Dosimetric Characteristics of Multileaf Collimator-based Intensity-modulated Arc Therapy for Stereotactic Radiosurgery (방사선수술 시 다엽 콜리메이터를 기초로 한 IMAT의 선량분포)

  • Yun, Sang-Mo;Kim, Sung-Kyu
    • Progress in Medical Physics
    • /
    • v.18 no.2
    • /
    • pp.93-97
    • /
    • 2007
  • This study was designed to evaluate radiosurgery technique using multiple noncoplanar arc therapy with intensity modulated fine MLC shaped photon beam. The stereotactic radiosurgery was performed with 6-MV X-ray beams from a Clinac 21EX LINAC (Varian, Palo Alto, CA, USA) with a MLC-120, which features a full $40{\times}40cm$ field and is the first MLC for general use that offers 0.5 cm resolution for high precision treatment of small and irregular fields. We used a single isocenter and five gantry-couch combinations with a set of intensity modulated arc therapy. We investigated dosimetric characteristics of 2 cm sized spherical target volume with film (X-OMAT V2 film, Kodak Inc, Rochester NY, USA) dosimetry within $25{\times}25cm$ acrylic phantom. A simulated single isocentric treatment using inversely Planned 3D radiotherapy planning system demonstrated the ability to conform the dose distribution to an spherical target volume. The 80% dose level was adequate to encompass the target volume in frontal, sagittal, and transverse planes, and the region between the 40% and 80% isodose lines was $4.0{\sim}4.5mm$ and comparable to the dose distribution of the Boston Arcs. We expect that our radiosurgery technique could be a treatment option for irregular-shaped large intracranial target.

  • PDF

Quality assurance for computed-tomography simulator : Report of the AAPM Radiation Therapy Committee Task Group No.66 (Report of the AAPM Radiation Therapy Committee의 Task Group No.66에 의한 전산화 단층촬영 모의치료기의 정도 관리)

  • Lee, Yun-Seok
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.17 no.1
    • /
    • pp.41-43
    • /
    • 2005
  • Purpose : Wish to present degree management process that is efficient confirm radiation treatment exclusive use CT simulator's Q.A item that become Q.A and Differentiation of diagnosis area that present Report of the AAPM Task Group No.66 using Q.A tool that produce itself and secure safe and correct CT-simulation process and equip convenience. Method and material : Manufacture CT simulator's Q.A tool on source and confirm virtue between isocenter of wall laser system, patient table, CT scanner's imaging plane that present in Report of the AAPM Task Group No.66 by daily publication unit. Result : Confirmed measured value from Report of the AAPM Task Group No.66 to confirmation of presenting degree management item in wall laser's ${\pm}2mm$, table's ${\pm}2mm$, imaging plane's ${\pm}2mm$ tolerance extent. Conclusion : There is unconfirmed item from CT-simulation process for therapy to CT Q.A protocol of existent diagnosis area, premising suitable degree management of radiation treatment exclusive use CT-simulator equipment confirming presenting Q.A item in Report of the AAPM Task Group No.66 safe and correct CT-simulation process guarantee can

  • PDF

Dosimetry and Three Dimensional Planning for Stereotactic Radiosurgery with SIEMENS 6-MV LINAC (6-MV선형가속기를 이용한 입체방사선수술의 선량측정 및 3차원적 치료계획)

  • Choi Dong-Rak;Cho Byong Chul;Suh Tae-Suk;Chung Su Mi;Choi Il Bong;Shinn Kyung Sub
    • Radiation Oncology Journal
    • /
    • v.11 no.1
    • /
    • pp.175-181
    • /
    • 1993
  • Radiosurgery requires integral procedure where special devices and computer systems are needed for localization, dose planning and treatment. The aim of this work is to verify the overall mechanical accuracy of our LINAC and develop dose calculation algorithm for LINAC radiosurgery. The alignment of treatment machine and the performance testing of the entire system were extensively carried out and the basic data such as percent depth dose, off-axis ratio and output factor were measured. A three dimensional treatment planning system for stereotactic radiosurgery has been developed. We used an IBM personal computer with C programming language (IBM personal system/2, Model 80386, IBM Co., USA) for calculating the dose distribution. As a result, deviations at isocenter on gantry and table rotation for our treatment machine were acceptable since they were less than 2 mm. According to the phantom experiments, the focusing isocenter were successful by the error of less than 2 mm. Finally, the mechanical accuracy of our three dimensional planning system was confirmed by film dosimetry in sphere phantom.

