• Title/Summary/Keyword: Quality aduit

Search Result 1, Processing Time 0.017 seconds

KFDA TLD Dose Quality Audit and Measurement Uncertainty (식품의약품안전청의 치료방사선 선량보증과 측정불확도)

  • Jeong, Hee-Kyo;Lee, Hyun-Ku;Kim, Gwe-Ya;Yang, Hyun-Kyu;Lim, Chun-Il
    • Proceedings of the Korean Society of Medical Physics Conference
    • /
    • 2004.11a
    • /
    • pp.153-156
    • /
    • 2004
  • Korea Food and Drug Administration(KFDA) has peformed the calibration of therapy level dosimeters for Co-60 radiation since 1979. The reference standard ionization chamber has been calibrated at BIPM in France. The uncertainty on the KFDA calibration coefficients is 0.9 %(k=2) for air kerma and absorbed dose to water. Since 1999 a national quality audit program for ensuring dosimetry accuracy in Korea radiotherapy centers has been performed by the KFDA. The uncertainty associated with the determination of the absorbed dose to water from the TLD readings for high energy x-ray is 1.6 %(k=1). The correction factors for energy, non-linearity dose response, and TLD holder are used in the dose determination. Agreement between the user stated dose and KFDA measured dose within ${\pm}$ 5 % is considered acceptable. KFDA TLD postal dose quality audit program was peformed for 71 beam qualities of 53 domestic radiotherapy centers in 2003. The results for quality assurance showed that 63 out of 71 beam qualifies (89 %) satisfied the acceptance limit. The second audit was carried out for the centers outside the limit and ail of them have been corrected.

  • PDF