Evaluation of the Space Scattered Dose According to the Position of the Radiation Workers in Mammography Room (유방촬영 시 방사선관계종사자의 위치에 따른 공간선량평가)

  • Lee, Dong-Yeon;Lee, Jin-Soo
    • Journal of radiological science and technology
    • /
    • v.39 no.3
    • /
    • pp.297-303
    • /
    • 2016
  • This study was conducted to evaluate the dose of the space to the controller located within the mammography room conducted a research on ways to the reduction exposure to the radiation workers. Results, the dose of 6.18 mGy/year was measured when there is no difference in the hilar area of the controller position, the dose of 2.35E-11 mGy/year was measured when installing the Shielding door. In addition, when the direction of the X-ray tube anode be heading this direction controller, low average level measured was 0.30 mGy/year. Based on this study, the mammography should be considered when installing the anode and cathod directions. And, by installing the shielding door, it must be able to completely separate shooting space and control room. This is the best way radiation protection method in radiation workers.

MOSFET Dosimetry for Evaluation of Gonad Shielding during Radiotherapy (방사선 치료시 생식선 차폐체 성능 평가를 위한 MOSFET 선량 측정)

  • Kim, Hwi-Young;Choi, Yun-Seok;Park, So-Yeon;Park, Yang-Kyun;Ye, Sung-Joon
    • Journal of Radiation Protection and Research
    • /
    • v.36 no.1
    • /
    • pp.23-27
    • /
    • 2011
  • In order to confirm feasibility of MOSFET modality in use of in.vivo dosimetry, evaluation of gonad shielding in order to minimize gonadal dose of patients undergoing radiotherapy by using MOSFET modality was performed. Gonadal dose of patients undergoing radiotherapy for rectal cancer in the department of radiation oncology of Seoul National University Hospital since 2009 was measured. 6 MV and 15 MV photon beams emitted from Varian 21EX LINAC were used for radiotherapy. In order to minimize exposed dose caused by the scattered ray not only from collimator of LINAC but also from treatment region inside radiation field, we used box.shaped lead shielding material. The shielding material was made of the lead block and consists of $7.5\; cm\;{\times}\;9.5\;cm\;{\times}5.5\;cm$ sized case and $9\;cm\;{\times}\;9.5\;cm\;{\times}\;1\;cm$ sized cover. Dosimetry for evaluation of gonad shielding was done with MOSFET modality. By protecting with gonad shielding material, average gonadal dose of patients was decreased by 23.07% compared with reference dose outside of the shielding material. Average delivered gonadal dose inside the shielding material was 0.01 Gy. By the result of MOSFET dosimetry, we verified that gonadal dose was decreased by using gonad shielding material. In compare with TLD dosimetry, we could measure the exposed dose easily and precisely with MOSFET modality.

Real Time Image Acquisition System using a Image Intensifier and Position Error Verification (영상증배관을 이용한 실시간 영상획득시스템과 위치오차검증)

  • Lee, Dong-Hoon;Kim, Nam-Hoon;Jeong, Jong-Beom
    • Journal of rehabilitation welfare engineering & assistive technology
    • /
    • v.11 no.4
    • /
    • pp.331-338
    • /
    • 2017
  • In this study, a portable x-ray generator was manufactured and a real-time image acquisition system was constructed using the image intensifier from the generated generator. We have developed a real - time position error verification system that can verify whether the artificial joint position is different from the initial image from the acquired image. The template image of the region of interest is extracted from the reference image using the pattern matching technique and compared with the image to be compared. As a result, It is shown that real - time position error verification is achieved by displaying the difference angle. This system is portable type, has a self-shielding facility, and the output of the irradiation device can be manufactured in a small size of 1kw and can be used as a portable type. In case of emergency patients in the non-destructive field for industrial use, It has proved effective for use in small areas such as feet.

