Screw loosening and changes in removal torque relative to abutment screw length in a dental implant with external abutment connection after oblique cyclic loading

  • Lee, Joo-Hee;Cha, Hyun-Suk
    • The Journal of Advanced Prosthodontics
    • /
    • 제10권6호
    • /
    • pp.415-421
    • /
    • 2018
  • PURPOSE. This study investigated the effects of abutment screw lengths on screw loosening and removal torque in external connection implants after oblique cyclic loading. MATERIALS AND METHODS. External connection implants were secured with abutment screws to straight abutments. The abutment-implant assemblies were classified into seven groups based on the abutment screw length, with each group consisting of five assemblies. A cyclic load of 300 N was applied at a $30^{\circ}$ angle to the loading axis until one million cycles were achieved. Removal torque values (RTVs) before and after loading, and RTV differences were evaluated. The measured values were analyzed using repeated measures of analysis of variance with the Student-Newman-Keuls multiple comparisons. RESULTS. All assemblies survived the oblique cyclic loading test without screw loosening. There was a significant decrease in the RTVs throughout the observed abutment screw lengths when the abutment-implant assemblies were loaded repeatedly (P<.001). However, the abutment screw length did not show significant difference on the RTVs before and after the experiment when the abutment screw length ranged from 1.4 to 3.8 mm (P=.647). CONCLUSION. Within the limit of this experiment, our results indicate that the abutment screw length did not significantly affect RTV differences after oblique cyclic loading when a minimum length of 1.4 mm (3.5 threads) was engaged. These findings suggest that short abutment screws may yield stable clinical outcomes comparable to long screws in terms of load resistance.

국내제작 임프란트 고정체와 지대주 유지나사의 적합도에 대한 주사전자현미경적 비교연구 (A STUDY ON FITNESS OF SEVERAL DOMESTIC IMPLANT FIXTURE AND ABUTMENT SCREWS)

  • 박영선;방몽숙;이석형
    • 대한치과보철학회지
    • /
    • 제42권1호
    • /
    • pp.94-109
    • /
    • 2004
  • Statement of problem : Several domestic dental implants have been developed since dental treatment being implants are increasing and popularized. However, they have not been used in domestic market like imported goods. Material and methods : This study was carried out to compare domestic and imported implants in aspect of fitness, dental implant fixture and abutment screw using scanning electron microscope. All experiments were performed under dry condition. Results : 1. Only in aspect of relation of dental implant fkxture and abutment screw, except only one group with point contact, good fitness was existed. 2. Home products must presevere in their efforts, so as excellent to fit.

임플랜트와 지대주 간 내측연결 시스템에서 Friction Fit와 Slip Fit에 따른 유한요소 응력분석 (Finite Element Stress Analysis of Implant Prosthesis According to Friction Fit or Slip Fit of Internal Connection System between Implant and Abutment)

  • 장두익;정승미;정재헌
    • 구강회복응용과학지
    • /
    • 제21권2호
    • /
    • pp.113-132
    • /
    • 2005
  • The purpose of this study was to assess the stress-induced pattern at the supporting bone, the implant fixture, the abutment and the abutment screw according to a friction-fit joint (Astra; Model 1) or slip- fit joint (Frialit-2; Model 2) in the internal connection system under vertical and inclined loading using finite element analysis. In conclusion, in the internal connection system of the implant and the abutment connection methods, the stress-induced pattern at the supporting bone, the implant fixture, the abutment and the abutment screw according to the abutment connection form had difference among them, and the stress distribution pattern usually had a widely distributed tendency along the inner surface of the implant fixture contacting the abutment post. The magnitude of the stress distributed in the supporting bone, the implant fixture, the abutment and the abutment screw was higher in the friction-fit joint than in the slip-fit joint. But it is considered that the further study is necessary about how this difference in the magnitude of the stress have an effect on the practical clinic.