  • PDF

A Comparison of the Independent Verification Methods for the Results of Leksell GammaPlan for Gamma Knife Predecessor with the Hemispherical Collimators (반구형 시준기를 가진 감마나이프에 대한 렉셀감마플랜 결과물의 독립적인 검증방법들의 비교)

  • Hur, Beong Ik
    • Journal of the Korean Society of Radiology
    • /
    • v.10 no.7
    • /
    • pp.521-529
    • /
    • 2016
  • Since Gamma Knife(R) radiosurgery(GKRS) is based on a single-fraction high dose treatment strategy, independent verification for the results of Leksell GammaPlan(R) (LGP) is an important procedure in assuring patient safety and minimizing the risk of treatment errors. Several verification methods have been developed and reported previously. Thus these methods were tested statistically and tried on Leksell Gamma Knife(LGK) target treatments through the embodiment of the previously proposed algorithms(PPA). The purpose of this study was to apply and evaluate the accuracy of verification methods for LGK target treatments using PPA. In the study 10 patients with intracranial lesion treated by GKRS were included. We compared the data from PPA and LGP in terms of maximum dose, arbitrary point dose, and treatment time at the isocenter locations. All data were analyzed by Paired t-test, which is statistical method used to compare two different measurement techniques. No statistical significance in maximal dose at 10 cases was observed between PPA and LGP. Differences in average maximal dose ranged from -0.53 Gy to 3.71 Gy. The arbitrary point dose calculated by PPA and LGP was not statistically significant too. But we found out the statistical difference with p=0.021 between TMR and LGP for treatment time at the isocenter locations. PPA can be incorporated as part of a routine quality assurance(QA) procedure to minimize the chance of a wrong overdose. Statistical analyses demonstrated that PPA was in excellent agreement with LGP when considering the maximal dose and the arbitrary point dose for the best plan of GKRS. Due to the easy applicability we hope PPA can be widely used.

Verification of Target Position in Stereotactic Radiosurgery Based on Photon Knife System (Photon Knife 시스템에 근거한 뇌정위 방사선수술에서 표적위치 확인)

  • 최태진;김진희;김옥배
    • Progress in Medical Physics
    • /
    • v.14 no.2
    • /
    • pp.99-107
    • /
    • 2003
  • This study was performed to prepare the verification film for localizing beam-target position with the Photon Knife radiosurgery system (PKRS) using linear accelerator(Mitsubishi, Model ML-15MDX). We developed a laser calibration system using a reticle of transparent lucite to detect Inlet and outlet beams. We verified fixation of the second collimator with film mounted on a holder in the shape of an octagon block 5cm apart from the isocenter. The film was exposed to photon beams of linear accelerator at an interval of 45 degrees during the gantry movement. There were no shifts in the beam of the second collimator during gantry movement. We used a position marker which is designed a head-shaped small lead block and a 10 mm in diameter of steel bead in the plastic tube. The position marker helped to verify the beam directions with patient position in multi-arc and trans-multi-arc of PKRS The verification of beam alignments showed an average 0.8$\pm$0.26 mm discrepancy in LINAC-gram images of PKRS. In our study, the couch movement was $\pm$5 mm laterally, while it shook $\pm$ 2 mm toward the couch axis. The couch, however, was immediately returned to the initial site after shaking. Thus, we postulate that the beam-target position(s) should be verified with LINAC-gram in a stereotactic radiosurgery system to achieve the accuracy of beam-target alignment.

