Abstract
The dosimetric characteristics were experimentally evaluated for electron beams from the prototype linac developed for radiotherapy units. This paper focuses on the electron beam output and energy variations as a function of electron gun heater current. The electron energy was derived from its mean and most probable energies measured by film dosimetry. The electron beam output at the maximum electron energy was measured with the plane parallel ionization chamber in water using TRS-398 dosimetry protocol. The mean energy and the most probable energy of the electron beam were 6.54~3.31 MeV and 5.94~2.80 MeV at electron gun current of 2.02~2.50 A respectively. The output dose rate for an electron beam of mean energy 6.54 MeV was 5.41 Gy/min ${\pm}1.5%$ at the reference depth in water.
본 연구에서는 의료용 선형가속기 제작을 위해 개발된 연구용 선형가속기의 전자선에 대한 선량학적 특성을 실험적으로 평가하였다. 본 논문에서는 전자총 가열 전류에 따른 에너지의 변화와 출력 흡수선량 측정 결과를 보고하고자 한다. 전자선의 에너지는 필름 측정법을 써서 평균에너지와 최빈에너지의 관점에서 결정하였다. 출력 흡수선량은 최적 에너지에 대하여 평행평판형 전리함을 사용하여 물속 깊이선량율을 측정하고 TRS-398 프로토콜에 따라 결정하였다. 측정 결과 전자총 가열 전류 2.02~2.50 A에서 평균에너지와 최빈에너지는 5.94~2.80 MeV와 6.54~3.31 MeV로 변화하였다. 그리고 평균에너지 5.94 MeV의 전자선에 대해 물속 기준 깊이에서 출력 흡수선량은 5.41 Gy/min으로 나타났다.