Comparison of the Efficacy of 2D Dosimetry Systems in the Pre-treatment Verification of IMRT (세기조절방사선치료의 환자별 정도관리를 위한 2차원적 선량계의 유용성 평가)

  • Hong, Chae-Seon;Lim, Jong-Soo;Ju, Sang-Gyu;Shin, Eun-Hyuk;Han, Young-Yih;Ahn, Yong-Chan
    • Radiation Oncology Journal
    • /
    • v.27 no.2
    • /
    • pp.91-102
    • /
    • 2009
  • Purpose: To compare the accuracy and efficacy of EDR2 film, a 2D ionization chamber array (MatriXX) and an amorphous silicon electronic portal imaging device (EPID) in the pre-treatment QA of IMRT. Materials and Methods: Fluence patterns, shaped as a wedge with 10 steps (segments) by a multi-leaf collimator (MLC), of reference and test IMRT fields were measured using EDR2 film, the MatriXX, and EPID. Test fields were designed to simulate leaf positioning errors. The absolute dose at a point in each step of the reference fields was measured in a water phantom with an ionization chamber and was compared to the dose obtained with the use of EDR2 film, the MatriXX and EPID. For qualitative analysis, all measured fluence patterns of both reference and test fields were compared with calculated dose maps from a radiation treatment planning system (Pinnacle, Philips, USA) using profiles and $\gamma$ evaluation with 3%/3 mm and 2%/2 mm criteria. By measurement of the time to perform QA, we compared the workload of EDR2 film, the MatriXX and EPID. Results: The percent absolute dose difference between the measured and ionization chamber dose was within 1% for the EPID, 2% for the MatriXX and 3% for EDR2 film. The percentage of pixels with $\gamma$%>1 for the 3%/3 mm and 2%/2 mm criteria was within 2% for use of both EDR2 film and the EPID. However, differences for the use of the MatriXX were seen with a maximum difference as great as 5.94% with the 2%/2 mm criteria. For the test fields, EDR2 film and EPID could detect leaf-positioning errors on the order of -3 mm and -2 mm, respectively. However it was difficult to differentiate leaf-positioning errors with the MatriXX due to its poor resolution. The approximate time to perform QA was 110 minutes for the use of EDR2 film, 80 minutes for the use of the MatriXX and approximately 55 minutes for the use of the EPID. Conclusion: This study has evaluated the accuracy and efficacy of EDR2 film, the MatriXX and EPID in the pre-treatment verification of IMRT. EDR2 film and the EPID showed better performance for accuracy, while the use of the MatriXX significantly reduced measurement and analysis times. We propose practical and useful methods to establish an effective QA system in a clinical environment.

Analysis of the Dose Distribution of Moving Organ using a Moving Phantom System (구동팬텀 시스템에 의한 움직이는 장기의 선량분포 분석)

  • Kim, Yon-Lae;Park, Byung-Moon;Bae, Yong-Ki;Kang, Min-Young;Lee, Gui-Won;Bang, Dong-Wan
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.18 no.2
    • /
    • pp.81-87
    • /
    • 2006
  • Purpose: Few researches have been peformed on the dose distribution of the moving organ for radiotherapy so far. In order to simulate the organ motion caused by respiratory function, multipurpose phantom and moving device was used and dosimetric measurements for dose distribution of the moving organs were conducted in this study. The purpose of our study was to evaluate how dose distributions are changed due to respiratory motion. Materials and Methods: A multipurpose phantom and a moving device were developed for the measurement of the dose distribution of the moving organ due to respiratory function. Acryl chosen design of the phantom was considered the most obvious choice for phantom material. For construction of the phantom, we used acryl and cork with density of $1.14g/cm^3,\;0.32g/cm^3$ respectively. Acryl and cork slab in the phantom were used to simulate the normal organ and lung respectively. The moving phantom system was composed of moving device, moving control system, and acryl and cork phantom. Gafchromic film and EDR2 film were used to measure dose ditrbutions. The moving device system may be driven by two directional step motors and able to perform 2 dimensional movements (x, z axis), but only 1 dimensional movement(z axis) was used for this study. Results: Larger penumbra was shown in the cork phantom than in the acryl phantom. The dose profile and isodose curve of Gafchromic EBT film were not uniform since the film has small optical density responding to the dose. As the organ motion was increased, the blurrings in penumbra, flatness, and symmetry were increased. Most of measurements of dose distrbutions, Gafchromic EBT film has poor flatness and symmetry than EDR2 film, but both penumbra distributions were more or less comparable. Conclusion: The Gafchromic EBT film is more useful as it does not need development and more radiation dose could be exposed than EDR2 film without losing film characteristics. But as response of the optical density of Gafchromic EBT film to dose is low, beam profiles have more fluctuation at Gafchromic EBT. If the multipurpose phantom and moving device are used for treatment Q.A, and its corrections are made, treatment quality should be improved for the moving organs.

