Effectiveness Evaluation of Displacement Accommodatable Pressure Measuring Jig for Quality Assessment of Pressure Application Device

압력 인가 장치의 품질관리를 위한 변위 수용이 가능한 압력 측정용 지그의 유효성 평가

  • 문창수 (KMG(거명) 기업부설연구소) ;
  • 전성철 (부산테크노파크 고령친화산업지원센터) ;
  • 노시철 (KMG(거명) 기업부설연구소)
  • Received : 2020.05.04
  • Accepted : 2020.06.22
  • Published : 2020.06.30

Abstract

Recently, a variety of electric anesthetics devices have been developed and used in clinical practice to reduce the fatigue of the operator during local anesthesia for dental procedures and to compensate for the disadvantages of manual anesthesia device. In this electric anesthesia injection device, the accurate and constant delivery of pressure for drug infusion is a very important performance factor. In order to evaluate the accuracy of the transfer pressure, a small pressure gauge using a load cell is often used, but since the elastic body inside the load cell may not be able to accommodate a sufficient displacement, an error may occur when evaluating pressure performance. For these reasons, in this study, we proposed and evaluated a silicon-chrome steel (Si-Cr steel) spring jig that can accommodate relatively large displacements that can be used when evaluating the performance of a pressure-controlled pressure application device using a load cell type pressure gauge. As a result of the pressure transmissibility test and repeated measurement results using a commercial dental anesthesia injection device, a more stable result was obtained when using a spring jig, and it was confirmed that the frequency of abnormally high measurement was reduced.

최근 치과 시술을 위한 국소마취 시 시술자의 피로감을 줄이고 수동 마취기의 단점을 보완한 다양한 전동형 마취액 주입기가 개발되어 임상에서 사용되고 있다. 이러한 전동형 마취액 주입기기에서는 약물 주입을 위한 압력을 정확하고 일정하게 전달하는 것이 매우 중요한 성능 요소이다. 이러한 전달 압력의 정확성을 평가하기 위하여 로드셀을 이용한 소형의 압력계가 많이 사용되는데, 로드셀 내부의 탄성체가 충분한 변위를 수용할 수 없는 경우가 있어 압력 성능 평가 시 오차를 발생시킬 수도 있다. 이에 본 연구에서는 로드셀 방식의 압력측정기를 이용하여 압력 제어형 압력 인가 장치의 성능을 평가할 때 사용 가능한 비교적 큰 변위 수용이 가능한 실리콘-크롬강(Si-Cr강) 스프링 지그를 제안하였으며 유효성을 평가하였다. 제안된 스프링 지그의 압력 전달률과 상용의 치과용 마취액 주입장치를 이용한 반복 측정 결과, 스프링 지그 사용 시 보다 안정적인 결과를 얻을 수 있었으며, 비정상적으로 높게 측정되는 빈도가 줄어드는 것을 확인할 수 있었다.

Keywords

References

  1. M. A. Ghavimi, Y. Kananizadeh, S. Hajizadeh, et. al., "Overview of local anesthesia techniques," in A textbook of advanced oral and maxillofacial surgery, vol. 2, 1st ed., Rijeka, Croatia: Intech, pp.18-22, 2015.
  2. H. Khalil, "Local anesthetics dosage still a problem for most dentists: A survey of current knowledge and awareness," The Saudi. Journal for Dental Research, vol. 5, no. 1, pp. 49-53, 2014. https://doi.org/10.1016/j.ksujds.2013.08.002
  3. S. S. Sharma, S. Aruna Sharma, C. Saravanan, "Sathyabama. Newer local anaesthetic drugs and delivery systems in dentistry-an update," J. Dent. Med. Sci, vol. 1, pp. 10-16, 2012.
  4. Y. K. Kim, I. H. Cho, J. H. Kwon, H. S. Kim, "Comparison of pain perception using computer-controlled anesthetic device and aspirating syringe," The Journal of Korean Academy of Periodontology, vol. 34, no. 3, pp. 639-646, 2004. https://doi.org/10.5051/jkape.2004.34.3.639
  5. S. G. Cho, Q. S. Auh, Y. H. Chun, J. P. Hong, "Blood flow changes in the masseter muscle and overlying skin following various functional waves of transcutaneous electrical nerve stimulation," Korean Journal of Oral Medicine, vol. 34, no. 1, pp. 115-122, 2009.
  6. G. Lappegard, J. Kohler, M. Jackson, J. O. Hagen, "Characteristics of subglacial drainage systems deduced from load-cell measurements," Journal of Glaciology, vol. 52, no. 176, pp. 137-148, 2006. https://doi.org/10.3189/172756506781828908
  7. J. H. Choi, J. Y. Seo, J. M. Park, "Development of a foot pressure distribution measuring device for lower limb rehabilitaion," The Journal of Korea Institute of Convergence Signal Processing, vol. 18, no. 1, pp. 1-5, 2017.
  8. F. Wang, Y. Chen, "Dynamic characteristics of pressure compensator in underwater hydraulic system," IEEE/ASME Transactions on Mechatronics, vol. 19, no. 2, pp. 777-787, 2014. https://doi.org/10.1109/TMECH.2013.2260829