• Title/Summary/Keyword: Magnetic Susceptibility Artifact

Search Result 28, Processing Time 0.023 seconds

Evaluation of the Usefulness of PROPELLER (periodically rotated overlapping parallel lines with enhanced reconstruction) Technique to Reduce the Magnetic susceptibility artifact (Magnetic susceptibility artifact를 줄이기 위한 PROPELLER 확산강조영상기법의 유용성에 대한 평가)

  • Cho, Jae-Hwan
    • Journal of Digital Contents Society
    • /
    • v.11 no.1
    • /
    • pp.73-78
    • /
    • 2010
  • This study attempted to examine whether the propeller diffusion weighted image method may remove magnetic susceptibility artifacts caused by metallic materials. A comparison of occurrence rates of magnetic susceptibility artifacts in the four regions, both temporal lobes, pons, and orbit, between b = 0 and b = 1,000 s/mm2 images was made after obtaining echo-planar diffusion weighted image, propeller diffusion weighted image, and ADC map images, respectively, from a total of 20 patients who had MRI shots taken of their brain and were found to be with retained metallic foreign bodies within their teeth using a 3.0T MR scanner. In the case of echo-planar diffusion weighted image technique, the presence of metallic materials may bring in some limits on accurate diagnosis due to magnetic susceptibility artifacts, while the propeller diffusion weighted image technique where magnetic susceptibility artifacts decrease is expected to be more useful in ensuring accurate diagnosis in the clinical context.

MRI Artifacts

  • 최순섭
    • Investigative Magnetic Resonance Imaging
    • /
    • v.1 no.1
    • /
    • pp.51-57
    • /
    • 1997
  • MRI의 artifact는 대부분 신호의 부호화 방향에 따라서 방향성을 가지는데, 이를 요약해보면, 위상부호화 방향의 artifact에는 motion artifact, flow artifact, RF noise등이 있고, 주파수 부호화 방향의 artfact는 susceptibility artfact, chemical shift artifact, central line artifact등이 있으며, 양방향 모두 생길수 있는 것은 Aliasing artifact와 Gibb's phenomenon이고, 전체적으로 영샹의 질을 떨어뜨리는 것은 susceptibility artifact, Eddy current, cross talk등이 있다. 이런 artifact는 대부분은 MRI 자체의 물리적 특성에 다소간 기인하므로, artifact가 없는 양호한 영상을 얻기 위해서는 MRI의 설치 단계부터 관심이 필요하고, MRI의 기본원리와 다양한 artifact에 대해 이해함으로써, 제거 가능한 artifact는 제거하여 양질의 영상을 만들고 판독시의 오류를 피할 수 있도록 해야할 것이다.

  • PDF

A Study on the Artifact Reduction Method of Magnetic Resonance Imaging in Dental Implants and Prostheses (치아 임플란트와 보철에서 발생하는 자기공명영상의 인공물 감소방안 연구)

  • Shin, Woon-Jae
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.7
    • /
    • pp.1025-1033
    • /
    • 2019
  • Although magnetic resonance imaging without linear hardening of CT is recognized as a method of obtaining high contrast of tissue and excellent resolution image in brain disease and head and neck examination, magnetic susceptibility artifact is generated in case of metal implants in the oral cavity, which is an obstacle to image diagnosis. Therefore, an effort was made in this thesis to find a method to reduce artifacts caused by dental implants and prosthesis in MRI. Implant-induced artifacts in magnetic resonance imaging showed that the signal size increased with shorter TE in GE technique and was inconsistent with water temperature change. In SE technique as well, the signal size of water was generally higher than that of air, but the signal to noise ratio (SNR) was not different by air and temperature. In EPI technique, images with fewer artifacts were obtained quantitatively and qualitatively when there was more water than air, and the signal to noise ratio was measured the highest, especially at water temperatures of 20° and 30°. In conclusion, when examining using the EPI technique rather than the SE or the GE technique, obtaining brain diffusion using a 20° and 30° water bag reduces the magnetic susceptibility artifacts caused by implants and prosthesis, suggesting that it may provide images with high diagnostic value.

Evaluation of O-MAR XD Technique for Reduction of Magnetic Susceptibility Artifact of Knee Implant (인공 무릎관절에서 자화율 인공물의 감소를 위한 O-MAR XD 기법의 평가)

