Intercomparison Exercise at Harshaw 6600, DVG-02TM, and D-Shuttle Dosimeters for the Individual Monitoring of Ionizing Radiation

  • Received : 2018.10.16
  • Accepted : 2019.05.20
  • Published : 2019.06.30

Abstract

Background: The main goal of experiments is to compare various operational and technical characteristics of D-Shuttle semiconductor personal dosimeters of the Japanese company "Chiyoda Technol Corporation" and Harshaw thermoluminescent dosimeters (TLD) manufactured by "Thermo Fisher Scientific" and DTL-02 of the Russian Research and Production Enterprise (RPE) "Doza" by their occupational and calibration exposure at various dose equivalents from 0.5 to 20 mSv of gamma-radiation. Materials and Methods: Besides dosimeters DTL-02, D-Shuttle and Harshaw TLD, there were also used: (1) the primary reference radionuclide source Hopewell Designs IAEA: G10-1-12 with $^{137}Cs$ isotope (an error is not more than 6% and activity is 20 Ci), and (2) the verification device UPGD-2M of RPE "Doza" and installed in the National Center for Expertise and Certification of the Republic of Kazakhstan (Kapchagai, the National Center for Expertise and Certification). Results and Discussion: The main results of researches are the following: (1) TLDs for Harshaw 6600 and DVG-02TM have an approximately equal measurement accuracy of the individual dose equivalents in the range from 0.5 to 20 mSv of gamma-radiation. (2) Advantages of dosimeters for Harshaw 6600 are due to the high measurement productivity and opportunity to indicate the dose on the skin $H_p$(0.07). Advantages of DVG-02TM consist of operation simplicity and lower cost than of Harshaw 6600. (3) D-Shuttles are convenient for use in the current and the operational monitoring of ionizing radiation. Measurement accuracy and 10% linearity of measurements are ensured when D-Shuttle is irradiated with dose equivalents below 1 mSv at the equivalent dose rate not higher than $3mSv{\cdot}hr^{-1}$. This allows using D-Shuttle at a routine technological activity. Conclusion: The obtained results of experiments demonstrate advantages and disadvantages of D-Shuttle semiconductor dosimeters in comparison with two TLD systems of DVG-02TM and Harshaw 6600.

Keywords

References

  1. Ministry of National Economy of the Republic of Kazakhstan. Sanitary-epidemiological requirements for ensuring radiation safety. Sanitary rules 261. 2015;12-43.
  2. Kudryashev VA, Kim DS, Nurtazin YR, On the standards of individual dosimetry of external radioactive irradiation. Metrological scientific and technical journal "The world of measurements". 2016;174:11-14.
  3. Martynyuk AN. DVG-02TM dosimetric thermoluminescent unit; Operating Manual PIGU.412113.003RE. Moscow. JSC SIE Doza. 2005;10-18.
  4. Thermo Scientific Harshaw Model 6600 LITE TLD Reader; Operator's Manual #6600LT-W-O-0810-002. Oakwood Village, Ohio. One Thermo Fisher Way. 2010;250.
  5. D-Shuttle Dosimeter Management System (for Research Institutions); Instruction Manual. Oarai. Chyoda Technol Corporation. 2015;27.
  6. Cross WG, Böhm J, Charles M, Piesch E, Seltzer SM, Dosimetry of External Beta Rays for Radiation Protection (Report 56). Journal of the International Commission on Radiation Units and Measurements. 1997;29:133.
  7. International Commission on Radiological Protection. Conversion Coefficients for use in Radiological Protection against External Radiation. ICRP Publication 74. 1997;18-26.