• 제목/요약/키워드: Digital X-ray System

검색결과 225건 처리시간 0.03초

a-Se에 첨가된 Iodine의 조성비 변화에 따른 X선 검출특성 연구 (The Study on X-ray Detection Characteristics of Radiation Detective Sensor with Changing Composition Ratio of Iodine in a-Se)

  • 차병열;강상식;이규홍;김재형;남상희
    • 한국전기전자재료학회:학술대회논문집
    • /
    • 한국전기전자재료학회 2002년도 추계학술대회 논문집 Vol.15
    • /
    • pp.399-402
    • /
    • 2002
  • This paper is researched that electric characteristic of Digital x-ray radiography technique with changing composition ratio of Iodine. Dopant material, Iodine is evaporated with amorphous selenium. Thorugh the old papers say, doponted Iodine will be down the created trap level because of Arsenic dopant material in amorphous selenium. Arsenic material of Composition ratio in amorphous selenium is fixed with 0.3% and test sample is deposited composition of 30, 100, 300, 500, 700ppm with thermal evaporate system. Experimental measurement is performed by dark current and x-ray sensitivity in amorphous selenium based radition detector sensor. Fabricated test sample thickness is $30{\mu}m$ and injected voltage is $3{\mu}m$$6{\mu}m$$9{\mu}m$ to both electrode. Experimental results showed that the net charge of composition rate of 30ppm is 398.88 pc/mR/$cm^2$ very high. And increase of the Iodine composition ratio is tendency to the decrease of net charge. Doping changing composition of Iodine in amorphous selenium detector offered to basical information of amorphous selenium material.

  • PDF

Computed Radiograhpy (CR)를 통한 목재 수종별 X선 투과 이미지 해석을 위한 기초연구 (1) (Preliminary Study (1) for Development of Computed Radiography (CR) Image Analysis according to X-ray Non-destructive Test by Wood Species)

  • 송정일;김한슬
    • 보존과학회지
    • /
    • 제37권3호
    • /
    • pp.220-231
    • /
    • 2021
  • 문화재 방사선 비파괴 투과 조사 방식은 이미지플레이트(Image Plate, IP)를 사용하는 디지털 방사선영상시스템(Computed Radiography, CR)이 도입되면서 필름방식의 아날로그 이미지에서 점차 디지털 이미지로 교체되고 있다. 방사선 비파괴 투과 조사의 이미지 품질은 촬영 조건(관전압, 관전류, 노출시간), 영상 획득 매체, 촬영 거리, 유물의 두께 등이 영향을 미친다. 이 논문에서는 목재 문화재에 주로 사용되는 수종(소나무, 밤나무, 은행나무, 상수리나무)에 X선 선원의 투과 특성을 GE사(社)의 CR 영상 시스템을 통해 획득된 이미지의 그레이 스케일(Gray Scale)을 분석한 후 신호 대 잡음비(signal to noise ratio) 및 대조도를 평가하고, 이중-선 상질계(Duplex-wire image quality indicator), 구형파 차트(line-pair gauges), 선형 식별계(Image Quality Indicator)를 이용하여 이미지를 분석하였다.

Ektaspeed plus 필름을 이용한 일반 방사선시스템과 Digora를 이용한 디지탈 영상시스템의 밀도변화 비교연구 (Consideration of density matching technique of the plate type direct radiologic image system and the conventional X-ray film;first step for the subtraction)

