DOI QR코드

DOI QR Code

7-Diethylamino-4-methylcoumarin 기반 섬광체 제작 및 방사능 검출특성평가

Fabrication of 7-Diethylamino-4-methylcoumarin-based Scintillator for Gamma Radiation Detection

  • 민수정 (한국원자력연구원 해체기술개발부) ;
  • 노창현 (한국원자력연구원 해체기술개발부) ;
  • 서범경 (한국원자력연구원 해체기술개발부) ;
  • 홍상범 (한국원자력연구원 해체기술개발부)
  • Sujung Min (Decommissioning Technology Research Division, Korea Atomic Energy Research Institute) ;
  • Changhyun Roh (Decommissioning Technology Research Division, Korea Atomic Energy Research Institute) ;
  • Bumkyoung Seo (Decommissioning Technology Research Division, Korea Atomic Energy Research Institute) ;
  • Sangbum Hong (Decommissioning Technology Research Division, Korea Atomic Energy Research Institute)
  • 투고 : 2023.02.19
  • 심사 : 2023.03.17
  • 발행 : 2023.03.31

초록

Commercially used organic scintillation materials (1,4 di[2-(5phenyloxazolyl)] benzene) have low solubility in solvents and a wide emission energy range, which causes a decrease in detection efficiency. In this study, an organic liquid scintillator with improved detection efficiency was developed using 7-Diethylamino-4-methylcoumarin material to compensate for the disadvantages of existing organic scintillation detectors. And to evaluate the applicability of radiation measurement, the performance of a commercial plastic detector was compared. As a result of analyzing the 60Co detection characteristics by applying 7-Diethylamino-4-methylcoumarin as an alternative to 1,4 di[2-(5phenyloxazolyl)] benzene, the detection efficiency was improved around 2% compared with commercial scintillator when the 7-Diethylamino-4-methylcoumarin content was 0.04 wt%. Based on the results of this study, the possibility of improving detection efficiency through scintillator material modification was confirmed. In addition, since it is possible to discriminate nuclide through the spectrum correction algorithm, it will be possible to inspect and classify various decommissioning wastes generated during the decommissioning process.

키워드

과제정보

본 연구는 정부(과학기술정보통신부)의 재원으로 한국연구재단(NRF)의 지원을 받아 수행(2022M2E9A2063396와 RS-2022-00154985)되었습니다.

참고문헌

  1. Lee BS. 2015. Management of decommissioning waste from nuclear power plant. Autumn Workshop of korean radioactive waste society. Pusan, Korea.
  2. El-Daly SA, El-Azim SA, Elmekawey FM, Elbaradei BY, Shama SA and Asiri AM. 2012. Photophysical parameters, excitation energy transfer, and photoreactivity of 1,4-Bis(5-phenyl-2-oxazolyl)benzene (POPOP) laser dye. Int. J. Photoenergy 2012(458126):1-10. https://doi.org/10.1155/2012/458126
  3. Skorotetcky M, Borshchev OV, Surin NM, Odarchenko Y, Pisarev SA, Peregudova SM, Tornroos KW, Chernyshov D, Ivanov DA and Ponomarenko SA. 2017. Synthesis and photostability of 1,4-bis(5-phenyloxazol-2-yl)benzene (POPOP) structural isomers and their trimethylsilyl derivatives. Dyes Pigm. 141:128-136. https://doi.org/10.1016/j.dyepig.2017. 02.006
  4. May B, Poteau X, Yuan D and Brown RG. 1999. A Study of a highly efficiency resonance energy transfer between 7-N, N-diethylamino4-methylcoumarin and 9-butyl-4-butylamino-1,8-naphthalimide. Dyes Pigm. 42(1):79-84. https://doi.org/10.1016/S0143-7208(99)00011-X
  5. Yip RW and Wen YX. 1990. Photophysics of 7-diethylamino4-methylcoumarin: picosecond time-resolved absorption and amplified emission study. J. Photochem. Photobiol. A: Chem. 54(2):263-270. https://doi.org/10.1016/1010-6030(90)85039-Y
  6. Tolba MS, Abdil-Malik MA, Kamal El-Dean AM, Geies AA, Radwan SM, Zaki RM, Sayed M, Mohanmed SK and Abdel-Raheen SAA. 2022. An overview on synthesis and reactions of coumarin based compound. Curr. Chem. Lett. 11(1):29-42. https://doi.org/10.5267/j.ccl.2021.9. 007
  7. Ramanlingam A, Sivaram BM, Palanisamy PK and Masilamani V. 2000. Photophysics of TICT states of 7-diethylamino-4-methylcoumarin dye by energy transfer techniques. Spectrochim. Acta -A: Mol. Biomol. Spectrosc. 56(6):1205-1210. https://doi.org/10.1016/S1386-1425(99)00222-X
  8. Sastikumar D and Masilamani V. 1997. Influence of solvents on amplified spontaneous emission characteristics of 7-diethylamino4-methylcoumarin. Proc. Indian. Acad. Sci. 109:325-331. https://doi.org/10.1007/BF02875974
  9. Shin WG, Lee HC, Choi CI, Park CS, Kim HS and Min CH. 2015. A Monte Carlo study of an energy weighted algorithm for radionuclide analysis with a plastic scintillation detector. Appl. Radiat. Isot. 101:53-59. https://doi.org/10.1016/j.apradiso.2015.03.014
  10. Lee HC, Shin WG, Park HJ, Yoo DH, Choi CI, Park CS, Kim HS and Min CH. 2016. Validation of energy-weighted algorithm for radiation portal monitor using plastic scintillator. Appl. Radiat. Isot. 107:160-164. https://doi.org/10.1016/j.apradiso.2015.10.019
  11. Hamada MM, Rela PR, Costa FE and Mesquita CH. 1999. Radiation damage studies on the optical and mechanical properties of plastic scintillators. Nucl. Instrum. Methods Phys. Res. A. 422(1-3):148-154. https://doi.org/10.1016/S0168-9002(98)01091-2