Comparison of Metal Artifact Reduction Algorithms in Patients with Hip Prostheses: Virtual Monoenergetic Images vs. Orthopedic Metal Artifact Reduction

고관절 인공치환술 환자에서 금속 인공물 감소 방법의 비교: 가상 단일에너지영상 대 금속 인공물 감소기법

  • Hye Jin Yoo (Department of Radiology, Seoul National University Hospital) ;
  • Sung Hwan Hong (Department of Radiology, Seoul National University Hospital) ;
  • Ja-Young Choi (Department of Radiology, Seoul National University Hospital) ;
  • Hee Dong Chae (Department of Radiology, Seoul National University Hospital)
  • 류혜진 (서울대학교병원 영상의학과) ;
  • 홍성환 (서울대학교병원 영상의학과) ;
  • 최자영 (서울대학교병원 영상의학과) ;
  • 채희동 (서울대학교병원 영상의학과)
  • Received : 2021.07.22
  • Accepted : 2021.12.14
  • Published : 2022.11.01

Abstract

Purpose To assess the usefulness of various metal artifact reduction (MAR) methods in patients with hip prostheses. Materials and Methods This retrospective study included 47 consecutive patients who underwent hip arthroplasty and dual-energy CT. Conventional polyenergetic image (CI), orthopedic-MAR (OMAR), and virtual monoenergetic image (VMI, 50-200 keV) were tested for MAR. Quantitative analysis was performed in seven regions around the prostheses. Qualitative assessments included evaluation of the degree of artifacts and the presence of secondary artifacts. Results The lowest amount of image noise was observed in the O-MAR, followed by the VMI. O-MAR also showed the lowest artifact index, followed by high-keV VMI in the range of 120-200 keV (soft tissue) or 200 keV (bone). O-MAR had the highest contrast-to-noise ratio (CNR) in regions with severe hypodense artifacts, while VMI had the highest CNR in other regions, including the periprosthetic bone. On assessment of the CI of pelvic soft tissues, VMI showed a higher structural similarity than O-MAR. Upon qualitative analysis, metal artifacts were significantly reduced in O-MAR, followed by that in VMI, while secondary artifacts were the most frequently found in the O-MAR (p < 0.001). Conclusion O-MAR is the best technique for severe MAR, but it can generate secondary artifacts. VMI at high keV can be advantageous for evaluating periprosthetic bone.

목적 고관절 인공치환물을 가진 환자에서 여러 가지 금속인공물 감소 효과를 비교하였다. 대상과 방법 이 연구는 고관절 인공치환술과 이중에너지 전산화단층촬영을 시행한 47명 환자에서 시행하였다. 금속에서 발생한 인공물 감소효과는 서로 다른 3개의 영상(고식적 영상, 금속인공물 감소영상, 가상 단일에너지 영상)에서 비교하였다. 이를 위해 인공관절 주변 7곳에서 금속인공물에 대한 정량적 분석과 정성적 분석을 시행하였다. 결과 금속인공물 감소영상에서 가장 낮은 영상잡음과 인공물 지수를 보였고, 다음으로는 가상 단일에너지 영상이었다. 금속인공물 감소영상은 저음영 인공물이 매우 심한 영역에서만 높은 대조도를 보인 반면, 가상 단일에너지 영상은 인공물 주변 골조직과 그 외 영역에서 높은 대조도를 보였다. 연부조직 분석에서도 금속인공물 감소영상이 더 우수함을 보여 주었다. 정성적 분석에서도 금속인공물 감소영상이 가상 단일에너지 영상보다 인공물 감소 효과가 뛰어남을 밝혔지만, 이차적인 인공물 발생도 가장 흔히 발생하였다. 결론 금속인공물 감소영상이 심한 금속인공물감소에 가장 뛰어난 효과를 보였지만 새로운 이차적 인공물을 발생시켰다. 가상 단일에너지 영상은 인공물 주변 골조직 평가에서 우수함을 보였다.

Keywords

Acknowledgement

This study was supported by Seoul National University Hospital Research Fund (800-20180100). The funders had no role in study design, data collection, and analysis, decision to publish, or preparation of the manuscript.