임플란트 고정체와 지대주의 체결방식에 따른 스크류 풀림토크에 관한연구 (A Study on the Screw Loosening Torque According to the Type of Tightening the Implant Fixture and Abutment)

  • 박수철;김홍식;함성원
    • 대한치과기공학회지
    • /
    • 제35권3호
    • /
    • pp.201-207
    • /
    • 2013
  • Purpose: In this study, the loosening torque test was conducted with three implant products that are produced, approved and sold in Korea, which are manufactured in different fixture and abutment tightening methods (internal submerged type, internal morse taper type, and external type) to examine the loosening torque of the screw according to the method of tightening the implant fixture and abutment. Methods: In the loosening torque test, the three types of fixtures and abutments with different tightening methods were tightened by rotating them clockwise with a $30N{\cdot}cm$ force using a driver equipped with an electric torque meter. The results of the test are as follows. Results: The loosening torque values of the internal submerged type, internal morse taper type and external type implants were $24.10{\pm}0.742N{\cdot}cm$, $29.10{\pm}1.710N{\cdot}cm$, and $26.60{\times}1.636N{\cdot}cm$, respectively. Conclusion: The screw loosening torque values of the three fixture and abutment tightening methods were analyzed via Kruskal Wallis test layout, and they were significantly different (p<0.05).

치조제가 심하게 흡수된 하악 무치악환자의 나사-시멘트 유지형 임플란트 보철 수복증례 (Screw and cement retained implant prosthesis rehabilitation of mandibular edentulous patients with severely absorbed ridge)

  • 임중재
    • 대한치과기공학회지
    • /
    • 제41권2호
    • /
    • pp.149-156
    • /
    • 2019
  • Purpose: Prosthodontics for edentulous patients is a treatment technique using implant, which has impactful results in retention and support effects. Methods: As a retention technique, SCRP (screw and cement retained implant prosthesis) has reported in many studies as a beneficial method for both patients and curers, which can reduce errors in process of making abutment and top implant. Results: Prosthesis manufacturing, as polymerization method of hardened resin teeth with thermoplastic resin, is helpful for patients with aesthetic and financial situations regarding residual ridge and interocclusal relationship, also indicates reliable results in both retention and care. Conclusion: Using SCRP technique, we notably obtained a clinical and aesthetic outcome from five implants in anterior tooth, which are half fixable and detachable implants on screw of implant abutment by the technicians at anytime.

THE EFFECT OF INTERNAL IMPLANT-ABUTMENT CONNECTION AND DIAMETER ON SCREW LOOSENING

  • Ha, Chun-Yeo;Kim, Chang-Whe;Lim, Young-Jun;Jang, Kyung-Soo
    • 대한치과보철학회지
    • /
    • 제43권3호
    • /
    • pp.379-392
    • /
    • 2005
  • Statement of problem. One of the common problems of dental implant prosthesis is the loosening of the screw that connects each component, and this problem is more common in single implant-supported prostheses with external connection, and in molars. Purpose. The purposes of this study were: (1) to compare the initial abutment screw detorque values of the six different implant-abutment interface designs, (2) to compare the detorque values of the six different implant-abutment interface designs after cyclic loading, (3) to compare the detorque values of regular and wide diameter implants and (4) to compare the initial detorque values with the detorque values after cyclic loading. Material and methods. Six different implant-abutment connection systems were used. The cement retained abutment and titanium screw of each system were assembled and tightened to 32Ncm with digital torque gauge. After 10 minutes, initial detorque values were measured. The custom titanium crown were cemented temporarily and a cyclic sine curve load(20 to 320N, 14Hz) was applied. The detorque values were measured after cyclic loading of one million times by loading machine. One-way ANOVA test, scheffe’s test and Mann-Whitney U test were used. Results. The results were as follows : 1. The initial detorque values of six different implant-abutment connections were not significantly different(p>0.05). 2. The detorque values after one million dynamic cyclic loading were significantly different (p<0.05). 3. The SS-II regular and wide implant both recorded the higher detorque values than other groups after cyclic loading(p<0.05). 4. Of the wide implants, the initial detorque values of Avana Self Tapping Implant, MIS and Tapered Screw Vent, and the detorque values of MIS implant after cyclic loading were higher than their regular counterparts(p<0.05). 5. After cyclic loading, SS-II regular and wide implants showed higher detorque values than before(p<0.05).

