-ray machines (치과 방사선 촬영기의 표면선량 변화)

  • Lee Jae-Seo;Kang Byung-Cheol;Yoon Suk-Ja
    • Imaging Science in Dentistry
    • /
    • v.35 no.2
    • /
    • pp.87-90
    • /
    • 2005
  • Purpose : The purpose of this study was to investigate variability of doses with same exposure parameters and evaluate radiographic density according to the variability of doses. Materials and methods Twenty-eight MAX-GLS (Shinhung Co, Seoul, Korea), twenty-one D-60-5 (DongSeo Med, Seoul, Korea), and eleven REX-601 (Yoshida Dental MFG, Tokyo, Japan) dental x-ray machines were selected for this study Surface doses were measured under selected combinations of tube voltage, tube current, exposure time, and constant distance 42 cm from the focal spot to the surface of the Multi-O-meter (Unfors Instruments, Billdal, Sweden). Radiographic densities were measured on the films at maximum, minimum and mean surface doses of each brand of x-ray units. Results With MAX-GLS, the maximum surface doses were thirteen to fourteen times as much as the minimum surface doses. With D-60-S, the maximum surface doses were three to eight times as much as the minimum surface doses. With REX-601, the maximum surface doses were six to ten times as much as the minimum surface doses. The differences in radiographic densities among maximum, mean, and minimum doses were significant (p<0.01). Conclusion : The surface exposure doses of each x-ray machine at the same exposure parameters were different within the same manufacturer's machines.

  • PDF

Comparison of cone beam CT and conventional CT in absorbed and effective dose (Cone beam CT와 일반 CT의 흡수선량 및 유효선량 비교평가)

  • Kim, Sang-Yeon;Han, Jin-Woo;Park, In-Woo
    • Imaging Science in Dentistry
    • /
    • v.38 no.1
    • /
    • pp.7-15
    • /
    • 2008
  • Purpose: This study provides comparative measurements of absorbed and effective doses for newly developed cone beam computed tomography (CT) in comparison with these doses for conventional CT. Materials and Methods: Thermoluminescent dosimeter rods (TLD rod: GR-200, Thermo Fisher Scientific Inc., Waltham, MA, USA) were placed at 25 sites throughout the layers of Male ART Head and Neck Phantom (Radiology Support Devices Inc., Long Beach, USA) for dosimetry. Implagraphy, DCT Pro (Vatech Co., Hwasung, Korea) units, SCT-6800TXL (Shimadzu Corp., Kyoto, Japan), and Crane x 3+(Soredex Orion Corp., Helsinki, Finland) were used for radiation exposures. Absorption doses were measured with Harshaw 3500TLD reader (Thermo Fisher Scientific Inc., Waltham, MA, USA). Radiation weighted doses and effective doses were measured and calculated by 2005 ICRP tissue weighting factors. Results: Absorbed doses in Rt. submandibular gland were 110.57 mGy for SCT 6800TXL (Implant), 24.56 mGy for SCT 6800TXL (3D), 22.39 mGy for Implagraphy 3, 7.19 mGy for DCT Pro, 5.96 mGy for Implagraphy 1, 0.70 mGy for Cranex 3+. Effective doses $(E_{2005draft)$ were 2.551 mSv for SCT 6800TXL (Implant), 1.272 mSv for SCT 6800TXL (3D), 0.598 mSv for Implagraphy 3, 0.428 mSv for DCT Pro and 0.146 mSv for Implagraphy 1. These are 108.6, 54.1, 25.5, 18.2 and 6.2 times greater than panoramic examination (Cranex 3+) doses (0.023mSv). Conclusion: Cone beam CT machines recently developed in Korea, showed lower effective doses than conventional CT. Cone beam CT provides a lower dose and cost alternative to conventional CT, promising to revolutionize the practice of oral and maxillofacial radiology.