References

  1. Wellenberg RHH, Hakvoort ET, Slump CH, Boomsma MF, Maas M, Streekstra GJ. Metal artifact reduction techniques in musculoskeletal CT-imaging. Eur J Radiol 2018;107:60-69 
  2. Boas FE, Fleischmann D. CT artifacts: causes and reduction techniques. Imaging Med 2012;4:229-240 
  3. Mallinson PI, Coupal TM, McLaughlin PD, Nicolaou S, Munk PL, Ouellette HA. Dual-energy CT for the musculoskeletal system. Radiology 2016;281:690-707 
  4. Kalender WA, Hebel R, Ebersberger J. Reduction of CT artifacts caused by metallic implants. Radiology 1987;164:576-577 
  5. Lell MM, Wildberger JE, Alkadhi H, Damilakis J, Kachelriess M. Evolution in computed tomography: the battle for speed and dose. Invest Radiol 2015;50:629-644 
  6. Shim E, Kang Y, Ahn JM, Lee E, Lee JW, Oh JH, et al. Metal artifact reduction for orthopedic implants (O-MAR): usefulness in CT evaluation of reverse total shoulder arthroplasty. AJR Am J Roentgenol 2017;209:860-866 
  7. Huang JY, Kerns JR, Nute JL, Liu X, Balter PA, Stingo FC, et al. An evaluation of three commercially available metal artifact reduction methods for CT imaging. Phys Med Biol 2015;60:1047-1067 
  8. Wellenberg RH, Boomsma MF, van Osch JA, Vlassenbroek A, Milles J, Edens MA, et al. Quantifying metal artefact reduction using virtual monochromatic dual-layer detector spectral CT imaging in unilateral and bilateral total hip prostheses. Eur J Radiol 2017;88:61-70 
  9. Grosse Hokamp N, Neuhaus V, Abdullayev N, Laukamp K, Lennartz S, Mpotsaris A, et al. Reduction of artifacts caused by orthopedic hardware in the spine in spectral detector CT examinations using virtual monoenergetic image reconstructions and metal-artifact-reduction algorithms. Skeletal Radiol 2018;47:195-201 
  10. Yoo HJ, Hong SH, Chung BM, Moon SJ, Choi JY, Chae HD, et al. Metal artifact reduction in virtual monoenergetic spectral dual-energy CT of patients with metallic orthopedic implants in the distal radius. AJR Am J Roentgenol 2018;211:1083-1091 
  11. Yue D, Fan Rong C, Ning C, Liang H, Ai Lian L, Ru Xin W, et al. Reduction of metal artifacts from unilateral hip arthroplasty on dual-energy CT with metal artifact reduction software. Acta Radiol 2018;59:853-860 
  12. Neuhaus V, Grosse Hokamp N, Zopfs D, Laukamp K, Lennartz S, Abdullayev N, et al. Reducing artifacts from total hip replacements in dual layer detector CT: combination of virtual monoenergetic images and orthopedic metal artifact reduction. Eur J Radiol 2019;111:14-20 
  13. Hegazy MA, Cho MH, Lee SY. A metal artifact reduction method for a dental CT based on adaptive local thresholding and prior image generation. Biomed Eng Online 2016;15:119 
  14. Wang Z, Bovik AC, Sheikh HR, Simoncelli EP. Image quality assessment: from error visibility to structural similarity. IEEE Trans Image Process 2004;13:600-612 
  15. Han SC, Chung YE, Lee YH, Park KK, Kim MJ, Kim KW. Metal artifact reduction software used with abdominopelvic dual-energy CT of patients with metal hip prostheses: assessment of image quality and clinical feasibility. AJR Am J Roentgenol 2014;203:788-795 
  16. Laukamp KR, Lennartz S, Neuhaus VF, Grosse Hokamp N, Rau R, Le Blanc M, et al. Correction to: CT metal artifacts in patients with total hip replacements: for artifact reduction monoenergetic reconstructions and post-processing algorithms are both efficient but not similar. Eur Radiol 2018;28:4524-4533 
  17. Kidoh M, Nakaura T, Nakamura S, Tokuyasu S, Osakabe H, Harada K, et al. Reduction of dental metallic artefacts in CT: value of a newly developed algorithm for metal artefact reduction (O-MAR). Clin Radiol 2014;69:e11-e16 
  18. Meinel FG, Bischoff B, Zhang Q, Bamberg F, Reiser MF, Johnson TR. Metal artifact reduction by dual-energy computed tomography using energetic extrapolation: a systematically optimized protocol. Invest Radiol 2012;47:406-414 
  19. Neuhaus V, Grosse Hokamp N, Abdullayev N, Rau R, Mpotsaris A, Maintz D, et al. Metal artifact reduction by dual-layer computed tomography using virtual monoenergetic images. Eur J Radiol 2017;93:143-148