DOI QR코드

DOI QR Code

Development of a Measurement System of the Transferred Pressure from Intermittent Pneumatic Compression Device

간헐적공기압박장치의 전달압력 측정시스템 개발

  • 이원희 (국립암센터 융합기술연구부 의공학연구과) ;
  • 서종현 (국립암센터 융합기술연구부 의공학연구과) ;
  • 김준 (국립암센터 융합기술연구부 의공학연구과) ;
  • 강승호 ((주)대성마리프 의공학연구소) ;
  • 김국한 ((주)대성마리프 의공학연구소) ;
  • 정승현 (국립암센터 재활의학클리닉) ;
  • 김광기 (국립암센터 융합기술연구부 의공학연구과) ;
  • 강현귀 (국립암센터 골연부종양클리닉)
  • Received : 2015.12.18
  • Accepted : 2016.02.21
  • Published : 2016.02.29

Abstract

A pressure measurement system was developed to verify magnitude and position of transferred pressure on the body surface during the intermittent pneumatic compression (IPC) which is one of the most well-known methods for the prevention of deep vein thrombosis (DVT). Eighty force sensing resistors (FSR) were arranged on a mannequin leg and a hardware controller sensed, digitized, and transferred pressure data every second while IPC was being applied. Finally, sensed pressure data were color coded and visualized on the 3D model with lab-developed software. The pressure data were also saved to files for further analysis. Using this measurement system, the changing pattern of pressure was measured on the mannequin leg by changing both chamber pressure and cuff tightness. As a result, net pressure transferred onto the body surface is dependent on chamber pressure and cuff tightness. Under the same chamber pressure, the tighter a cuff was worn, the wider compressed area was and the shorter compression cycle was. Also transferred pressure was proportional to both chamber pressure and cuff tightness.

