DOI QR코드

DOI QR Code

Effect of Reducing Scattering Radiation Exposure of Medical Staffs When Additional Shielding is Used in Interventional Radiology

중재적 방사선시술에서 부가 차폐체 사용 시 종사자의 산란선 피폭 감소효과

  • Kim, Min-Jun (Dept. of Radiological Science, Gachon University) ;
  • Baek, Kang-Nam (Gachon University Graduate School of Health Science) ;
  • Kim, Sungchul (Dept. of Radiological Science, Gachon University)
  • Received : 2021.11.29
  • Accepted : 2021.12.27
  • Published : 2021.12.31

Abstract

This article is designed to look into the radiation exposure dose to each body part and the shielding effect for workers using an additional shielding to reduce their radiation exposured by scattering radiation which is generated in a space between the operating table and lead curtain during interventional radiology(IR) procedures. After placing a human phantom on the table of SIEMENS' angiography machine, the following measurements were taken, depending on the presence of an additional shield of lead equivalent of 0.25 mmPb, manufactured for this purpose: dose to gonad, dose to an area where the personal dosimeter is placed, and dose to an area of eye lens is located. An ion chamber(chamber volume 1,800 cc) was utilized to measure scattering radiation. The two imaging tests were carried out as follows: fluoroscopy of the abdomen (66 kV, 100 mA, 60 seconds) and of the head (70 kV, 65 mA, 60 seconds); and digital subtraction angiography(DSA) of the abdomen (67 kV, 264 mA, 20 seconds) and of the head (79 kV, 300 mA, 20 seconds). In all the experiments, the shielding efficiency of the gonad position was the largest at 59.8%. In case an additional shielding was used as protection against scattering radiation that came through the operating table and the lead curtain during an IR, the radiation shielding efficiency was estimated to be up to 59.8%, leading to a conclusion that its presence may effectively reduce the radiation exposure dose of medical staffs.

Keywords

References

  1. Bor D, Sancak T, Olgar T, Elcim Y, Adanali A, Sanlidilek U, et al. Comparison of effective doses obtained from dose-area product and air kerma measurements in interventional radiology. The British Journal of Radiology. 2004;77(916):315-22. https://doi.org/10.1259/bjr/29942833
  2. ICRP Publication 85. Avoidance of Radiation Injuries from Medical Interventional Procedures; 2000.
  3. Radiation Protection Guidelines for Patient Dose Reduction in Interventional Radiation Procedures. Ministry of food and drug safety. 2014. [cited 2021 November 20]. Available from: https://www.mfds.go.kr/brd/m_1060/down.do?brd_id=data0011&seq=12392&data_tp=A&file_seq=1
  4. Miller DL. Society for Interventional Radiology. Interventional fluoroscopy: Reducing radiation risks for patients and staff. J Vasc Interv Radiol. 2009;20(7):S274. https://doi.org/10.1016/j.jvir.2009.04.057
  5. Artschan RE, Brettle DS, Chase K, Fender A, Howells PG, Buchan S. An investigation of the radiation doses to the lower legs and feet of staff undertaking interventional procedures. Br J Radiol. 2014;87(1038):20130746. https://doi.org/10.1259/bjr.20130746
  6. Koenig AM, Maas J, Viniol S, Etzel R, Fiebich M, Thomas RP, et al. Scatter radiation reduction with a radiation-absorbing pad in interventional radiology examinations. Eur J Radiol. 2020;132(5):109245. https://doi.org/10.1016/j.ejrad.2020.109245
  7. Kim KP, Miller DL, Balter S, Kleinerman RA, Linet MS, Kwon D, Simon SL. Occupational radiation doses to operators performing cardiac catheterization procedures. Health Phys. 2008;94(3):211-27. https://doi.org/10.1097/01.HP.0000290614.76386.35
  8. Kim SH, KIm YJ, Kwak JS. Development and Radiation shield effects of Dose Reduction Fiber for Scatter ray in CT Exams. Journal of the Korea Academia-Industrial cooperation Society. 2013;14(4):1871-6. https://doi.org/10.5762/KAIS.2013.14.4.1871
  9. Yoon CH, Yoon SH, Choi JG. Radiation Exposure According to Radiation Technologist' Working Departments. Journal of Radiological Science and Technology. 2008;31(3):217-22.
  10. Nowak M, Carbonez P, Krauss M, Verdun FR, Damet J. Characterisation and mapping of scattered radiation fields in interventional radiology theatres. Sci Rep. 2020;10(1):18754. https://doi.org/10.1038/s41598-020-75257-5
  11. Han SC, Hong DH. Measurement of Spatial Dose Distribution for evaluation operator dose during Neuro-interventional Procedures. Journal of Radiological Science and Technology. 2016;39(3):323-8. https://doi.org/10.17946/JRST.2016.39.3.04
  12. Kang YM, Cho JH, Kim SC. Changes in entrance surface dose in relation to the location of shielding material in chest computed tomography. Radiation Effects and Defects in Solids. 2015;170(7-8):645-50. https://doi.org/10.1080/10420150.2015.1078803
  13. Kim JH, Kim JS, Kim YE, Jang IH. Comparison of worker exposure dose according to pulse rate change during TACE and analysis of spatial scattering dose distribution in laboratory. Journal of Korean Society of Cardio-Vascular Interventional Technology. 2014;17(1):137-44. [cited 2021 November 20]. Available from: https://www.dbpia.co.kr/journal/articleDetail?nodeId=NODE06401740
  14. Kim SC, Kim S. Physics of radiology. Hanmi Medical Publishing. 2009:123-126.
  15. Jang MS, Her SB, Kim DY, Lee WS, Lee HK, Song CW. Comparison of patient's radiation dose according to whether or not collimation is used during TACE treatment. Journal of Korean Society of Cardio-Vascular Interventional Technology. 2014; 17(1):151-5. [cited 2021 November 20]. Available from: https://www.dbpia.co.kr/journal/article-Detail?nodeId=NODE06401742
  16. Jeon ME, Lim CH, Jung HR, You IG, Hong DH, Kang BS. A Study on Radiation Exposure Dose of Operator During Interventional Radiology Procedure. Journal of Radiological Science and Technology. 2012;35(3):219-226.