  • PDF

Review on Usefulness of EPID (Electronic Portal Imaging Device) (EPID (Electronic Portal Imaging Device)의 유용성에 관한 고찰)

  • Lee, Choong Won;Park, Do Keun;Choi, A Hyun;Ahn, Jong Ho;Song, Ki Weon
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.25 no.1
    • /
    • pp.57-67
    • /
    • 2013
  • Purpose: Replacing the film which used to be used for checking the set-up of the patient and dosimetry during radiation therapy, more and more EPID equipped devices are in use at present. Accordingly, this article tried to evaluated the accuracy of the position check-up and the usefulness of dosimetry during the use of an electronic portal imaging device. Materials and Methods: On 50 materials acquired with the search of Korea Society Radiotherapeutic Technology, The Korean Society for Radiation Oncology, and Pubmed using "EPID", "Portal dosimetry", "Portal image", "Dose verification", "Quality control", "Cine mode", "Quality - assurance", and "In vivo dosimetry" as indexes, the usefulness of EPID was analyzed by classifying them as history of EPID and dosimetry, set-up verification and characteristics of EPID. Results: EPID is developed from the first generation of Liquid-filled ionization chamber, through the second generation of Camera-based fluoroscopy, and to the third generation of Amorphous-silicon EPID imaging modes can be divided into EPID mode, Cine mode and Integrated mode. When evaluating absolute dose accuracy of films and EPID, it was found that EPID showed within 1% and EDR2 film showed within 3% errors. It was confirmed that EPID is better in error measurement accuracy than film. When gamma analyzing the dose distribution of the base exposure plane which was calculated from therapy planning system, and planes calculated by EDR2 film and EPID, both film and EPID showed less than 2% of pixels which exceeded 1 at gamma values (r%>1) with in the thresholds such as 3%/3 mm and 2%/2 mm respectively. For the time needed for full course QA in IMRT to compare loads, EDR2 film recorded approximately 110 minutes, and EPID recorded approximately 55 minutes. Conclusion: EPID could easily replace conventional complicated and troublesome film and ionization chamber which used to be used for dosimetry and set-up verification, and it was proved to be very efficient and accurate dosimetry device in quality assurance of IMRT (intensity modulated radiation therapy). As cine mode imaging using EPID allows locating tumors in real-time without additional dose in lung and liver which are mobile according to movements of diaphragm and in rectal cancer patients who have unstable position, it may help to implement the most optimal radiotherapy for patients.

  • PDF

Property of Dose Distribution in Accordance with Dose Rate Variation in Intensity Modulated Radiation Therapy (세기조절방사선치료에서 선량율 변화에 따른 선량분포 특성)

  • Kang, Min-Kyu;Kim, Sung-Joon;Shin, Hyun-Soo;Kim, Sung-Kyu
    • Progress in Medical Physics
    • /
    • v.21 no.2
    • /
    • pp.218-222
    • /
    • 2010
  • As radiation is irradiated from various directions in intensity modulated radiation therapy (IMRT), longer treatment time than conventional treatment method is taken. In case of the patients who have problem to keep same posture for long time because of pain and injury, reducing treatment time through increased dose rate is a way for effective treatment. This study measured and found out the variation of dose and dose distribution in accordance with dose rate variation. IMRT treatment plan was set up to investigate from 5 directions - $0^{\circ}$, $72^{\circ}$, $144^{\circ}$, $216^{\circ}$, $288^{\circ}$ - using ECLIPSE system (Varian, SomaVision 6.5, USA). To confirm dose and dose rate in accordance with dose rate variation, dose rate was set up as 100, 300, 500 MU/min, and dose and dose distribution were measured using ionization chamber (PTW, TN31014) and film dosimeter (EDR2, Kodak). At this time, film dosimeter was inserted into acrylic phantom, then installed to run parallel with beam's irradiating direction, 21EX-S (Varian, USA) was utilized as linear accelerator for irradiation. The measured film dosimeter was analyzed using VXR-16 (Vidar System Corporation) to confirm dose distribution.

