DOI QR코드

DOI QR Code

Performance Evaluation of the Developed Diagnostic Multi-Leaf Collimator and Implementation of Fusion Image of X-ray Image and Infrared Thermography Image

개발한 진단용 다엽조리개 성능평가 및 X선영상과 적외선체열영상의 융합영상 구현

  • Kwon, Soon-Mu (Department of Radiolgic Technology, Daegu Health College) ;
  • Shim, Jae-Goo (Department of Radiolgic Technology, Daegu Health College) ;
  • Chon, Kwon-Su (Department of Radiolgical Science, Catholic University of Daegu)
  • 권순무 (대구보건대학교 방사선과) ;
  • 심재구 (대구보건대학교 방사선과) ;
  • 천권수 (대구가톨릭대학교 방사선학과)
  • Received : 2019.09.06
  • Accepted : 2019.10.22
  • Published : 2019.10.31

Abstract

We have developed and applied a diagnostic Multi-Leaf Collimator (MLC) to optimized the X-ray field in medical imaging and the usefulness evaluated through the fusion of infrared image and X-ray image acquired by infrared camera. The hand and skull radiography with multi-leaf collimator(MLC) showed significant area dose reductions of 22.9% and 31.3% compared to ARC and leakage dose was compliant with KS A 4732. Also scattering doses of 50 cm and 100 cm showed a significant decrease to confirm the usefulness of MLC. It was confirmed that the fusion of infrared images with an adjustable degree of transparency was possible in the X-ray images. Therefore, fusion of anatomical information with physiological convergence is expected to contribute and improvement of diagnostic ability. In addition, the feasibility of convergence X-ray imaging and DITI devices and the possibility of driving MLC with infrared images were confirmed.

Keywords

References

  1. Teng J, Wang X, Zhang J, Wang S, Huo P. A multimodality medical image fusion algorithm based on wavelet transform. In International Conference in Swarm Intelligence, Springer, Berlin, Heidelberg. 2010;627-33.
  2. James AP, Dasarathy BV. Medical image fusion, A survey of the state of the art. Information Fusion. 2014;19:4-19. https://doi.org/10.1016/j.inffus.2013.12.002
  3. Jeong WK. Radiation exposure and its reduction in the fluoroscopic examination and fluoroscopyguided interventional radiology. Journal of the Korean Medical Association. 2011:54(12);1269-76. https://doi.org/10.5124/jkma.2011.54.12.1269
  4. Korean society of medical imaging technology. Textbook of radiographic positioning and clinical diagnosis. Chungku publisher; 2017.
  5. http://www.thermology.or.kr/ksdt/learn_021.php
  6. Research society of radiation equipment. Radiation equipment. Chungku publisher; 2017.
  7. KS A 4732. Variable X-ray beam limiting devices for diagnostic use, 1997.
  8. IEC 60601-1-33. General rule for X-ray equipment, 2008.
  9. Han SC, Park SW. Monte Carlo simulation for development of diagnostic multileaf collimator. Journal of Radiological Science and Technology. 2016;39(4):595-600. https://doi.org/10.17946/JRST.2016.39.4.15
  10. Han JH, Baek SC, Lee JH, Park DH. Development and utility evaluation of new Multi-Leaf Collimator for Diagnostic X-ray Equipment. Journal of Electrical Engineering & Technology. 2018;13(2):936-42. https://doi.org/10.5370/JEET.2018.13.2.936
  11. Le Heron JC. Estimation of effective dose to the patient during medical X-ray examinations from measurements of the dose-area product. Physics in Medicine & Biology. 1992;37(11):2117. https://doi.org/10.1088/0031-9155/37/11/008
  12. Kwon SM, Kim BS, Park HJ, Kang YH. Variation of Image Quality and Dose by Applying Multi-Leaf Collimator for Digital Mammography. Journal of the Korean Society of Radiology. 2015; 9(7):535-40. https://doi.org/10.7742/jksr.2015.9.7.535
  13. Kang Y, Cho P. Reduction of Radiation Dose according to Geometric Parameters from Digital Coronary Angiography. Journal of the Korean Society of Radiology. 2013;7(4):277-84. https://doi.org/10.7742/jksr.2013.7.4.277
  14. Bor D, Sancak T, Olgar T, Elcim Y, Adanali A, Sanlidilek U, et al. Comparison of effective doses obtained from dose-area product and air kerma measurements in interventional radiology. The British Journal of Radiology. 2004;77(916):315-22. https://doi.org/10.1259/bjr/29942833
  15. Choi SK. Change of the Scattered Dose by Field Size in X-ray Radiography. The Journal of the Korea Contents Association. 2013;13(3):198-203. https://doi.org/10.5392/JKCA.2013.13.03.198
  16. Cho PK. Distribution of the scatter ray in fluoroscopy X-ray room. The Journal of the Korea Contents Association. 2011;11(10):349-54. https://doi.org/10.5392/JKCA.2011.11.10.349
  17. Buchanan JB, Spratt JS, Heuser LS. Tumor growth, doubling times, and the inability of the radiologist to diagnose certain cancers. Radiologic Clinics of North America. 1983;21(1):115-26.
  18. Park JK. Measurement of ROI temperature in herniation of intervertebral disc patients using DITI. Journal of the Korean Society of Radiology. 2017;11(11):272-8.
  19. Sim JH. Diagnostic significance of digital infrared thermographic imaging after cerebrovascular stroke. Journal of Korean Neurosurgical Society. 1998;27(7):890-7.
  20. Sanchis-Sanchez E, Salvador-Palmer R, Codoner-Franch P, Martin J, Vergara-Hernandez C, Blasco J, Cibrian R. Infrared thermography is useful for ruling out fractures in pediatric emergencies. European Journal of Pediatrics, 2015;174(4):493-9. https://doi.org/10.1007/s00431-014-2425-0
  21. Kwon SM, Kang YH, Park BH. X-ray inspection device and its inspection method. Korea Patent. 2017;10-1743469.