  • PDF

Radiation absorbed doses of cone beam computed tomography (Cone beam형 전산화단층촬영에 의한 흡수선량)

  • Lee, Eui-Tae;Kim, Gyu-Tae;Choi, Yong-Suk;Hwang, Eui-Hwan
    • Imaging Science in Dentistry
    • /
    • v.37 no.2
    • /
    • pp.87-92
    • /
    • 2007
  • Purpose: To measure the absorbed doses of cone beam computed tomography (CBCT), which is recently being more frequently used, and to compare them with those of panoramic radiography. Materials and Methods: To measure the absorbed doses of CBCT ($PSR-9000N^{TM}$, Asahi Roentgen Ind. Co., Japan), we placed TLD chips on the skin regions above the parotid and thyroid glands, and on the dorsum of tongue in a dental head phantom. We used two image acquisition modes of the Dental and Panoramic modes of CBCT, which differed in the field of view. Also, panoramic radiographs (Auto IIIN, Asahi Roentgen Ind. Co., Japan) were taken to compare with the absorbed doses of CBCT. Result: In the Dental mode of CBCT, the absorbed doses of the parotid gland, dorsum of tongue, and thyroid gland were 3.53, 3.13, and 0.36 mGy, respectively. In the Panoramic mode of CBCT, they were 9.57, 9.15, and 0.85 mGy, respectively. The panoramic mode showed higher absorbed doses than those of the Dental mode. In the panoramic radiography, the absorbed doses of the parotid gland, dorsum of tongue, and thyroid gland were 1.21, 1.19, and 0.16 mGy, respectively. And they were about 1/3 of the Dental mode and 1/9 of the Panoramic mode of CBCT. Conclusion: Absorbed doses of CBCT are higher than those of panoramic radiography, and dependent upon the field of view.

  • PDF

Isolation of Torilin from Torilis japonica Fruit and Its Analgesic and Anti-inflammatory Activities (사상자 중 Torilin의 분리 및 진통소염작용)

  • Cho, Sung-Ig;Kang, Sam-Sik;Kim, Kyung-Ran;Kim, Tae-Hee;Lee, Eun-Bang
    • Korean Journal of Pharmacognosy
    • /
    • v.30 no.2
    • /
    • pp.137-144
    • /
    • 1999
  • Torilin was isolated from haxane fraction of Torilis Fructus extract. Torilin produced inhibition of the acetic acid-induced and phenylquinone-induced writhing syndrome at the oral doses of 30 and 90 mg/kg in mice. It also increased the pain threshold at the oral doses of 30, 90 and 270 mg/kg in the tail pressure method and the Randall-Selitto method. However, it did not show a hypothermic action at the oral doses of 30 and 90 mg/kg in mice. The compound exhibited strong anticarrageenan activity at the oral doses of 90 and 270 mg/kg in rats, and had inhibitory effect on the vascular permeability at the oral doses of 30 and 90 mg/kg in mice. It also showed potent inhibition of leucocyte emigration in CMC-pouch at the doses of 3 and 9 mg/rat, sc. The acute toxicity of torilin was very weak: the $LD_{50}$ values were more than 5000 mg/kg, po and 2000 mg/kg, ip in mice. From the above mentioned results, it was suggested that torilin had potent analgesic and anti-inflammatory activities.

  • PDF

Absorbed and effective dose from newly developed cone beam computed tomography in Korea (최근 개발된 cone beam computed tomography의 흡수선량 및 유효선량 평가)

  • Lee, Jong-Nyeong;Han, Won-Jeong;Kim, Eun-Kyung
    • Imaging Science in Dentistry
    • /
    • v.37 no.2
    • /
    • pp.93-102
    • /
    • 2007
  • Purpose: Cone beam computed tomography (CBCT) provides a lower dose and cost alternative to conventional CT, promising to revolutionize the practice of oral and maxillofacial radiology. The purpose of this study was to evaluate the absorbed and effective doses of Implagraphy and VCT (Vatech Co., Hwasung, Korea) and compare them with those of panoramic radiography. Materials and Methods: Thermoluminescent dosimeter (TLD) chips were placed at 27 sites throughout the layers of Female ART Head and Neck Phantom for dosimetry. Implagraphy, VCT units, and Planmeca Proline XC panoramic unit were used for radiation exposures. Radiation weighted doses and effective doses were measured and calculated using 1990 and 2005 ICRP tissue weighting factors. Results: Effective doses in Sv (ICRP 2005, ICRP 1990) were 90.19, 61.62 for Implagraphy at maxillay molar area, 123.20, 90.02 for Implagraphy at mandibular molar area, 183.55, 139.26 for VCT and 40.92, 27.16 for panoramic radiography. Conclusion: Effective doses for VCT and Implagraphy were only about 2.2 to 4.5 times greater than those for panoramic radiography. VCT and Implagraphy, CBCT machines recently developed in Korea, showed moderately low effective doses.