  • 소성수;노현수;김창성;최성호;김기덕;조규성
    • Journal of Periodontal and Implant Science
    • /
    • 제32권1호
    • /
    • pp.199-211
    • /
    • 2002
  • Digital substraction technique and computer-assisted densitometirc analysis detect minor change in bone density and thus increase the diagnostic accuracy. This advantage as well as high sensitivity and objectivity which precludes human bias have drawn interest in radiologic research area. The objectives of this study are to verify if Radiographic density can be recognized in linear pattern when density profile of standard periapical radiograph with the aluminium stepwedge as the reference, was investigated under varies circumstances which can be encountered in clinical situations, and in addition to that to obtain mutual relationship between the existing standard radiographic system, and future digital image systems, by confirming the corelationship between the standard radiograph and Digora system which is a digital image system currently being used. In order to make quantitative analysis of the bone tissue, digital image system which uses high resolution automatic slide scanner as an input device, and Digora system were compared and analyzed using multifunctional program, Brain3dsp. The following conclusions were obtained. 1. Under common clinical situation that is 70kVp, 0.2 sec., and focal distance 10cm, Al-Equivalent image equation was found to be Y=11.21X+46.62 $r^2=0.9898$ in standard radiographic system, and Y=12.68X+74.59, $r^2=0.9528$ in Digora system, and linear relation was confirmed in both the systems. 2. In standard radiographic system, when all conditions were maintained the same except for the condition of developing solution, Al-Equivalent image equation was Y=10.07X+41.64, $r^2=0.9861$ which shows high corelationship. 3. When all conditions were maintained the same except for the Kilovoltage peak, linear relationship was still maintained under 60kVp, and Al-Equivalent image equation was Y=14.60X+68.86, $r^2=0.9886$ in the standard radiograhic system, and Y=13.90X+80.68, $r^2=0.9238$ in Digora system. 4. When all conditions were maintained the same except for the exposure time which was varied from 0.01 sec. to 0.8 sec., Al-Equivalent image equation was found to be linear in both the standard radiographic system and Digora system. The R-square was distributed from 0.9188 to 0.9900, and in general, standard radiographic system showed higher R-square than Digora system. 5. When all conditions were maintained the same except for the focal distance which was varied from 5cm to 30cm, Al-Equivalent image equation was found to be linear in both the standard radiographic system and Digora system. The R-square was distributed from 0.9463 to 0.9925, and the standard radiographic system had the tendency to show higher R-square in shorter focal distances.

형광체 기반 X선 광 변조기를 위한 비정질 셀레늄 필름 특성 (Characterization of the a-Se Film for Phosphor based X-ray light Modulator)

  • 강상식;박지군;조성호;차병열;신정욱;이건환;문치웅;남상희
    • 대한의용생체공학회:의공학회지
    • /
    • 제28권2호
    • /
    • pp.306-309
    • /
    • 2007
  • PXLM(Phosphor based x-ray light modulator) has a combined structure by phosphor, photoconductor, and liquid crystal and it can realize x-ray image of high resolution in clinical diagnosis area. In this study, we fabricated a photoconductor and investigated electrical and optical properties to confirm application possibility of radiator detector of PXLM structure. As photoconductor, amorphous selenium(a-Se), which is used most in DR(Digital radiography) of direct conversion method, was used and for formation of thin film, it was formed as $20{\mu}m-thick$ by using thermal vacuum evaporation system. For a produced a-Se film, through XRD(X-ray diffraction) and SEM(Scanning electron microscope), we investigated that amorphous structure was uniformly established and through optical measurement, for visible light of 40 $0\sim630nm$, it had absorption efficiency of 95 % and more. After fabricated a-Se film on the top of ITP substrate, hybrid structure was manufactured through forming $Gd_2O_3:Eu$ phosphor of $270{\mu}m-thick$ on the bottom of the substrate. As the result to confirm electrical property of the manufactured hybrid structure, in the case of appling $10V/{\mu}m$, leakage current of $2.5nA/cm^2$ and x-ray sensitivity of $7.31nC/cm^2/mR$ were investigated. Finally, we manufactured PXLM structure combined with hybrid structure and liquid crystal cell of TN(Twisted nematic) mode and then, investigated T-V(Transmission vs. voltage) curve of external light source for induced x-ray energy. PXLM structure showed a similar optical response with T-V curve that common TN mode liquid crystal cell showed according to electric field increase and in appling $50\sim100V$, it showed linear transmission efficiency of $12\sim18%$. This result suggested an application possibility of PXLM structure as radiation detector.

미세 대조도 영상을 기반한 디지털 방사선 영상 시스템의 자동노출제어 조절인자 간의 상관관계 분석 (Correlation Analysis of Control Factors in Automatic Exposure Control of Digital Radiography System Based on Fine Contrast Images)

  • 임세훈;성열훈
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제44권1호
    • /
    • pp.1-8
    • /
    • 2021
  • The purpose of this study was to analyze the effect of automatic exposure control (AEC) control factors in digital radiography systems based on the fine contrast images using coin phantoms. The AEC control factors were targeted at the range of dominent zone, sensitivity, and density. The dominent zone was divided into cases where a single coin was used to cover the field configuration, and cases where seven coins were used to cover the field configuration. The sensitivity was classified into three stages (200, 400, 800) and the density was classified into three stages (2.5, 0, 2.5). Image quality was evaluated by signal to noise ratio (SNR) and contrast to noise ratio (CNR). Then, the automatically exposed tube current was measured. As a result, the X-ray image of seven coins obtained a result value of about 1.2 times higher for SNR and 1.9 times higher for CNR than the X-ray image for one coin. The tube current was also about 1.6 times higher. In conclusion, In AEC, the higher the field configuration and dominent zone are matched and the higher the density, the lower the sensitivity, which increases the tube current and CNR, which increases the image quality. Therefore, it is judged that the appropriate setting of the range of dominent zone, sensitivity, and density of the control, which is the AEC control factor, could improve the fine contrast of images.