Digital evaluation of axial displacement by implant-abutment connection type: An in vitro study

  • Kim, Sung-Jun;Son, KeunBaDa;Lee, Kyu-Bok
    • The Journal of Advanced Prosthodontics
    • /
    • 제10권5호
    • /
    • pp.388-394
    • /
    • 2018
  • PURPOSE. To measure axial displacement of different implant-abutment connection types and materials during screw tightening at the recommended torque by using a contact scanner for two-dimensional (2D) and three-dimensional (3D) analyses. MATERIALS AND METHODS. Twenty models of missing mandibular left second premolars were 3D-printed and implant fixtures were placed at the same position by using a surgical guide. External and internal fixtures were used. Three implant-abutment internal connection (INT) types and one implant-abutment external connection (EXT) type were prepared. Two of the INT types used titanium abutment and zirconia abutment; the other INT type was a customized abutment, fabricated by using a computer-controlled milling machine. The EXT type used titanium abutment. Screws were tightened at $10N{\cdot}cm$, simulating hand tightening, and then at the manufacturers' recommended torque ($30N{\cdot}cm$) 10 min later. Abutments and adjacent teeth were subsequently scanned with a contact scanner for 2D and 3D analyses using a 3D inspection software. RESULTS. Significant differences were observed in axial displacement according to the type of implant-abutment connection (P<.001). Vertical displacement of abutments was greater than overall displacement, and significant differences in vertical and overall displacement were observed among the four connection types (P<.05). CONCLUSION. Displacement according to connection type and material should be considered in choosing an implant abutment. When adjusting a prosthesis, tightening the screw at the manufacturers' recommended torque is advisable, rather than the level of hand tightening.

치과용 임플랜트에서 지대주 나사의 길이 및 반복 조임 횟수가 지대주 나사의 풀림에 미치는 영향 (THE INFLUENCE OF ABUTMENT SCREW LENGTH AND REPEATED TIGHTENING ON SCREW LOOSENING IN DENIAL IMPLANT)

  • 최진호;양재호;조원표;이재봉
    • 대한치과보철학회지
    • /
    • 제44권4호
    • /
    • pp.432-442
    • /
    • 2006
  • Statement of problem: One of common problems associated with dental implant is the loosening of abutment screws that retain the implants. Purpose : This study was performed to investigate the influence of abutment screw length and repeated tightening on screw loosening in dental implant. Material and method: Forty nine Hexplants (13mm length, 4.3mm diameter, Ti grade IV, Warantec. Co. Ltd. Seongnam, Korea) and cementation type abutments(straight abutment) and abutment screws (0.4mm/pitch) were divided into 7 groups, depending on abutment screw length. Each implant and abutment was tightened to 30Ncm by torque controller(MGT50, MARK-10 Inc., USA) and the removal torque values were measured during 10 consecutive closure/opening trials. Results and Conclusion: The results of comparing the removal torque value are as follows : 1. There is no significant difference in the removal torque value between groups in 10 consecutive closure/opening trials (p = 0.97). 2. If the fractured abutment screw is engaged in longer than 2.425 thread length, there is no significant difference in the preload between the fractured abutment screw and the new abutment screw when both are equally tightened to 30 Ncm. 3. The removal torque value in the 1st trial(24.510 Ncm) was lower than that in the 2nd, 3rd, 4th, 5th, 6th, 7th trials and the removal torque value in the 2nd trials(25.551 Ncm) was maximum and was decreased in 1311owing trials. The removal torque value in the 1st trial was significantly lower than that in the 2nd, 3rd, 4th trials and was significantly higher than that in the 8th, 9th, l0th trials(p<0.05). 4. In the 2nd, 3rd, 4th, 5th, 6th, 7th trials, the abutment screw was mainly influenced by settling effect and the higher preload was obtained In the 8th, 9th, l0th trials, the abutment screw was mainly influenced by adhesive wear and the progressively lower preload was obtained.

