Abstract
Purpose : The purpose of this study is to compare and analyze the set-up errors using the Combine IGRT with ExacTrac and CBCT phased in the treatment of Stereotatic Body Radiotherapy. Methods and materials : Patient who were treated Stereotatic Body Radiotherapy in the ulsan university hospital from May 2014 to november 2017 were classified as treatment area three brain, nine spine, three pelvis. First using ExacTrac Set-up error calibrated direction of Lateral(Lat), Longitudinal(Lng), Vertical(Vrt), Roll, Pitch, Yaw, after applied ExacTrac moving data in addition to use CBCT and set-up error calibrated direction of Lat, Lng, Vrt, Rotation(Rtn). Results : When using ExacTrac, the error in the brain region is Lat $0.18{\pm}0.25cm$, Lng $0.23{\pm}0.04cm$, Vrt $0.30{\pm}0.36cm$, Roll $0.36{\pm}0.21^{\circ}$, Pitch $1.72{\pm}0.62^{\circ}$, Yaw $1.80{\pm}1.21^{\circ}$, spine Lat $0.21{\pm}0.24cm$, Lng $0.27{\pm}0.36cm$, Vrt $0.26{\pm}0.42cm$, Roll $1.01{\pm}1.17^{\circ}$, Pitch $0.66{\pm}0.45^{\circ}$, Yaw $0.71{\pm}0.58^{\circ}$, pelvis Lat $0.20{\pm}0.16cm$, Lng $0.24{\pm}0.29cm$, Vrt $0.28{\pm}0.29cm$, Roll $0.83{\pm}0.21^{\circ}$, Pitch $0.57{\pm}0.45^{\circ}$, Yaw $0.52{\pm}0.27^{\circ}$ When CBCT is performed after the couch movement, the error in brain region is Lat $0.06{\pm}0.05cm$, Lng $0.07{\pm}0.06cm$, Vrt $0.00{\pm}0.00cm$, Rtn $0.0{\pm}0.0^{\circ}$, spine Lat $0.06{\pm}0.04cm$, Lng $0.16{\pm}0.30cm$, Vrt $0.08{\pm}0.08cm$, Rtn $0.00{\pm}0.00^{\circ}$, pelvis Lat $0.06{\pm}0.07cm$, Lng $0.04{\pm}0.05cm$, Vrt $0.06{\pm}0.04cm$, Rtn $0.0{\pm}0.0^{\circ}$. Conclusion : Combine IGRT with ExacTrac in addition to CBCT during Stereotatic Body Radiotherapy showed that it was possible to reduce the set-up error of patients compared to single ExacTrac. However, the application of Combine IGRT increases patient set-up verification time and absorption dose in the body for image acquisition. Therefore, depending on the patient's situation that using Combine IGRT to reduce the patient's set-up error can increase the radiation treatment effectiveness.
목 적 : 본 연구에서는 정위적체부방사선치료시 ExacTrac과 CBCT를 단계적으로 적용한 Combine IGRT를 사용하여 Set-up 오차를 비교 및 분석하여 Combine IGRT의 유용성을 평가하고자 한다. 대상 및 방법 : 2014년 5월에서 2017년 11월까지 본원에 내원하여 trueBEAM Stx(Varian medical System, USA)에서 정위적체부방사선치료를 받은 환자 중 부위별로 뇌(Brain) 3명, 척추(Spine) 9명, 골반(Pelvis) 3명으로 분류하였다. 치료전 ExacTrac을 사용하여 Lateral(Lat), Longitudinal(Lng), Vertical(Vrt) 방향과 Roll, Pitch, Yaw 방향으로 Set-up 오차를 보정하였고, ExacTrac 이동 값을 적용 후 CBCT를 추가적으로 수행하여 Lat, Lng, Vrt, Rotation(Rtn) 방향으로 보정하였다. 결 과 : ExacTrac 사용시 뇌 부위의 오차는 Lat $0.18{\pm}0.25cm$, Lng $0.23{\pm}0.04cm$, Vrt $0.30{\pm}0.36cm$, Roll $0.36{\pm}0.21^{\circ}$, Pitch $1.72{\pm}0.62^{\circ}$, Yaw $1.80{\pm}1.21^{\circ}$, 척추 부위는 Lat $0.21{\pm}0.24cm$, Lng $0.27{\pm}0.36cm$, Vrt $0.26{\pm}0.42cm$, Roll $1.01{\pm}1.17^{\circ}$, Pitch $0.66{\pm}0.45^{\circ}$, Yaw $0.71{\pm}0.58^{\circ}$, 골반 부위는 Lat $0.20{\pm}0.16cm$, Lng $0.24{\pm}0.29cm$, Vrt $0.28{\pm}0.29cm$, Roll $0.83{\pm}0.21^{\circ}$, Pitch $0.57{\pm}0.45^{\circ}$, Yaw $0.52{\pm}0.27^{\circ}$로 나타났다. Couch 이동 후 CBCT를 하였을 때, 뇌 부위는 Lat $0.06{\pm}0.05cm$, Lng $0.07{\pm}0.06cm$, Vrt $0.00{\pm}0.00cm$, Rtn $0.0{\pm}0.0^{\circ}$, 척추 부위는 Lat $0.06{\pm}0.04cm$, Lng $0.16{\pm}0.30cm$, Vrt $0.08{\pm}0.08cm$, Rtn $0.00{\pm}0.00^{\circ}$, 골반 부위는 Lat $0.06{\pm}0.07cm$, Lng $0.04{\pm}0.05cm$, Vrt $0.06{\pm}0.04cm$, Rtn $0.0{\pm}0.0^{\circ}$ 오차가 발생하였다. 결 론 : 정위적체부방사선치료시 ExacTrac에 추가적으로 CBCT를 사용한 Combine IGRT의 경우는 ExacTrac만 사용한 경우보다 환자의 Set-up 오차를 줄일 수 있게 나타났다. 그러나 Combine IGRT를 적용함으로 환자 Set-up 확인시간, 영상획득을 위한 체내 흡수선량이 증가한다. 그러므로 환자 상황에 따라 Combine IGRT를 사용하여 환자의 Set-up 오차를 적게 함으로써 방사선치료 효과비를 증가시킬 수 있을 것으로 사료된다.