  • Lee, Jung-Hoon
    • Journal of radiological science and technology
    • /
    • v.41 no.5
    • /
    • pp.413-419
    • /
    • 2018
  • Magnetic Resonance Imaging for patients with metallic implant has poor image quality, and signal loss and artifacts including distortion can occur. The purpose of this study is to carry out a comparative evaluation on high receive bandwidth(hiBW), O-MAR, O-MAR XD to reduce artifacts in knee implant. To take MRI, 3.0T scanner and dual-source radiofrequency transmission were used. O-MAR XD technique's strong option showed a significant difference (p<0.001) with O-MAR XD technique's weak option, O-MAR and hiBW excluding the medium option. O-MAR XD's medium option had a significant difference (p<0.01) with O-MAR XD's weak, O-MAR and hiBW. O-MAR XD technique's weak option had a significant difference (p<0.01) with O-MAR XD's strong and medium options, O-MAR and hiBW. O-MAR technique had a significant difference (p<0.001) with strong, medium, weak options of O-MAR XD technique except for hiBW. HiBW had a significant difference (p<0.001) with strong, medium and weak options of O-MAR XD technique except for O-MAR. The results showed that O-MAR XD technique was more useful for MRI scan for patients with knee replacement surgery than traditional techniques such as hiBW or O-MAR, and susceptibility artifacts decreased more when O-MAR XD technique's strong or medium option was applied. Based on the results above, it is considered that it will be possible to acquire images whose susceptibility artifacts were highly decreased by using O-MAR XD technique's strong or medium option when conducting MRI for artificial knee joint and it will be helpful for checking and monitoring patients with knee joint replacement.

A Study on the Reduction of the Magnetic Susceptibility Artifact in MRI of the Cervical Spine with Chemical Shift Selective Fat Suppression (CHESS 기법을 이용한 지방소거 경추 자기공명영상에서 자화감수성 인공물을 줄이기 위한 연구)

  • Lim, Woo-Taek;Jeong, Hong-Ryang
    • Proceedings of the Korea Contents Association Conference
    • /
    • 2014.11a
    • /
    • pp.197-198
    • /
    • 2014
  • CHESS(chemical shift selective saturation) 기법을 이용한 경추 자기공명영상에서 주로 나타나는 자화 감수성 인공물(susceptibility artifact)을 줄이기 위한 보조물질을 찾고자 하였다.

  • PDF

Reduction of Metal Artifact by Using VAT-SEMAC in MRI (VAT-SEMAC을 이용한 보철물에 의한 허상 감소)

  • Kim, Hyung-Tae;Lim, Jong-Nam;Chon, Kwon Su
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.2
    • /
    • pp.227-232
    • /
    • 2019
  • MRI examination for patients with metal objects has in poor image quality. Metallic implants can result in poor image because magnetic susceptibility causes signal loss and distortion and makes poor imaging, which is called magnetic susceptibility artifact or metal artifact. There are several approaches to reduce metal artifacts. In this study, we study the reduction of metal artifact by VAT and SEMAC techniques. A metal implant used for orthopedic surgery was attached to the phatom and the distortion caused by the artifact was measured under T1WI and T2WI protocols. Several techniques of VAT only and VAT and SEMAC for the reduction of metal artifact were compared. The metal artifact showed a reduction of at least 8% to a maximum of 26% in the VAT-SEMAC. The VAT-SEMAC technique can be applied to patients with orthopedic implants to improve image quality. If scan time and image quality are simultaneously considered in VAT-SEMAC technique, metal artifact will be reduced in clinical practice.

Effect of Manganese Content on the Magnetic Susceptibility of Ferrous-Manganese Alloys: Correlation between Microstructure on X-Ray Diffraction and Size of the Low-Intensity Area on MRI

  • Youn, Sung Won;Kim, Moon Jung;Yi, Seounghoon;Ahn, Hyun Jin;Park, Kwan Kyu;Lee, Jongmin;Lee, Young-Cheol
    • Investigative Magnetic Resonance Imaging
    • /
    • v.19 no.2
    • /
    • pp.76-87
    • /
    • 2015
  • Purpose: There is an ongoing search for a stent material that produces a reduced susceptibility artifact. This study evaluated the effect of manganese (Mn) content on the MRI susceptibility artifact of ferrous-manganese (Fe-Mn) alloys, and investigated the correlation between MRI findings and measurements of Fe-Mn microstructure on X-ray diffraction (XRD). Materials and Methods: Fe-Mn binary alloys were prepared with Mn contents varying from 10% to 35% by weight (i.e., 10%, 15%, 20%, 25%, 30%, and 35%; designated as Fe-10Mn, Fe-15Mn, Fe-20Mn, Fe-25Mn, Fe-30Mn, and Fe-35Mn, respectively), and their microstructure was evaluated using XRD. Three-dimensional spoiled gradient echo sequences of cylindrical specimens were obtained in parallel and perpendicular to the static magnetic field (B0). In addition, T1-weighted spin echo, T2-weighted fast spin echo, and $T2^*$weighted gradient echo images were obtained. The size of the low-intensity area on MRI was measured for each of the Fe-Mn binary alloys prepared. Results: Three phases of ${\alpha}^{\prime}$-martensite, ${\gamma}$-austenite, and ${\varepsilon}$-martensite were seen on XRD, and their composition changed from ${\alpha}^{\prime}$-martensite to ${\gamma}$-austenite and/or ${\varepsilon}$-martensite, with increasing Mn content. The Fe-10Mn and Fe-15Mn specimens comprised ${\alpha}^{\prime}$-martensite, the Fe-20Mn and Fe-25Mn specimens comprised ${\gamma}+{\varepsilon}$ phases, and the Fe-30Mn and Fe-35Mn specimens exhibited a single ${\gamma}$ phase. The size of the low-intensity areas of Fe-Mn on MRI decreased relative to its microstructure on XRD with increasing Mn content. Conclusion: Based on these findings, proper conditioning of the Mn content in Fe-Mn alloys will improve its visibility on MR angiography, and a Mn content of more than 25% is recommended to reduce the magnetic susceptibility artifacts on MRI. A reduced artifact of Fe-Mn alloys on MRI is closely related to the paramagnetic constitution of ${\gamma}$-austenite and/or ${\varepsilon}$-martensite.