The Comparative Analysis of Exposure Conditions between F/S and C/R System for an Ideal Image in Simple Abdomen (복부 단순촬영의 이상적 영상구현을 위한 F. S system과 C.R system의 촬영조건 비교분석)

  • Son, Sang-Hyuk;Song, Young-Geun;Kim, Je-Bong
    • Korean Journal of Digital Imaging in Medicine
    • /
    • v.9 no.1
    • /
    • pp.37-43
    • /
    • 2007
  • 1. Purpose : This study is to present effective exposure conditions to acquire the best image of simple abdomen in Film Screen (F.S) system and Computed Radiography (C.R) system. 2. Method : In the F.S system, while an exposure condition was fixed as 70kVp, images of a patients simple abdomen were taken under the different mAs exposure conditions. Among these images, the best one was chosen by radiologists and radiological technologists. In the C.R system, the best image of the same patient was acquired with the same method from the F.S system. Both characteristic curves from F.S system and C.R system were analyzed. 3. Results : In the F.S system, the best exposure condition of simple abdomen was 70kVp and 20mAs. In the CR system, with the fixed condition at 70kVp, the image densities of human organs, such as liver, kidney, spleen, psoas muscle, lumbar spine body and iliac crest, were almost same despite different environments (3.2mAs, 8mAs, 12mAs, 16mAs and 20mAs). However, when the exposure conditions were over or under (below) 12mAs, the images between the abdominal wall and the directly exposed part became blurred because the gap of density was decreased. In the C.R system, while the volume of mAs was decreased, an artifact of quantum mottle was increased. 4. Conclusion : This study shows that the exposure condition in the C.R system can be reduced 40% than in the F.S system. This paper concluded that when the exposure conditions are set in CR environment, after the analysis of equipment character, such as image processing system(EDR : Exposure Data Recognition processing), PACS and so on, the high quality of image with maximum information can be acquired with a minimum exposure dose.

  • PDF

Superficial Dosimetry for Helical Tomotherapy (토모테라피를 이용한 표면 치료 계획과 선량 분석)

  • Kim, Song-Yih;You, Sei-Hwan;Song, Tae-Soo;Kim, Yong-Nam;Keum, Ki-Chang;Cho, Jae-Ho;Lee, Chang-Geol;Seong, Jin-Sil
    • Radiation Oncology Journal
    • /
    • v.27 no.2
    • /
    • pp.103-110
    • /
    • 2009
  • Purpose: To investigate the feasibility of helical tomotherapy on a wide curved area of the skin, and its accuracy in calculating the absorbed dose in the superficial region. Materials and Methods: Two types of treatment plans were made with the cylinder-shaped 'cheese phantom'. In the first trial, 2 Gy was prescribed to a 1-cm depth from the surface. For the other trial, 2 Gy was prescribed to a 1-cm depth from the external side of the surface by 5 mm. The inner part of the phantom was completely blocked. To measure the surface dose and the depth dose profile, an EDR2 film was inserted into the phantom, while 6 TLD chips were attached to the surface. Results: The film indicated that the surface dose of the former case was 118.7 cGy and the latter case was 130.9 cGy. The TLD chips indicated that the surface dose was higher than these, but it was due to the finite thickness of the TLD chips. In the former case, 95% of the prescribed dose was obtained at a 2.1 mm depth, while the prescribed does was at 2.2 mm in the latter case. The maximum dose was about 110% of the prescribed dose. As the depth became deeper, the dose decreased rapidly. Accordingly, at a 2-cm depth, the dose was 20 % of the prescribed dose. Conclusion: Helical tomotherapy could be a useful application in the treatment of a wide area of the skin with curvature. However, for depths up to 2 mm, the planning system overestimated the superficial dose. For shallower targets, the use of a compensator such as a bolus is required.

