Proficiency Test for the Dosimetry Audit Service Provider

  • Chul-Young Yi (Ionizing Radiation Metrology Group, Korea Research Institute of Standards and Science (KRISS)) ;
  • In Jung Kim (Ionizing Radiation Metrology Group, Korea Research Institute of Standards and Science (KRISS)) ;
  • Jong In Park (Ionizing Radiation Metrology Group, Korea Research Institute of Standards and Science (KRISS)) ;
  • Yun Ho Kim (Ionizing Radiation Metrology Group, Korea Research Institute of Standards and Science (KRISS)) ;
  • Young Min Seong (Ionizing Radiation Metrology Group, Korea Research Institute of Standards and Science (KRISS))
  • Received : 2022.08.31
  • Accepted : 2022.10.18
  • Published : 2022.12.31

Abstract

Purpose: The proficiency test was conducted to assess the performance of the dosimetry audit service provider in the readout practice of the dose delivered to patients in medical institutions. Methods: A certain amount of the absorbed dose to water for the high-energy X-ray from the medical linear accelerator (LINAC) installed in the Korea Research Institute of Standards and Science (KRISS) was delivered to the postal dose audit package given by the dosimetry audit service provider, in which the radio-photoluminescence (RPL) glass dosimeters were mounted. The dosimetry audit service provider read the RPL glass dosimeters and sent the readout dose value with its uncertainty to KRISS. The performance of the dosimetry audit service provider was evaluated based on the En number given in ISO/IEC 17043:2010. Results: The evaluated En number was -0.954. Based on the ISO/IEC 17043, the performance of the dosimetry service provider is "satisfactory." Conclusions: As part of the conformity assessment, the KRISS performed the proficiency test over the postal dose audit practice run by the dosimetry audit service provider. The proficiency test is in line with confirming the traceability of the medical institutions to the primary standard of absorbed dose to the water of the KRISS and ensuring the confidence of the dosimetry audit service provider.

Keywords

Acknowledgement

The KRISS has supported the present work under the project "Development of Measurement Standards for Medical Radiation (2022)." We thank the dosimetry audit service provider for allowing us to report the PT result.

References

  1. The International Atomic Energy Agency (IAEA). IAEA TRS-398. Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. Vienna: IAEA; 2006.
  2. International Organization for Standardization (ISO). Reference number ISO/IEC 17025. General requirements for the competence of testing and calibration laboratories. Geneva: ISO; 2017.
  3. International Organization for Standardization (ISO). Reference number ISO/IEC 17043. Conformity assessment-general requirements for proficiency testing. Geneva: ISO; 2010.
  4. Kim IJ, Kim BC, Kim JH, Chung JP, Kim HM, Yi CY. Building a graphite calorimetry system for the dosimetry of therapeutic X-ray beams. Nucl Eng Technol. 2017;49:810-816. https://doi.org/10.1016/j.net.2017.01.015
  5. Kim YH, Yi CY, Kim IJ, Kim BC, Kim JH, Seong YM, et al. Monte Carlo studies on dose conversion factors from graphite to water for high energy X-ray beams. Radiat Phys Chem. 2020;171:108760.
  6. Kim IJ, Kim BC, Yi CY, Shimizu M, Morishita Y, Saito N. Bilateral comparison of the absorbed dose to water in high energy X-ray beams between the KRISS and the NMIJ. Nucl Eng Technol. 2020;52:1511-1516.
  7. Thomas C. Rapport BIPM-05/07. The BIPM key comparison database (KCDB): technical aspects and data management. Sevres: Bureau International des Poids et Mesures; 2005.