비정질 셀레늄을 이용한 직접방식의 디지털 방사선 검출기와 X-ray film과의 MTF측정을 통한 영상 질(quality) 비교평가에 관한 연구 (Comparison Study of the Modulation Transfer Function of a Prototype a-Se based Flat Panel Detector with Conventional Speed Class 400 Film/screen System)

  • Park, Jang-Yong;Park, Ji-Koon;Kang, Sang-Sik;Moon, Chi-Woong;Lee, Hyung-Won;Nam, Sang-Hee
    • 전자공학회논문지SC
    • /
    • 제40권3호
    • /
    • pp.163-171
    • /
    • 2003
  • 디지털 방사선 검출기의 성능을 측정하기 위해 변조전달할수(Modulation Transfer Function)을 측정하여 기존 방사선 영상 시스템인 필름/스크린과 비교하였다. 제작된 디지털 방사선 검출기는 TFT 패널위에 진공 증착법을 통해 비정질 셀레늄이 코팅되었다. 비교 측정물이 감도 400의 필름/스크린은 아날로그 방식의 방사선 검출기로서 현재 임상에서 이용되고 있는 필름이다. Square wave & slit 측정법을 통해 두 시스템의 변조전달함수를 측정하였다. 필름/스핀린의 반치폭(FWHM)은 357㎛(1.4 lp/mm at 50% spatial frequency)으로 측정되었으며, 디지털 방사선 검출 시스템은 200㎛(2.5 lp/mm at 50%)였다. 영상의 표현성능 평가의 결과 디지털 방사선 시스템은 필름 시스템보다 높은 것으로 나타났다.

X선 검출 물질의 전극 면적에 따른 신호특성 분석 (Analysis of Signal Properties in accordance with electrode area of x-ray conversion material)

  • 전승표;김성헌;조규석;정숙희;박지군;강상식;한용희;김강수;문치웅;남상희
    • 한국방사선학회논문지
    • /
    • 제4권1호
    • /
    • pp.5-9
    • /
    • 2010
  • 최근 광도전체와 형광체 기반의 디지털 방사선 검출기가 많은 관심을 받고 있으며, 이를 상업화 하기위한 많은 연구들이 이뤄지고 있다. 디지털 방사선 검출기률 제작하는 방법에서 크게 직접변환방식과 간접변환방식이 있다. 본 연구는 기존에 직접변환방식에 널리 사용되어 지고 있는 비정질 셀레늄 (Amorphous seleinum) 기반의 디지털 방사선검출기 보다 좋은 SNR (Signal-to-noise ratio) 동작 특성을 가지는 X선 검출 물질을 제작하여 X선 조사시 두께와 전기장 형성에 따른 차이점올 알아보기 위한 것이다. 본 연구에서는 기존의 진공 증착법이 두꺼운 대면적 필름제조가 어렵다는 문제점을 해결하고자 Screen-Print Method를 이용하여 전도성을 가진 ITO (Indium-tin-oxide)가 코팅 된 유리판에 제작하였다. 본 연구에 사용된 X-선 검출물질로는 다결정 $HgI_2$를 사용하였으며, 시편의 두께를 $150{\mu}m$로 제작하였으며, $3cm{\times}3cm$ 크기로 제작하였다. 상부전극은 Magnetron sputtering system을 사용하여 $3cm{\times}3cm$, $2cm{\times}2cm$, $1cm{\times}1cm$의 크기로 각각 다르게 하여 ITO(Indium-tin-oxide)를 증착 시킨 후, X선 조사시 $HgI_2$ 의 민감도와 누설전류, SNR 등을 측정하여 필름의 전기적 검출 특성을 정량적으로 평가하였으며, I-V테스트는 전류 적분(integration) 모드를 사용하였다. 그 결과 전극의 크기에 따라 신호량 증가 특성을 확인할 수 있었지만 신호량의 증가와 동시에 누설전류 또한 증가함으로써 전극의 크기에 따라 오히려 SNR 특성이 감소됨을 확인하였다. 향후 다양한 두께와 최적의 전극물질을 통해 신호대 잡음비를 개선시키기 위한 연구를 통해 최적화해야 할 것이다.

