• 제목/요약/키워드: low drift flux meter

검색결과 1건 처리시간 0.019초

펠티어 소자를 사용한 Low Drift Flux Meter의 기초연구 (A Basic Study on the Low Drift Flux Meter by Using a Peltier Device)

  • 김철한;허진;신광호;사공건
    • 한국전기전자재료학회논문지
    • /
    • 제14권11호
    • /
    • pp.912-916
    • /
    • 2001
  • Fluxmeter is a measuring instrument the magnetic flux intensity by means of an integration of the voltage induced to a search coil to unit time. It also is required to a precise integrator since the voltage induced to a search coil has a differential value of the flux ${\Phi}$ to unit time. In this study, a bias current which is a main problem of the integrator in a drift troublesome depending on the temperature of a FET is investigated. We have confirmed that the temperature dependence of both the bias current of a integrator using the FET and the reversal saturated current of the minor carrier in a P-N junction of a semiconductor were the same. The property of a commercial integrator goes rapidly down with increasing temperature. The bias current of a FET is increased twice as much with 10$^{\circ}C$ increment. As a result, the low drift integrator could be developed by setting the lower temperature up with a pottier device.

  • PDF