Keywords

References

  1. D. McGrath, W.G. Dennyson, and M. Rolland, "Death rate from pulmonary embolism following joint replacement surgery", J R Coll Surg Edinb, vol. 41, no. 4, pp. 265-266, 1996.
  2. T.P. Pierce, J.J. Cherian, J.J. Jauregui, R.K. Elmallah, J.R. Lieberman, and M.A. Mont, "A Current Review of Mechanical Compression and Its Role in Venous Thromboembolic Prophylaxis in Total Knee and Total Hip Arthroplasty", J Arthroplasty, vol. 30, no. 12, pp. 2279-2284, 2015. https://doi.org/10.1016/j.arth.2015.05.045
  3. P. Roderick, G. Ferris, K. Wilson, H. Halls, D. Jackson, R. Collins, and C. Baigent, "Towards evidence-based guidelines for the prevention of venous thromboembolism: systematic reviews of mechanical methods, oral anticoagulation, dextran and regional anaesthesia as thromboprophylaxis", Health Technol Assess, vol. 9, no. 49, pp. iii-iv, ix-x, 1-78, 2005.
  4. S.K. Kakkos, G. Szendro, M. Griffin, S.S. Daskalopoulou, and A.N. Nicolaides, "The efficacy of the new SCD response compression system in the prevention of venous stasis", J Vasc Surg, vol. 32, no. 5, pp. 932-940, 2000. https://doi.org/10.1067/mva.2000.110358
  5. G.H. Guyatt, E.A. Akl, M. Crowther, D.D. Gutterman, H.J. Schuunemann, for the American College of Chest Physicians Antithrombotic Therapy and Prevention of Thrombosis Panel, "Executive summary: Antithrombotic Therapy and Prevention of Thrombosis, 9th ed: American College of Chest Physicians Evidence-Based Clinical Practice Guidelines", Chest, vol. 141, pp. 7S-47S, 2012. https://doi.org/10.1378/chest.1412S3
  6. E. Kesieme, C. Kesieme, N. Jebbin, E. Irekpita, and A. Dongo, "Deep vein thrombosis: a clinical review", J Blood Med, vol. 2, pp. 59-69, 2011.
  7. J.C. Giddings, R.J. Morris, H.M. Ralis, G.M. Jennings, D.A. Davies, and J.P. Woodcock, "Systemic haemostasis after intermittent pneumatic compression. Clues for the investigation of DVT prophylaxis and travellers thrombosis", Clin Lab Haematol, vol. 26, no. 4, pp. 269-273, 2004. https://doi.org/10.1111/j.1365-2257.2004.00614.x
  8. A.H. Chen, S.G. Frangos, S. Kilaru, and B.E. Sumpio, "Intermittent pneumatic compression devices - physiological mechanisms of action", Eur J Vasc Endovasc Surg, vol. 21, no. 5, pp. 383-392, 2001. https://doi.org/10.1053/ejvs.2001.1348
  9. A.J. Comerota, V. Chouhan, R.N. Harada, L. Sun, J. Hosking, R. Veermansunemi, A.J. Comerota Jr, D. Schlappy, and A.K. Rao, "The fibrinolytic effects of intermittent pneumatic compression: mechanism of enhanced fibrinolysis", Ann Surg, vol. 226, no. 3, pp. 306-314, 1997. https://doi.org/10.1097/00000658-199709000-00010
  10. W.H. Geerts, D. Bergqvist, G.F. Pineo, J.A. Heit, C.M. Samama, M.R. Lassen, C.W. Colwell, and American College of Chest Physicians, "Prevention of venous thromboembolism: American College of Chest Physicians Evidence-Based Clinical Practice Guidelines (8th Edition)", Chest, vol. 133, pp. 381S-453S, 2008. https://doi.org/10.1378/chest.08-0656
  11. J.M. Zhao, M.L. He, Z.M. Xiao, T.S. Li, H. Wu, and H. Jiang, "Different types of intermittent pneumatic compression devices for preventing venous thromboembolism in patients after total hip replacement", Cochrane Database Syst Rev, vol. 12, p. CD009543, 2014.
  12. D. Kim, H. Chung, C.H. Yi, Y.S. Yoon, J. Son, Y. Kim, M.G. On, and J. Yang, "Effect of glenohumeral position on contact pressure between the capsulolabral complex and the glenoid in free ALPSA and Bankart lesions", Knee Surg Sports Traumatol Arthrosc, vol. 24, no. 2, pp. 350-356, 2016. https://doi.org/10.1007/s00167-015-3962-4
  13. W.G. Yoo, "Effect of a suspension seat support chair on the trunk flexion angle and gluteal pressure during computer work", J Phys Ther Sci, vol. 27, no. 9, pp. 2989-2990, 2015. https://doi.org/10.1589/jpts.27.2989
  14. J.S. Park, M.H. McGarry, S.T. Campbell, H.J. Seo, Y.S. Lee, S.H. Kim, T.Q. Lee, and J.H. Oh, "The optimum tension for bridging sutures in transosseous-equivalent rotator cuff repair: a cadaveric biomechanical study", Am J Sports Med, vol. 43, no. 9, pp. 2118-2125, 2015. https://doi.org/10.1177/0363546515590596
  15. S.G. Kim and J.H. Park, "The effects of dual-task gait training on foot pressure in elderly women", J Phys Ther Sci, vol. 27, no. 1, pp. 143-144, 2015. https://doi.org/10.1589/jpts.27.143
  16. W.G. Yoo, "Effects of different computer typing speeds on acceleration and peak contact pressure of the fingertips during computer typing", J Phys Ther Sci, vol. 27, no. 1, pp. 57-58, 2015. https://doi.org/10.1589/jpts.27.57
  17. M.H. Kim and W.G. Yoo, "Comparison of Center of Force Trajectory during Sit-to-stand Movements Performed by Elderly and Old-old Elderly Subjects", J Phys Ther Sci, vol. 26, no. 9, pp. 1403-1404, 2014. https://doi.org/10.1589/jpts.26.1403
  18. M. Gibson, "A practical demonstration of Boyle's Law", Occup Med (Lond), vol. 65, no. 2, p. 109, 2015. https://doi.org/10.1093/occmed/kqu101
  19. J.B. West, "The original presentation of Boyle's law", J Appl Physiol (1985), vol. 87, no. 4, pp. 1543-1545, 1999. https://doi.org/10.1152/jappl.1999.87.4.1543
  20. R.R. Demers, M. Sullivan, and M. Saklad, "Boyle's law reconsidered: a graphical analysis", Respir Care, vol. 21, no. 7, pp. 610-616, 1976.
  21. J.H. Seo, W. Lee, S.H. Kang, G.H. Kim, Y.H. Jung, M.G. Lim, H.G. Kang, K.G. Kim and S.H. Chung, "Development of Quantitative Force Detecting System on IPC Using FSR sensor", The Magazine of the IEE, vol. 2015, no. 6, pp. 1093-1095, 2015.
  22. Interlink-electronics, "FSR 402 Data Sheet (FSR 400 Series Round Force Sensing Resistor)", Interlink-electronics, 1994, pp. 1-3.