Study on rock fracture behavior under hydromechanical loading by 3-D digital reconstruction

  • Kou, Miaomiao;Liu, Xinrong;Wang, Yunteng
    • Structural Engineering and Mechanics
    • /
    • 제74권2호
    • /
    • pp.283-296
    • /
    • 2020
  • The coupled hydro-mechanical loading conditions commonly occur in the geothermal and petroleum engineering projects, which is significantly important influence on the stability of rock masses. In this article, the influence of flaw inclination angle of fracture behaviors in rock-like materials subjected to both mechanical loads and internal hydraulic pressures is experimentally studied using the 3-D X-ray computed tomography combined with 3-D reconstruction techniques. Triaxial compression experiments under confining pressure of 8.0 MPa are first conducted for intact rock-like specimens using a rock mechanics testing system. Four pre-flawed rock-like specimens containing a single open flaw with different inclination angle under the coupled hydro-mechanical loading conditions are carried out. Then, the broken pre-flawed rock-like specimens are analyzed using a 3-D X-ray computed tomography (CT) scanning system. Subsequently, the internal damage behaviors of failed pre-flawed rock-like specimens are evaluated by the 3-D reconstruction techniques, according to the horizontal and vertical cross-sectional CT images. The present experimental does not only focus on the mechanical responses, but also pays attentions to the internal fracture characteristics of rock-like materials under the coupled hydro-mechanical loading conditions. The conclusion remarks are significant for predicting the rock instability in geothermal and unconventional petroleum engineering.

위장내 음식물에 따른 방사선 감약 및 화질에 관한 연구 (The Study on the Attenuation of X-ray and Imaging Quality by Contents in Stomach)

  • 동경래;지연상;김창복;최성관;문상인;디터케빈
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제32권1호
    • /
    • pp.53-60
    • /
    • 2009
  • 인체의 $40{\sim}50%$인 연부조직에 X선을 조사하면 연부조직의 두께에 따라 투과선량이 감소하며 영상에도 질적 저하를 가져온다. 본 연구는 복부 촬영시 위장내 내용물에 따라 X선 감약 및 화질에 차이가 나타날 수 있다는 가정 하에 실제 위의 조직과 비슷한 조직등가물질 phantom을 제작하여 실제 위장내 내용물에 따른 X선의 감약 변화와 DR (Digital Ridiography)의 ROI수치를 비교해보았다. 단백질 두께가 증가할수록 X-선 감약에 의한 투과선량이 감소되었으며(p < 0.001), Film과 DR 영상에서의 ROI Mean값의 변화도 감소하였다(p <0.001). Film과 DR 영상에서의 ROI Mean값에 대한 비교는 Film이 DR 영상에 비해 단백질 두께에 따른 농도의 변화 값이 크게 나타났다(p < 0.001). 이와 같은 결과를 종합해 볼 때 DR system 촬영 보다는 필름을 사용하는 단순촬영에서 금식(NPO ; nothing by mouth)의 필요성이 더 중요함을 알 수 있다.

  • PDF

교육용 의료방사선 시뮬레이터 시스템 개발 및 연구 모델 제안 (Development of a Medical Radiation Simulator System for Education and Proposal of a Research Model)

  • 한창화;전영황;한재복;공창기;송종남
    • 한국방사선학회논문지
    • /
    • 제17권3호
    • /
    • pp.459-464
    • /
    • 2023
  • 첨단 기술의 발전으로 디지털 방사선영상 장비들이 많이 개발되어 정확한 진단과 치료에 많은 도움을 받고 있으며, 이를 안전하고 효과적으로 사용하기 위하여 전문적인 지식을 습득한 인력양성이 매우 중요하다. 진단용 X선 장비를 활용한 촬영 실습 교육에 있어서 학생들은 방사선 피폭의 위험성에 노출되어 있고, 일부 교육기관에서는 원자력안전법에 따른 관리의 어려움으로 X선 장비를 사용하지 않는 경우도 발생하고 있다. 이에 대한 해결책으로, 본 연구에서는 비전 센서와 자체 개발한 소프트웨어를 활용하여 방사선이 발생하지 않는 교육용 의료방사선 시뮬레이터를 개발하였고, 이를 통해 교육기관은 법에 따른 행정 이행 사항의 부담을 줄일 수 있고, 학생들은 방사선 피폭이 없는 건강한 실습환경에서 높은 수준의 교육 효과를 얻을 수 있게 하였다.