Radiation safety for pain physicians: principles and recommendations

  • Park, Sewon;Kim, Minjung;Kim, Jae Hun
    • The Korean Journal of Pain
    • /
    • v.35 no.2
    • /
    • pp.129-139
    • /
    • 2022
  • C-arm fluoroscopy is a useful tool for interventional pain management. However, with the increasing use of C-arm fluoroscopy, the risk of accumulated radiation exposure is a significant concern for pain physicians. Therefore, efforts are needed to reduce radiation exposure. There are three types of radiation exposure sources: (1) the primary X-ray beam, (2) scattered radiation, and (3) leakage from the X-ray tube. The major radiation exposure risk for most medical staff members is scattered radiation, the amount of which is affected by many factors. Pain physicians can reduce their radiation exposure by use of several effective methods, which utilize the following main principles: reducing the exposure time, increasing the distance from the radiation source, and radiation shielding. Some methods reduce not only the pain physician's but also the patient's radiation exposure. Taking images with collimation and minimal use of magnification are ways to reduce the intensity of the primary X-ray beam and the amount of scattered radiation. It is also important to carefully select the C-arm fluoroscopy mode, such as pulsed mode or low-dose mode, for ensuring the physician's and patient's radiation safety. Pain physicians should practice these principles and also be aware of the annual permissible radiation dose as well as checking their radiation exposure. This article aimed to review the literature on radiation safety in relation to C-arm fluoroscopy and provide recommendations to pain physicians during C-arm fluoroscopy-guided interventional pain management.

Development of Lead Free Shielding Material for Diagnostic Radiation Beams (의료영상용 방사선방호를 위한 무납차폐체 개발)

  • Choi, Tae-Jin;Oh, Young-Kee;Kim, Jin-Hee;Kim, Ok-Bae
    • Progress in Medical Physics
    • /
    • v.21 no.2
    • /
    • pp.232-237
    • /
    • 2010
  • The shielding materials designed for replacement of lead equivalent materials for lighter apron than that of lead in diagnostic photon beams. The absorption characteristics of elements were applied to investigate the lead free material for design the shielding materials through the 50 kVp to 110 kVp x-ray energy in interval of 20 kVp respectively. The idea focused to the effect of K-edge absorption of variable elements excluding the lead material for weight reduction. The designed shielding materials composited of Tin 34.1%, Antimon 33.8% and Iodine 26.8% and Polyisoprene 5.3% gram weight account for 84 percent of weight of lead equivalent of 0.5 mm thickness. The size of lead-free shielder was $200{\times}200{\times}1.5\;mm^3$ and $3.2\;g/cm^3$ of density which is equivalent to 0.42 mm of Pb. The lead equivalent of 0.5 mm thickness generally used for shielding apron of diagnostic X rays which is transmitted 0.1% for 50 kVp, 0.9% for 70 kVp and 3.2% for 90 kVp and 4.8% for 110 kVp in experimental measurements. The experiment of transmittance for lead-free shielder has showed 0.3% for 50 kVp, 0.6% for 70 kVp, 2.0% for 90 kVp and 4.2% for 110 kVp within ${\pm}0.1%$. respectively. Using the attenuation coefficient of experiments for 0.5 mm Pb equivalent of lead-free materials showed 0.1%. 0.3%, 1.0% and 2.4%, respectively. Furthermore, the transmittance of lead-free shielder for scatter rays has showed the 2.4% in operation energy of 50 kVp and 5.9% in energy of 110 kVp against 2.4% and 5.1% for standard lead thickness within ${\pm}0.2%$ discrepancy, respectively. In this experiment shows the designed lead-free shielder is very effective for reduction the apron weight in diagnostic radiation fields.

The evaluation of contralateral breast's dose and shielding efficiency by breast size about breast implant patient for radiation therapy (인공 유방 확대술을 받은 환자의 유방암 치료 시 크기에 따른 반대 측 유방의 피폭 선량 및 차폐 효율 평가)