  • PDF

Development of A Fractionated Stereotactic Radiotherapy System (분할 정위방사선 치료 시스템 개발 연구)

  • 이동한;지영훈;이동훈;조철구;김미숙;유형준;류성렬
    • Progress in Medical Physics
    • /
    • v.13 no.1
    • /
    • pp.9-14
    • /
    • 2002
  • We invented the newly developed Fractionated Stereotactic Radiotherapy(F.S.R.T) system using combined techniques of couch mounting and pedestal mounting system. Head fixation frame consists of a milled alluminium alloy(duralumin) and is placed to the couch. This frame immobilized patient head using the dental bite, 3.2 mm frontal and occipital thermoplastic mask. To evaluate the coordinate of target isocenter, Brown-Revert-Walls C.T localizer can be attached to this frame. And also, we developed the frame mounting system by developing the modification of pedestal mounting system. This system is fixed to couch floor and can be used to evaluate the isocenteric accuracy of gantry, couch and collimator in Q.A procedure. In order to measure the relocation accuracy, the acrylic phantom and the accurate pointers have been made. The repositioning of the targets in the phantom were estimated by comparing C.T coordinates and E.C.L portal films taken with anterior-posterior and right-left direction. From the results of experiments, the average distance errors between the target isocenter and its mean position were 0.71$\pm$0.19 for lateral, 0.45$\pm$0.15 for inferior-superior, 0.63$\pm$0.18 for anterior-posterior. And the maximum distance error was less than 1.3 mm. The new head fixation frame and frame mounting system were non-invasive, accurately relocatable, easy to use, very light and well tolerable by the results of phantom tests. The major advantage of using this frame mounting system is complete access to any point in the Patients cranium especially posterior direction

  • PDF

The convergence study on patient position and exposure dose in abdominal CT examination using AEC (AEC를 적용한 복부 CT 검사 시 환자 자세와 피폭선량에 대한 융합 연구)

  • Lee, Chun-Kyu;Oh, Jeong-Sub;Choi, Seon-Wook;Kim, Gab-Jung;Yoo, Se-Jong;Jeon, Min-Cheol
    • Journal of the Korea Convergence Society
    • /
    • v.9 no.12
    • /
    • pp.107-113
    • /
    • 2018
  • The purpose of this study was to evaluate the dose and image quality according to the rotation of the X-axis direction in the abdominal CT scan, and to find ways to reduce the exposure dose. The phantom was scanned by rotating in the X-axis direction at 0, 5, 10, and 15 degrees, respectively. The CTDIvol value, HU, noise, and signal-to-noise ratio were measured at each rotation. ANOVA analysis was performed using the SPSSWIN (ver 19.0) program. The radiation exposure dose was 5.44mGy, 5.70mGy, 5.98mGy and 6.38mGy at 0, 5, 10 and 15 degrees, respectively. HU, noise, and signal-to-noise ratio were not statistically significant. In the CT scan, if the patient is located in the isocenter of the gantry aperture and there is no rotation in the X-axis direction, the exposure dose is reduced.

The Study on Applicability of Semi-conductive Compound for Radioactive Source Tracing Dosimeter in NDT Field (비파괴 검사 분야의 방사성 동위원소 위치추적을 위한 반도체 화합물의 적용 가능성 연구)

  • Shin, Yohan;Han, Moojae;Jung, Jaehoon;Kim, Kyotae;Heo, Yeji;Lee, Deukhee;Cho, Heunglae;Park, Sungkwang
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.1
    • /
    • pp.39-44
    • /
    • 2019
  • Radiation safety management is being considered very important since radioactive isotopes such as Co-60 and Ir-192 are widely used in fields such as non-destructive test(NDT). In this study, the applicability of Mercury(II) Iodide($HgI_2$) source for tracing system was evaluated. To make sure the unit cell sensor's reliability, we evaluated the electrical properties of the sensor made with $HgI_2$, and then position dependence of the sensor was analyzed and compared with the dose distribution from the planning system. As a result of the evaluation, high reliability of the sensor was shown through the linearity of R-sq > 0.990 and reproducibility of CV < 0.015. In the position dependence evaluation, the maximum value was measured at the isocenter of the sensor and gradually decreased according to the distance. However, the dose distribution data from the planning system was turned out that has difference with that of the sensor up to 30%. This seems to come from the difference between single-point measuring based planning system and area measuring based sensor.

