Evaluation of the Resolution Characteristics by Using ATS 535H Phantom for Ultrasound Medical Imaging

초음파 의료영상에서 ATS 535H 팬텀을 이용한 해상력 특성 평가

  • Jung-Whan, Min (Department of Radiological Science, Shingu University) ;
  • Hoi-Woun, Jeong (Department of Radiological Science, Baekseok Culture University) ;
  • Hea-Kyung, Kang (Department of Radiological Science, Shingu University)
  • 민정환 (신구대학교 방사선과) ;
  • 정회원 (백석문화대학교 방사선과 ) ;
  • 강혜경 (신구대학교 방사선과)
  • Received : 2023.01.06
  • Accepted : 2023.01.25
  • Published : 2023.02.28

Abstract

This study was purpose to assessment of the resolution characteristics by using ATS 535H Basic quality assurance (QA) phantom for ultrasound. The ultrasound equipment was used Logiq P6 (Ultrasound, GE Healthcare System, Chicago, IL, USA). And the ultrasound transducer were used Convex 4C (4~5.5 MHz), Linear 11L (10~13 MHz), Sector 3SP (3~5.5 MHz) probe. As for the noise power spectrum (NPS) comparison results by using ATS 535H Basic QA ultrasound phantom and Convex 4C, Linear 11L, Sector 3SP probe. The NPS value of the Convex 4C probe image was 0.0049, Linear 11L probe image was 0.0049, Sector 3SP probe image was 0.1422 when the frequency is 1.0 mm-1. The modulation transfer function (MTF) comparison results by using ATS 535H Basic QA ultrasound phantom and Linear 11L probe the MTF value of the 3 cm focus image was 0.7511 and 4 cm focus image was 0.9001 when the frequency is 1.0 mm-1. This study was presented characteristics of spatial resolution a quantitative evaluation methods by using ultrasound medical images for QA of ultrasound medical QA phantom. The quality control (QC) for equipment maintenance can be efficiently used in the clinic due to the quantitative evaluation of the NPS and MTF as the standard methods. It is meaningful in that it is applied mutatis mutandis and presented the results of physical resolution characteristics of the ultrasound medical image.

Keywords

Acknowledgement

This study was supported by from the Shingu University Industry-Academic Cooperation Foundation Grant 2023.

References

  1. Sipila O, Mannila V, Vartiainen E. Quality assurance in diagnostic ultrasound. European Journal of Radiology. 2011;80:519-25. https://doi.org/10.1016/j.ejrad.2010.11.015
  2. Kaar M, Semturs F, Figl M, et al. Quality Assurance for Ultrasound Scanners using a durable tissue-mimicking Phantom and radial MTF. SPIE. 2014;V9040;3;90401C7
  3. Min JW, Jeong HW, Kim KW, et al. Comparison of Noise Power Spectrum in Measurements by Using International Electrotechnical Commission Standard Devices in Indirect Digital Radiography. Journal of Radiological Science and Technology. 2018;41(5): 457-62. https://doi.org/10.17946/JRST.2018.41.5.457
  4. Min JW, Jeong HW, Kim SC. Evaluation of Noise Power Spectrum Characteristics by Using Magnetic Resonance Imaging 3.0T. Journal of Radiological Science and Technology. 2021;44(1):279-88.
  5. Min JW, Jeong HW, Han JH, et al. Evaluation of the Resolution Characteristics by Using American College of Radiology Phantom for Magnetic Resonance Imaging. Journal of Radiological Science and Technology. 2022;45(1):11-7. https://doi.org/10.17946/JRST.2022.45.1.11
  6. Lee JW, Ahn KJ, Lee SK, Na DG, Oh CH, Chang YM, et al. Usefulness of ACR MRI phantom for quality assurance of MRI instruments. Journal Korean Society Radiology. 2006;54(1):47-55. https://doi.org/10.3348/jkrs.2006.54.1.47
  7. Miyati T, Fujita H, Kasuga T, et al. Measurements of MTF and SNR(f) using a subtraction method im MRI. Physics in Medicine and Biology. 2002;47: 2961-72. https://doi.org/10.1088/0031-9155/47/16/311
  8. Steckner MC, Drost DJ, Prato FS. Computing the modulation transfer function of a magnetic resonance imager. Med. Phys. 1994;21:483-9. https://doi.org/10.1118/1.597310
  9. Min JW, Jeong HW, Kim KW, et al. Study on the Resolution Characteristics by Using Magnetic Resonance Imaging 3.0T. Journal of Radiological Science and Technology. 2020;43(4):251-7. https://doi.org/10.17946/JRST.2020.43.4.251
  10. Jeong HW, Min JW, Kim JM, et al. Performance Characteristic of a CsI(Tl) Flat Panel Detector Radiography System. Journal of Radiological Science and Technology. 2012;35(2):109-17.
  11. Kim KW, Jeong HW, Min JW, et al. Evaluation of the Performance Characteristic for Mammography by Using Edge device. Journal of Radiological Science and Technology. 2016;39(3):415-20. https://doi.org/10.17946/JRST.2016.39.3.16
  12. Fujita H, Tasai DY, Itoh T, et al. A simple method for determining the modulation transfer function in digital radiography. IEEE Trans Med Imaging. 1992;11(1):34-9. https://doi.org/10.1109/42.126908
  13. Samei E, Flynn MJ, Reimann DA, et al. A method for measuring the presampled MTF of digital radiographic systems using an edge test device. Medical Physics. 1998;25:102.
  14. Greer PB, Van Doorn T. Evaluation of an algorithm for the assessment of the MTF using an edge method. Medical Physics. 2000;27:2048.
  15. IEC(International Electro-technical Commission) 62220-1. Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency. Geneva; 2003.
  16. Kim KW, Jeong HW, Min JW, et al. Measurement of Image Quality According to the Time of Computed Radiography System. Journal of Radiological Science and Technology. 2015;38(4):365-74. https://doi.org/10.17946/JRST.2015.38.4.05
  17. Kim KW, Jeong HW, Min JW, et al. Evaluation of the Modulation Transfer Function for Computed Tomography by Using American Association Physics Medicine Phantom. Journal of Radiological Science and Technology. 2016;39(2):193-8.  https://doi.org/10.17946/JRST.2016.39.2.09