  • Kim, Jong Wook;Woo, Heon;Jeong, Hyeon Hak;Kim, Kyeong Ah;Kim, Chan Yong;Yoo, Suk Hyun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.26 no.2
    • /
    • pp.329-336
    • /
    • 2014
  • Purpose : To evaluate the dose on a contralateral breast and the usefulness of shielding according to the distance between the contralateral breast and the side of the beam by breast size when patients who got breast implant receive radiation therapy. Materials and Methods : We equipped 200 cc, 300 cc, 400 cc, and 500 cc breast model on the human phantom (Rando-phantom), acquired CT images (philips 16channel, Netherlands) and established the radiation treatment plan, 180 cGy per day on the left breast (EclipseTM ver10.0.42, Varian Medical Systems, USA) by size. We set up each points, A, B, C, and D on the right(contralateral) breast model for measurement by size and by the distance from the beam and attached MOSFET at each points. The 6 MV, 10 MV and 15 MV X-ray were irradiated to the left(target) breast model and we measured exposure dose of contralateral breast model using MOSFET. Also, at the same condition, we acquired the dose value after shielding using only Pb 2 mm and bolus 3 mm under the Pb 2 mm together. Results : As the breast model is bigger from 200 cc to 500 cc, The surface of the contralateral breast is closer to the beam. As a result, from 200 cc to 500 cc, on 180 cGy basis, the measurement value of the scattered ray inclined by 3.22 ~ 4.17% at A point, 4.06 ~ 6.22% at B point, 0.4~0.5% at C point, and was under 0.4% at D point. As the X-ray energy is higher, from 6 MV to 15 MV, on 180 cGy basis, the measurement value of the scattered ray inclined by 4.06~5% at A point, 2.85~4.94% at B point, 0.74~1.65% at C point, and was under 0.4% at D point. As using Pb 2 mm for shield, scattered ray declined by average 9.74% at A and B point, 2.8% at C point, and is under 1% at D point. As using Pb 2 mm and bolus together for shield, scattered ray declined by average 9.76% at A and B point, 2.2% at C point, and is under 1% at D point. Conclusion : Commonly, in case of patients who got breast implant, there is a distance difference by breast size between the contralateral breast and the side of beam. As the distance is closer to the beam, the scattered ray inclined. At the same size of the breast, as the X-ray energy is higher, the exposure dose by scattered ray tends to incline. As a result, as low as possible energy wihtin the plan dose is good for reducing the exposure dose.

Dose Assessment during Pregnancy in Chest PA Examination (흉부 후전방향 검사 시 임산부의 선량 평가)

  • Woo, Ri-Won;Cho, Yong-In;Kim, Jung-Hoon
    • Journal of the Korean Society of Radiology
    • /
    • v.14 no.5
    • /
    • pp.661-668
    • /
    • 2020
  • One of the causes of death for pregnant female is embolism, when a chest PA examination is performed. In addition, due to small doses, the examinations are performed for the purpose of preparing for pre-delivery emergency surgery or basic examination for pregnant female. Evaluating fetal doses through actual measurements is subject to ethical problems, Monte Carlo simulations assesses the organ and fetal doses of pregnant females according to week of pregnancy. The results of the simulations showed that the fetal dose decreased according to weeks of pregnancy and it showed a dose of about 0.1 mGy. The higher the density and thickness of the shielding material, the better the shielding effect. In addition, the dose reduction effect for each shielding material is between 40 and 98%. Afterwards, it is deemed necessary to study the reduction of fetal doses through various shielding characteristics and methods.

