Effect of the speed sintering schedule on the marginal and internal gaps of multi-layered zirconia prothesis

급속소결이 다층 지르코니아 치과보철물의 변연 및 내면 적합도에 미치는 영향

  • Kim, Se-Yeon (Department of Dental Laboratory Science, College of Health Science, Catholic University of Pusan) ;
  • Kim, Jae-Hong (Department of Dental Laboratory Science, College of Health Science, Catholic University of Pusan)
  • 김세연 (부산가톨릭대학교 보건과학대학 치기공학과) ;
  • 김재홍 (부산가톨릭대학교 보건과학대학 치기공학과)
  • Received : 2022.02.11
  • Accepted : 2022.03.16
  • Published : 2022.03.30

Abstract

Purpose: The purpose of this in vitro study is to evaluate the clinical acceptability of the marginal discrepancies and internal spacing of multi-layered zirconia prothesis by speed sintering. Methods: The two specimen types come from the same manufacturer but differ in the ingredients in layered zirconia and indication, that is anterior and posterior. The designs of both specimens were full contour crowns, and a total of 20 pieces were repeatedly fabricated, 10 for each group. The specimens were divided into two subgroups (n=10) and sintered with various total times (4 hours, 8 hours) at the maximum temperature (1,530℃). The gap between the two groups of multi-layered zirconia prostheses was measured using a silicone replica technique of 2D analysis method. The independent sample t-test was then used to compare and analyze the data obtained from the two groups (α=0.05). Results: The marginal and internal gap was superior in the 8-hour compared to the 4-hour group, and the results exhibited significant differences (p<0.05). All specimens showed that using the speed sintering schedule does not exceed the clinically permitted value of 120 ㎛, meaning zirconia prothesis using the speed sintering was adequate. Conclusion: The sintering condition is shown to affect the marginal and internal gaps of multi-layered zirconia restoration.

Keywords

Acknowledgement

This paper was supported by 2020 RESEARCH FUND offered from Catholic University of Pusan.

References

  1. Spyropoulou PE, Kamposiora P, Eliades G, Papavasiliou G, Razzoog ME, Bayne SC. Cyclic loading effect on color stability of unshaded versus shaded zirconia. J Esthet Restor Dent. 2016;28:77-84. https://doi.org/10.1111/jerd.12176
  2. Ueda K, Guth JF, Erdelt K, Stimmelmayr M, Kappert H, Beuer F. Light transmittance by a multi-coloured zirconia material. Dent Mater J. 2015;34:310-314. https://doi.org/10.4012/dmj.2014-238
  3. Miyazaki T, Hotta Y, Kunii J, Kuriyama S, Tamaki Y. A review of dental CAD/CAM: current status and future perspectives from 20 years of experience. Dent Mater J. 2009;28:44-56. https://doi.org/10.4012/dmj.28.44
  4. Juntavee N, Attashu S. Effect of different sintering process on flexural strength of translucency monolithic zirconia. J Clin Exp Dent. 2018;10:e821-e830. https://doi.org/10.4317/jced.54749
  5. Sabet H, Wahsh M, Sherif A, Salah T. Effect of different immersion times and sintering temperatures on translucency of monolithic nanocrystalline zirconia. Future Dent J. 2018;4:84-89. https://doi.org/10.1016/j.fdj.2017.09.003
  6. Ueda K, Watanabe F, Katsuta Y, Seto M, Ueno D, Hiroyasu K, et al. Marginal and internal fit of three-unit fixed dental prostheses fabricated from translucent multicolored zirconia: framework versus complete contour design. J Prosthet Dent. 2021;125:340.e1-340.e6. https://doi.org/10.1016/j.prosdent.2020.08.023
  7. Holmes JR, Bayne SC, Holland GA, Sulik WD. Considerations in measurement of marginal fit. J Prosthet Dent. 1989;62:405-408. https://doi.org/10.1016/0022-3913(89)90170-4
  8. Kim JH, Kim KB. An evaluation validity of the silicone replica technique at measurement on fit of fixed dental prostheses. J Dent Hyg Sci. 2012;12:566-571.
  9. Luthardt RG, Kuhmstedt P, Walter MH. A new method for the computer-aided evaluation of three-dimensional changes in gypsum materials. Dent Mater. 2003;19:19-24. https://doi.org/10.1016/S0109-5641(02)00013-1
  10. Price RB, Gerrow JD, Sutow EJ, MacSween R. The dimensional accuracy of 12 impression material and die stone combinations. Int J Prosthodont. 1991;4:169-174.
  11. Wagman SS. The role of coronal contour in gingival health. J Prosthet Dent. 1977;37:280-287. https://doi.org/10.1016/0022-3913(77)90070-1
  12. Ostlund LE. Cavity design and mathematics: their effect on gaps at the margins of cast restorations. Oper Dent. 1985;10:122-137.
  13. McLean JW, von Fraunhofer JA. The estimation of cement film thickness by an in vivo technique. Br Dent J. 1971. 131:107-111. https://doi.org/10.1038/sj.bdj.4802708
  14. Gulker I. Margins. N Y State Dent J. 1985;51:213-215, 217.
  15. Sorensen JA, Munksgaard EC. Interfacial gaps of resin cemented ceramic inlays. Eur J Oral Sci. 1995;103(2 Pt 1):116-120. https://doi.org/10.1111/j.1600-0722.1995.tb00128.x
  16. Assif D, Rimer Y, Aviv I. The flow of zinc phosphate cement under a full-coverage restoration and its effect on marginal adaptation according to the location of cement application. Quintessence Int. 1987;18:765-774.
  17. Sulaiman F, Chai J, Jameson LM, Wozniak WT. A comparison of the marginal fit of In-Ceram, IPS Empress, and Procera crowns. Int J Prosthodont. 1997;10:478-484.
  18. Jorgensen KD, Esbensen AL. The relationship between the film thickness of zinc phosphate cement and the retention of veneer crowns. Acta Odontol Scand. 1968;26:169-175. https://doi.org/10.3109/00016356809026130
  19. Passon C, Lambert RH, Lambert RL, Newman S. The effect of multiple layers of die-spacer on crown retention. Oper Dent. 1992;17:42-49.
  20. Grey NJ, Piddock V, Wilson MA. In vitro comparison of conventional crowns and a new all-ceramic system. J Dent. 1993;21:47-51. https://doi.org/10.1016/0300-5712(93)90051-Q
  21. Tuntiprawon M, Wilson PR. The effect of cement thickness on the fracture strength of all-ceramic crowns. Aust Dent J. 1995;40:17-21. https://doi.org/10.1111/j.1834-7819.1995.tb05607.x
  22. Kim JH, Kim KB. Effect of zirconia ceramic sintering condition on the precision of fit in dental restorations. J Korean Acad Dent Technol. 2020;42:121-128. https://doi.org/10.14347/KADT.2020.42.2.121