Reduction of Susceptibility Artifact in MR Gradient Echo Imaging (MR Gradient Echo 영상술에서 자화율에 의한 에러의 감소에 관한 연구)

  • Ro, Y.M.;Cho, Z.H.
    • Proceedings of the KOSOMBE Conference
    • /
    • v.1990 no.11
    • /
    • pp.19-23
    • /
    • 1990
  • A new technique for reducing the susceptibility artifact in gradient echo imaging which uses a tailored RF pulse is proposed. It is applied to the case of imaging where artifacts and distortions arise due to the high local magnetic field inhomogeneity i. e., the susceptibility. The signal loss and void phenomena due to susceptibility in a voxel are studied and a correction method using a tailored RF pulse is proposed. Applications of this method in imaging are given and experimental results obtained using an human volunteer with a 2.0 T KAIS NMR system are presented.

  • PDF

In Vitro Assessment of MRI Safety at 1.5 T and 3.0 T for Bone-Anchored Hearing Aid Implant (Bone-Anchored Hearing Aid Implant에 대한 1.5 T와 3.0 T에서 MRI 안전성의 생체외 평가)

  • Yeon, Kyoo-Jin;Kim, Hyun-Soo;Lee, Seung-keun;Lee, Tae-Soo
    • Journal of radiological science and technology
    • /
    • v.40 no.1
    • /
    • pp.19-25
    • /
    • 2017
  • The aim of this study was to evaluate Magnetic Resonance Imaging safety by measuring the translational attraction, torque and susceptibility artifact for Bone-Anchored Hearing Aid (BAHA) implant at 1.5 T and 3.0 T MRI by standard criteria. In vitro assessment tools were made of acrylic-resin by American Society for Testing and Materials (ASTM) F2052-06 and F2119-07 standard. Translational attraction of BAHA implant was measured by the maximum deflection angle at 96 cm position, where the magnetically induced deflection was the greatest. The torque was assessed by the qualitative criteria of evaluating the alignment and rotation pattern, when the BAHA implant was positioned on a line with $45^{\circ}$ intervals inside the circular container in the center of the bore. The susceptibility artifact images were obtained using the hanged test tool, which was filled with $CuSO_4$ solution. And then the artifact size was measured using Susceptibility A rtifact Measurement (SA M) software. In results, the translational attraction was 0 mm at both 1.5 T and 3.0 T and the torque was 0(no torque) at 1.5 T, and +1(mild torque) at 3.0 T. The size of susceptibility artifacts was between 13.20 mm and 38.91 mm. Therefore, The BAHA implant was safe for the patient in clinical MR environment.

Evaluation of Image Usability by SEMAC Turbo Factor Change using Susceptibility Artifact Reduction (Susceptibility Artifact를 감소시키는 SEMAC 사용 시 Turbo Factor 변화에 따른 영상의 유용성 평가)

  • Choi, Young-Jae;Kim, Sang-Hyun
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.1
    • /
    • pp.31-37
    • /
    • 2019
  • The study analyzes Non SEMAC and SEMAC to reduce susceptibility artifacts that may occur when performing magnetic resonance imaging(MRI) of metal patients. The Foot and Ankle Phantom was used as the experimental tool and the 3.8 cm general screw was used to make the magnetic susceptibility artifact. The experimental equipment was used 3.0T Magnetom Skyra and the area was measured with the 17th image where the signal off is the most noticeable in the obtained image. Statistical analysis was performed using the SPSS(Ver.25) program and the significance was assessed by the Wilcoxon Signed Rank Test. As a result, the area of Non SEMAC which is the lowest signal was $289.53{\pm}23.07197mm$. When the Turbo Factor was changed to 3, 4, and 5 after SEMAC use, it decreased to $125.02{\pm}7.45875mm$, $120.96{\pm}12.01704mm$ and $108.79{\pm}16.53498mm$, respectively. In conclusion, this study demonstrates that Using SEMAC with Turbo Factor effectively reduces the susceptibility artifacts.