Analysis of changes in dose distribution due to respiration during IMRT

  • Shin, Jung-Suk;Shin, Eun-Hyuk;Han, Young-Yih;Ju, Sang-Gyu;Kim, Jin-Sung;Ahn, Sung-Hwan;Kim, Tae-Gyu;Jeong, Bae-Kwon;Park, Hee-Chul;Ahn, Young-Chan;Choi, Doo-Ho
    • Radiation Oncology Journal
    • /
    • v.29 no.3
    • /
    • pp.206-213
    • /
    • 2011
  • Purpose: Intensity modulated radiation therapy (IMRT) is a high precision therapy technique that can achieve a conformal dose distribution on a given target. However, organ motion induced by respiration can result in significant dosimetric error. Therefore, this study explores the dosimetric error that result from various patterns of respiration. Materials and Methods: Experiments were designed to deliver a treatment plan made for a real patient to an in-house developed motion phantom. The motion pattern; the amplitude and period as well as inhale-exhale period, could be controlled by in-house developed software. Dose distribution was measured using EDR2 film and analysis was performed by RIT113 software. Three respiratory patterns were generated for the purpose of this study; first the 'even inhale-exhale pattern', second the slightly long exhale pattern (0.35 seconds longer than inhale period) named 'general signal pattern', and third a 'long exhale pattern' (0.7 seconds longer than inhale period). One dimensional dose profile comparisons and gamma index analysis on 2 dimensions were performed. Results: In one-dimensional dose profile comparisons, 5% in the target and 30% dose difference at the boundary were observed in the long exhale pattern. The center of high dose region in the profile was shifted 1 mm to inhale (caudal) direction for the 'even inhale-exhale pattern', 2 mm and 5 mm shifts to exhale (cranial) direction were observed for 'slightly long exhale pattern' and 'long exhale pattern', respectively. The areas of gamma index >1 were 11.88 %, 15.11%, and 24.33% for 'even inhale-exhale pattern', 'general pattern', and 'long exhale pattern', respectively. The long exhale pattern showed largest errors. Conclusion: To reduce the dosimetric error due to respiratory motions, controlling patient's breathing to be closer to even inhaleexhale period is helpful with minimizing the motion amplitude.

Partial transmission block production for real efficient method of block and MLC (Partial transmission block 제작 시 real block과 MLC를 이용한 방법 중 효율적인 방법에 대한 고찰)

  • Choi JiMin;Park JuYoung;Ju SangGyu;Ahn JongHo
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.16 no.2
    • /
    • pp.19-24
    • /
    • 2004
  • Introduction : The Vaginal, the urethra, the vulva and anal cancer avoid the many dose to femur head and the additional treatment is necessary in inguinal LN. The partial transmission block to use inguinal LN addition there is to a method which it treats and produce partial transmission block a method and the MLC which to it analyzes. Material & Methode : The Inguinal the LN treatment patient partial transmission it used block and the MLC in the object and with solid water phantom with the patient it reappeared the same depth. In order to analyze the error of the junction the EDR2 (Extended dose range, the Kodak and the U.S) it used the Film and it got film scanner it got the beam profile. The partial transmission block and the MLC bias characteristic, accuracy and stability of production for, it shared at hour and comparison it analyzed. Result : The partial the transmission block compares in the MLC and the block production is difficult and production hour also above 1 hours. The custom the block the place where it revises the error of the junction is a difficult problem. If use of the MLC the fabrication will be break and only the periodical calibration of the MLC it will do and it will be able to use easily. Conclusion : The Inguinal there is to LN treatment and partial transmission block and the MLC there is efficiency of each one but there is a place where the junction of block for partial transmission block the production hour is caught long and it fixes and a point where the control of the block is difficult. like this problem it transfers with the MLC and if it treats, it means the effective treatment will be possible.