Long-term cumulative survival and mechanical complications of single-tooth Ankylos Implants: focus on the abutment neck fractures

  • Shim, Hye Won;Yang, Byoung-Eun
    • The Journal of Advanced Prosthodontics
    • /
    • 제7권6호
    • /
    • pp.423-430
    • /
    • 2015
  • PURPOSE. To evaluate the cumulative survival rate (CSR) and mechanical complications of single-tooth $Ankylos^{(R)}$ implants. MATERIALS AND METHODS. This was a retrospective clinical study that analyzed 450 single $Ankylos^{(R)}$ implants installed in 275 patients between December 2005 and December 2012. The main outcomes were survival results CSR and implant failure) and mechanical complications (screw loosening, fracture, and cumulative fracture rate [CFR]). The main outcomes were analyzed according to age, sex, implant length or diameter, bone graft, arch, and position. RESULTS. The 8-year CSR was 96.9%. Thirteen (2.9%) implants failed because of early osseointegration failure in 3, marginal bone loss in 6, and abutment fracture in 4. Screw loosening occurred in 10 implants (2.2%), and 10 abutment fractures occurred. All abutment fractures were located in the neck, and concurrent screw fractures were observed. The CSR and rate of screw loosening did not differ significantly according to factors. The CFR was higher in middle-aged patients (5.3% vs 0.0% in younger and older patients); for teeth in a molar position (5.8% vs 0.0% for premolar or 1.1% for anterior position); and for larger-diameter implants (4.5% for 4.5 mm and 6.7% for 5.5 mm diameter vs 0.5% for 3.5 mm diameter) (all P<.05). CONCLUSION. The $Ankylos^{(R)}$ implant is suitable for single-tooth restoration in Koreans. However, relatively frequent abutment fractures (2.2%) were observed and some fractures resulted in implant failures. Middle-aged patients, the molar position, and a large implant diameter were associated with a high incidence of abutment fracture.

임플란트 지대주 재질에 따른 지대주나사의 유한요소 응력 분석 (A Finite element stress analysis of abutment screw according to the implant abutment material)

  • 김남식;이명곤;홍민호
    • 대한치과기공학회지
    • /
    • 제38권1호
    • /
    • pp.1-6
    • /
    • 2016
  • Purpose: The present study was to determine the stress distribution of an abutment screw according to implant abutment material. Methods: This study was a tightening torque 10 Ncm, 20 Ncm, set to 30 Ncm, and a titanium alloy (Ti-6Al-4V), PEEK (polyetheretherketone), Endoligns (60% Carbon Fiber Reinforced PEEK) material of the custom abutment titanium alloy (Ti-6Al-4V) the stress distribution in the material of the abutment screw will be evaluated by the finite element analysis. Results: Abutment screw most stress has been concentrated on the interface between the fixture and the abutment was also part of the interface that the threads are started. Depending on the abutment of the abutment screw Material von Mises stress values are shown differently. 10Ncm T10 under the tightening torque of 294.2 MPa, P10 is 562.8 MPa, appeared to E10 is 295.8 MPa, 20Ncm tightening torque under T20 is 581.1 MPa, P20 is 1125 MPa, E20 was shown to 585.1 MPa, 30Ncm tightening torque under T30 is 918.2 MPa, P30 is 1795 MPa, E30 has appeared 925.1 MPa. Conclusion: If the abutment is used as Endoligns, it was confirmed that the abutment screw exhibits of von Mises stress value is similar to the titanium alloy abutment.