  • PDF

Effective dose from direct and indirect digital panoramic units

  • Lee, Gun-Sun;Kim, Jin-Soo;Seo, Yo-Seob;Kim, Jae-Duk
    • Imaging Science in Dentistry
    • /
    • v.43 no.2
    • /
    • pp.77-84
    • /
    • 2013
  • Purpose: This study aimed to provide comparative measurements of the effective dose from direct and indirect digital panoramic units according to phantoms and exposure parameters. Materials and Methods: Dose measurements were carried out using a head phantom representing an average man (175 cm tall, 73.5 kg male) and a limbless whole body phantom representing an average woman (155 cm tall, 50 kg female). Lithium fluoride thermoluminescent dosimeter (TLD) chips were used for the dosimeter. Two direct and 2 indirect digital panoramic units were evaluated in this study. Effective doses were derived using 2007 International Commission on Radiological Protection (ICRP) recommendations. Results: The effective doses of the 4 digital panoramic units ranged between $8.9{\mu}Sv$ and $37.8{\mu}Sv$. By using the head phantom, the effective doses from the direct digital panoramic units ($37.8{\mu}Sv$, $27.6{\mu}Sv$) were higher than those from the indirect units ($8.9{\mu}Sv$, $15.9{\mu}Sv$). The same panoramic unit showed the difference in effective doses according to the gender of the phantom, numbers and locations of TLDs, and kVp. Conclusion: To reasonably assess the radiation risk from various dental radiographic units, the effective doses should be obtained with the same numbers and locations of TLDs, and with standard hospital exposure. After that, it is necessary to survey the effective doses from various dental radiographic units according to the gender with the corresponding phantom.

Reference dose levels for dental periapical radiography in Chonnam Province (전남 지방에서 치근단방사선사진의 참고 선량 수준)

  • Han, Mi-Ra;Kang, Byung-Cheol;Lee, Jae-Seo;Yoon, Suk-Ja;Kim, Young-Hee
    • Imaging Science in Dentistry
    • /
    • v.39 no.4
    • /
    • pp.195-198
    • /
    • 2009
  • Purpose : To establish reference doses of periapical radiography in Chonnam Province, Korea. Materials and Methods : The target-skin distances were measured for dental patient's 1235 exposures including 345 mandibular molar areas. Each periapical radiation exposure was simulated with exactly the same patients exposure parameters and the simulated radiation doses were measured utilizing Mult-O-Meter (Unfors Instruments, Billadal, Sweden). The measurements were done in 44 dental clinics with 49 dental x-ray sets in Chonnam Province for one or two weeks at each dental clinic during year 2006. Results : The third quartile patient surface doses were 2.8 mGy for overall periapical exposures and 3.2 mGy for periapical mandibular molar exposures. Conclusion : The third quartile patient surface doses in Chonnam Province can be used as a guide to accepted clinical practice to reduce patient radiation exposure for the surveyed reference doses were below the recommended dental periapical radiography dose of 7 mGy by IAEA.