New Techniques for Optimal Treatment Planning for LINAC-based Stereotactic Radiosurgery (LINAC 뇌정의적 방사선 수술시 새로운 최적 선량분포계획 시스템의 개발)

  • Suh Tae-suk
    • Radiation Oncology Journal
    • /
    • v.10 no.1
    • /
    • pp.95-100
    • /
    • 1992
  • Since LINAC-based stereotactic radiosurgery uses multiple noncoplanar arcs, three-dimensional dose evaluation and many beam parameters, a lengthy computation time is required to optimize even the simplest case by a trial and error. The basic approach presented in this paper is to show promising methods using an experimental optimization and an analytic optimization The purpose of this paper is not to describe the detailed methods, but introduce briefly, proceeding research done currently or in near future. A more detailed description will be shown in ongoing published papers. Experimental optimization is based on two approaches. One is shaping the target volumes through the use of multiple isocenters determined from dose experience and testing. The other method is conformal therapy using a beam's eye view technique and field shaping. The analytic approach is to adapt computer-aided design optimization in finding optimum irradiation parameters automatically.

  • PDF

치료중 실시간 모니터링을 위한 투과형 빔측정장치 개발

  • Kim, Jae-Hong;Swanepoel, M.W.;Dekock, E.A.;Park, Yeon-Su;Yang, Tae-Geon
    • Proceedings of the Korean Vacuum Society Conference
    • /
    • 2010.08a
    • /
    • pp.315-315
    • /
    • 2010
  • 양성자 빔을 이용하여 두경부 암 치료를 South Africa의 iTHEMBA에서 시행하고 있다. 200 MeV의 양성자 빔라인으로부터 진공에서 대기로 인출하여 노즐을 통과하여 종양세포에 조사된다. 치료계획에 적합하게 빔에너지와 모양을 변환하고, 빔을 모니터링하는 기계적 장치들이 노즐에 구성된다. 빔라인에는 이온챔버, Steering Magnet, Multi-wire 이온챔버, Range trimmer plates, lead scattering plate, Double-wedge energy degrader, Multi-layer Faraday cup, Range modulator, Range monitor, occluding ring, Shielding collimators, Quadrant and monitor ionization chamber, Treatment collimator, 그리고 Wellhofer dosimetry tank로 구성되어 있다. 총길이는 6.6m이며 노즐 끝에서 환자의 isocenter 까지는 30cm 정도 아래에 위치한다. 상기의 배치를 갖는 시스템의 양성자 scattering system의 성능을 MCNPX v2.5.0 Monte Carlo simulation을 실시하였다. 또한 정확한 선량을 실시간으로 측정하는 방법인 투과형 검출기를 개발하여 치료와 빔 특성을 동시에 수행하는 기술개발연구가 보고되고 있다. 본 연구에서는 Multileaf Faraday Cup (MLPC) 검출기 설계구조와 데이터 측정방법에 관한 연구를 수행하고자 한다. 빔의 전송 방향으로 3개층의 $4{\times}4$ 배열의 구조로 48 channel의 전류값을 측정하여 입자빔의 분포를 실시간으로 관측하고, 측정된 전류는 ADC를 거쳐 치료계획에 의해 선택된 영역의 SOBP를 유지하도록 range modulation propeller를 조절하는 feed-back system을 갖춘 방사선치료빔 실시간 측정장치 개발에 관한 결과를 보고하고자 한다.

  • PDF