  • PDF

The Analysis of a Cerrobend Compensator and a Electronic Compensator Designed by a Radiation Treatment Planning System (방사선치료계획장치로 설계된 Cerrobend 선량보상체와 전자 선량보상체의 제작 및 특성 분석)

  • Nah Byung-Sik;Chung Woong-Ki;Ahn Sung-Ja;Nam Taek-keun;Yoon Mi-Sun;Song Ju-Young
    • Progress in Medical Physics
    • /
    • v.16 no.2
    • /
    • pp.82-88
    • /
    • 2005
  • In this study, the physical compensator made with the high density material, Cerrobend, and the electronic compensator realized by the movement of a dynamic multileaf collimator were analyzed in order to verify the properness of a design function in the commercial RTP (radiation treatment planning) system, Eclipse. The CT images of a phantom composed of the regions of five different thickness were acquired and the proper compensator which can make homogeneous dose distribution at the reference depth was designed in the RTP. The frame for the casting of Cerrobend compensator was made with a computerized automatic styrofoam cutting device and the Millennium MLC-120 was used for the electronic compensator. All the dose values and isodose distributions were measured with a radiographic EDR2 film. The deviation of a dose distribution was $\pm0.99 cGy\;and\;\pm1.82cGy$ in each case of a Cerrobend compensator and a electronic compensator compared with a $\pm13.93 cGy$ deviation in an open beam condition. Which showed the proper function of the designed compensators in the view point of a homogeneous dose distribution. When the absolute dose value was analyzed, the Cerrobend compensator showed a $+3.83\%$ error and the electronic compensator showed a $-4.37\%$ error in comparison with a dose value which was calculated in the RTP. These errors can be admtted as an reasonable results that approve the accuracy of the compensator design in the RTP considering the error in the process of the manufacturing of the Cerrobend compensator and the limitation of a film in the absolute dosimetry.

  • PDF

Evaluation of a Water-based Bolus Device for Radiotherapy to the Extremities in Kaposi's Sarcoma Patients (사지에 발병한 카포시육종의 방사선치료를 위한 물볼루스 기구의 유용성 고찰)

  • Ahn, Seung-Kwon;Kim, Yong-Bae;Lee, Ik-Jae;Song, Tae-Soo;Son, Dong-Min;Jang, Yung-Jae;Cho, Jung-Hee;Kim, Joo-Ho;Kim, Dong-Wook;Cho, Jae-Ho;Suh, Chang-Ok
    • Radiation Oncology Journal
    • /
    • v.26 no.3
    • /
    • pp.189-194
    • /
    • 2008
  • Purpose: We designed a water-based bolus device for radiation therapy in Kaposi's sarcoma. This study evaluated the usefulness of this new device and compared it with the currently used rice-based bolus. Materials and Methods: We fashioned a polystyrene box and cut a hole in order to insert patient's extremities while the patient was in the supine position. We used a vacuum-vinyl based polymer to reduce water leakage. Next, we eliminated air using a vacuum pump and a vacuum valve to reduce the air gap between the water and extremities in the vacuum-vinyl box. We performed CT scans to evaluate the density difference of the fabricated water-based bolus device when the device in which the rice-based bolus was placed directly, the rice-based bolus with polymer-vinyl packed rice, and the water were all put in. We analyzed the density change with the air gap volume using a planning system. In addition, we measured the homogeneity and dose in the low-extremities phantom, attached to six TLD, and wrapped film exposed in parallel-opposite fields with the LINAC under the same conditions as the set-up of the CT-simulator. Results: The density value of the rice-based bolus with the rice put in directly was 14% lower than that of the water-based bolus. Moreover, the value of the other experiments in the rice-based bolus with the polymer-vinyl packed rice showed an 18% reduction in density. The analysis of the EDR2 film revealed that the water-based bolus shows a more homogeneous dose plan, which was superior by $4{\sim}4.4%$ to the rice-base bolus. The mean TLD readings of the rice-based bolus, with the rice put directly into the polystyrene box had a 3.4% higher density value. Moreover, the density value in the case of the rice-based bolus with polymer-vinyl packed rice had a 4.3% higher reading compared to the water-based bolus. Conclusion: Our custom-made water-based bolus device increases the accuracy of the set-up by confirming the treatment field. It also improves the accuracy of the therapy owing to the reduction of the air gap using a vacuum pump and a vacuum valve. This set-up represents a promising alternative device for delivering a homogenous dose to the target volume.