  • PDF

General Pharmacology of LB71350, a New HIV-1 Pretense Inhibitor

  • Kim, Hee-Jin;Oh, Jeng-In;Park, Hee-Dong;Kang, Ju-Seop;Ko, Hyun-Chul;Lee, Chang-Ho
    • Biomolecules & Therapeutics
    • /
    • v.7 no.3
    • /
    • pp.271-277
    • /
    • 1999
  • Safety evaluation of LB71350, a new HIV-1 protease inhibitor, was performed in mice, rats and dogs. For the general behavior of mice, LB71350 at an oral dose of 200 mg/kg did not show any significant effects on muscle tone and locomotor activity. In terms of central nervous system, at oral doses of 200 mg/kg and 1000 mg/kg, LB71350 inhibited acetic acid-induced pain response approximately 41% and 83% of control. respectively. At oral doses of 200 mg/kg and 500 mg/kg, it reduced the rectal body temperature in rats. Pentylenetetrazole-induced seizure in mice was slightly potentiated by oral administration of LB71350 at doses ranging from 200 mg/kg to 1000 mg/Ag. Single or five day treatment of LB71350 doubled the hexobarbital- induced sleeping time in mice at oral doses ranging from 50 mg/kg to 500 mg/kg. It did not cause any effects on gastric secretion and acidity in rat at oral doses of 200 mg/kg and 1000 mg/kg and also it did not change intestinal motility in mice up to 1000 mg/kg. Blood coagulation indices such as prothrombin time (PT), activated partial thromboplastin time (aPTT), and thrombin time (TT) in rats were not affected by the treatment of LB71350 up to 500 mg/kg. LB71350 caused no significant effects on the cardiac output, stroke volume, heart rate, and mean blood pressure when infused intravenously to the anesthetized rats and dogs. Taken together, LB71350 at high oral doses caused significant pharmacological effects on the central nervous system and the hexobarbital-induced sleeping time.

  • PDF

The absorbed doses from each exposure program of the Orthopos$\textregistered$ panoramic machine (Orthopos$\textregistered$ 파노라마방사선촬영기의 각종 프로그램에 따른 흡수선량)

  • Choi Soon-Chul;Lee Sul-Mi
    • Imaging Science in Dentistry
    • /
    • v.31 no.4
    • /
    • pp.215-219
    • /
    • 2001
  • Purpose: The objective of this study was to estimate the radiation absorbed doses in certain critical organs in the head and neck region with 16 imaging programs available on the Orthopos/sup (R)/ panoramic machine. Materials and Methods: A Rando phantom and LiF TLD chips were used for dosimetry. The absorbed doses were measured at the thyroid gland, the submandibular gland, the parotid gland, the mouth floor. the maxillary sinus, the brain, the mandibular body, the mandibular ramus. the 2nd cervical spine and the skin over TMJ area. Results : The overall absorbed doses with imaging programs available on the Orthopos/sup (R)/ panoramic machine were much less than that of standard program (program 1) except program 8, 11, and 16. Generally, the absorbed doses to the bone marrow of the mandibular ramus and the parotid gland were high, but the absorbed doses to the bone marrow in the mandibular body, brain, maxillary sinus, and, especially, the thyroid gland were very low. Conclusion : The modified imaging programs available on the Orthopos/sup (R)/ panoramic machine can be effectively used in aspect of radiation protection.

  • PDF

Absorbed and effective dose in direct and indirect digital panoramic radiography (직.간접디지털 파노라마 방사선촬영시 흡수선량과 유효선량)

  • Lee, Gun-Sun;Kim, Jin-Soo;Kim, Jae-Duk
    • Imaging Science in Dentistry
    • /
    • v.40 no.1
    • /
    • pp.9-14
    • /
    • 2010
  • Purpose : We evaluated the absorbed doses to the organs and calculated the effective doses when using the digital panoramic radiography. Materials and Methods : The absorbed dose averages in major organs of oral and maxillofacial region were measured using the Dental head phantom (CIRS Co., USA), $^nLi_2B_4O_7$ TLD chip and UD-716AGL dosimeter (Matsushita Electric Industrial Co., JPN) when performing indirect and direct digital panoramic radiography. Effective doses were calculated from correspond to ICRP 2007 recommendations for two panoramic radiography. Results : The absorbed dose average on indirect and direct digital panoramic radiography was highest in parotid glands as measured 1259.6 mGy and 680.7 mGy respectively. Absorbed dose average in another organs were high in order of esophagus, submandibular gland, tongue and thyroid gland on both types of digital panoramic radiography. The absorbed dose average was higher on indirect type than direct one (p<0.05). The effective dose was higher on indirect type than direct one as measured 13.28 mSv and 8.70 mSv respectively. Conclusion : The absorbed doses in salivary gland and oral mucosa were high. However, thyroid gland also demands the attention on radiography due to high tissue weighting factor in spite of the low absorbed dose.