DOI QR코드

DOI QR Code

Study on Characteristics of ECG Electrodes for Motion Artifact Reduction

동잡음 저감을 위한 심전도 전극 특성에 대한 연구

  • Kang, Young-Hwan (Department of Electronics and Control Engineering, Hanbat National University) ;
  • Park, Jae-Soon (Department of Electronics and Control Engineering, Hanbat National University) ;
  • Cho, Bum-Ki (Department of Electronics and Control Engineering, Hanbat National University) ;
  • Choi, Sang-Dong (M-tech Co. Ltd.) ;
  • Joung, Yeun-Ho (Department of Electronics and Control Engineering, Hanbat National University)
  • Received : 2017.03.22
  • Accepted : 2017.04.13
  • Published : 2017.06.01

Abstract

In this paper, we introduce an electrocardiogram (ECG) system designed to solve problems caused by wetgels and motion artifacts in measuring active movement. The system is called a dry-contact ECG and was designed by considering impedance matching between skin and electrode as well as the frictional electricity between electrode and clothes. In order to create the system, we measured impedance on the skin-electrode interface, and the result was applied to the electronic circuit scheme. Moreover, we added an electrode on the back of the measurement electrode to make a flow path to ground the electrical noise. The final ECG circuit and novel electrode were used to detect real human cardiac signals from a subject who was tested while standing still and walking. The signals obtained from the two activities were nicely shaped, without any motion artifact noise. We took electrode size into account in this study because the impedance depended on the area of the electrode. An electrode of 50 mm diameter showed the best curve for the ECG signal without any electrical noise.

Keywords

References

  1. H. J. Park, Ph. D. Implementation of Wireless Portable ECG Monitoring System, p. 4, Chosun University, Gwangju (2010).
  2. D. C. Lee, S. K. Choi, G. J. Lee, Y. B. Jeon, and W. S. Yu, Korean Circ. J., 21, 646 (1991). [DOI: http://dx.doi.org/10.4070/kcj.1991.21.3.646]
  3. D. N. Mathias, S. I. Kim, J. S. Park, and Y. H. Joung, Trans. Elec. Electron. Mater., 16, 16 (2015). [DOI: https://doi.org/10.4313/TEEM.2015.16.1.16]
  4. J. Pan and W. J. Tompkins, IEEE Transactions on Biomedical Engineering, 3, 230 (1985). [DOI: http://dx.doi.org/10.1109/TBME.1985.325532]
  5. S. Y. Kang, K. S. Bang, J. P. Choi, J. K. Kim, and W. Heo, Journal of Myongji Institute of Industrial Technology, 22, 111 (2003).
  6. A. A. Uvarov, I. A. Lezhnina, K. V. Overchuk, A. S. Starchak, S. D. Akhmedov, and I. A. Larioshina, J. Phys. Conf. Ser., 671, 1 (2016). [DOI: https://doi.org/10.1088/1742-6596/671/1/012032]
  7. D. N. Mathias, S. I. Kim, J. S. Park, Y. H. Joung, and W. S. Choi, Trans. Elec. Electron. Mater., 16, 42 (2015). [DOI: https://doi.org/10.4313/TEEM.2015.16.1.42]