Particle loading as a design parameter for composite radiation shielding

  • Baumann, N.;Diaz, K. Marquez;Simmons-Potter, K.;Potter, B.G. Jr.;Bucay, J.
    • Nuclear Engineering and Technology
    • /
    • v.54 no.10
    • /
    • pp.3855-3863
    • /
    • 2022
  • An evaluation of the radiation shielding performance of high-Z-particle-loaded polylactic acid (PLA) composite materials was pursued. Specimens were produced via fused deposition modeling (FDM) using copper-PLA, steel-PLA, and BaSO4-PLA composite filaments containing 82.7, 75.2, and 44.6 wt% particulate phase contents, respectively, and were tested under broad-band flash x-ray conditions at the Sandia National Laboratories HERMES III facility. The experimental results for the mass attenuation coefficients of the composites were found to be in good agreement with GEANT4 simulations carried out using the same exposure conditions and an atomistic mixture as a model for the composite materials. Further simulation studies, focusing on the Cu-PLA composite system, were used to explore a shield design parameter space (in this case, defined by Cu-particle loading and shield areal density) to assess performance under both high-energy photon and electron fluxes over an incident energy range of 0.5-15 MeV. Based on these results, a method is proposed that can assist in the visualization and isolation of shield parameter coordinate sets that optimize performance under targeted radiation characteristics (type, energy). For electron flux shielding, an empirical relationship was found between areal density (AD), electron energy (E), composition and performance. In cases where ${\frac{E}{AD}}{\geq}2MeV{\bullet}cm{\bullet}g^{-1}$, a shield composed of >85 wt% Cu results in optimal performance. In contrast, a shield composed of <10 wt% Cu is anticipated to perform best against electron irradiation when ${\frac{E}{AD}}<2MeV{\bullet}cm{\bullet}g^{-1}$.

Evaluation of Metal Composite Filaments for 3D Printing (3D 프린팅용 금속 입자 필라멘트의 물성 및 차폐 능력 평가)

  • Park, Ki-Seok;Choi, Woo-Jeon;Kim, Dong-Hyun
    • Journal of the Korean Society of Radiology
    • /
    • v.15 no.5
    • /
    • pp.697-704
    • /
    • 2021
  • It is hard to get Filaments which are materials of the 3D printing Fused Deposition Modeling(FDM) method as radiation shielding in Korea. and also related research is insufficient. This study aims to provide basic data for the development of radiation shields using 3D printing by evaluating the physical properties and radiation shielding capabilities of filaments containing metal particles. after selecting five metal filaments containing metal particle reinforcement materials, the radiation shielding rate was calculated according to the Korean Industrial Standard's protective equipment test method to evaluate physical properties such as tensile strength, density, X-ray Diffraction(XRD), and weight measurement using ASTM's evaluation method. In the tensile strength evaluation, PLA + SS was the highest, ABS + W was the lowest, and ABS + W is 3.13 g/cm3 which value was the highest among the composite filaments in the density evaluation. As a result of the XRD, it may be confirmed that the XRD peak pattern of the particles on the surface of the specimen coincides with the pattern of each particle reinforcing material powder metal, and thus it was confirmed that the printed specimen contained powder metal. The shielding effect for each 3D printed composite filament was found to have a high shielding rate in proportion to the effective atomic number and density in the order of ABS + W, ABS + Bi, PLA+SS, PLA + Cu, and PLA + Al. In this study, it was confirmed that the metal particle composite filament containing metal powder as a reinforcing material has radiation shielding ability, and the possibility of using a radiation shielding filament in the future.

Evaluation of effective dose in panorama, cone beam CT and the usefulness of x-ray protective (치과방사선검사에서 방사선방어용구 사용 전, 후의 유효선량에 대한 평가)

  • Kim, Jae In;Choi, Won Keun;Lee, So La;Lee, Jung Hwa;Lee, Kwan Sub
    • Korean Journal of Digital Imaging in Medicine
    • /
    • v.14 no.2
    • /
    • pp.15-22
    • /
    • 2012
  • The purpose of this study was to measure the absorbed dose and calculate the effective dose for cone beam computed tomography (CBCT) and panorama units and to estimate usefulness of x-ray protective. Rando phantom and glass dosimeters were used for dosimetry. The absorbed doses were measured at 15 organs and 14 remainder from correspond to ICRP 2007 recommendations. The absorbed dose was highest in salivary glands as measured CBCT 2.420mGy, panorama 0.307mGy. Absorbed dose in another organs were high in order of thyroid, brain, skin, esophagus. The effective dose was CBCT 0.100mSv, panorama 0.011mSv and effective dose of panorama was higher than that of CBCT by 10 times. In case of wearing x-ray protective, reducing effective dose of CBCT by 0.066mSv (66%) and panorama by 0.008mSv (72%). Effective dose were reduced by radiological shielding but it needs further optimization studies, where dosimetric data are analyzed in combination with image quality with keep the patients' exposure as low as possible.

  • PDF