Abstract
For the mobile healthcare environment, it is important to measure the exact biomedical signals in real time, and another key point is to design mobile healthcare devices with low power consumption. In this paper, we propose a method in which the piezo film sensor(PFS), having a low power characteristic, is used to measure the pulse signal synchronized with the heart rate from the radial artery. The critical issue in the bio-signal processing is the existence of the motion artifacts. To dissolve this problem, we have applied the periodic moving average filter using the quasi-periodicity of the pulse signal in addition to the conventional method of the adaptive filtering using the reference signal. Results of simulation and experiments are presented to confirm that the quasi-periodicity of the PFS signal can be used to eliminate completely the motion artifacts which still appears after the adaptive filtering.
모바일 헬스케어 환경에서는 정확한 생체 정보를 실시간으로 측정하는 것과 함께, 저전력의 장치를 설계하는 것이 무엇보다 중요한 관건이다. 본 논문에서는 저전력 특성을 갖는 피에조 필름 센서(PFS)를 이용하여 손목의 요골 동맥으로부터 심장의 동기 신호인 맥파를 측정하는 방법을 제안하였다. 생체 신호처리에 있어서 문제점은 동잡음의 존재이다. 본 연구에서는 맥파 신호의 동잡음을 제거하기 위해서 일반적으로 적용되는 참조신호를 이용한 적응 필터링 외에 펄스 신호의 유사 주기성을 이용한 주기 이동 평균 필터 방법을 추가로 적용하였다. 시물레이션 및 실험 결과 PFS 신호에 유사 주기성을 이용함으로써 적응 필터 이후에도 존재하는 동잡음을 완벽히 제거할 수 있음을 확인 하였다.