Abstract

An electrical signals of a conventional radiation medical imaging sensor are obtained by charge integration method. In this study, the polycrystalline cadmium telluride(p-CdTe) film was fabricated by a thermal evaporation method for the photon counting sensor development with excellent resolution in low exposure dose. From the fabricated p-CdTe sensor, the physical properties(SEM, XRD) and the electrical properties(leakage current, x-ray sensitivity, SNR) were evaluated. As a result, the leakage current of below $5nA/cm^2$ and $7{\mu}C/cm^2-R$ of the X-ray sensitivity were showed in below $1V/{\mu}m$. In addition, the signal to noise ratio showed the values of above 5000 at operating voltage.

종래의 방사선 의료영상 센서는 방사선에 의해 생성된 전하신호를 화소내에서 적분하는 방식이다. 본 연구에서는 저선량에서 우수한 해상력이 가능한 광계수형 영상센서를 개발하기 위해 진공 열증착법을 이용하여 다결정 CdTe(p-CdTe) 필름을 제작하였다. 또한, 광계수형 센서의 성능평가를 위하여 물리적 특성(SEM, XRD) 및 전기적 특성(leakage current, x-ray sensitivity, SNR) 평가를 하였다. 측정 결과, $1V/{\mu}m$ 이하의 인가전압에서 $5nA/cm^2$ 이하의 누설 전류밀도를 보였으며, X선 발생신호량은 $7{\mu}C/cm^2{\cdot}R$으로 광계수형 센서로의 적용에 적합한 것으로 평가되었다. 또한, 신호대잡음비는 동작영역에서 5000 이상의 값을 보였다.

Keywords

References

  1. Kyung-O KIM, "Measurement of the electrical properties of a polycrystalline cadmium telluride for direct conversion flat panel x-ray detector", 2014 JINST, pp. 1-4, 2014
  2. Kyung-O KIM, et al, "Radiation detector material development with milti-layer by hetero-junction for the reduction of leakage current", The Korean Society of Radiology, Vol. 3, No. 1, pp 11-15, 2009.
  3. Ji-Koon Park, et al, "X-ray sensitivity of hybrid-type sensor based on CaWO4-selenium for digital X-ray imager", Transactions on Electrical and ELectronic Materials, Vol. 5, No. 4, pp 133-134, 2004. https://doi.org/10.4313/TEEM.2004.5.4.133
  4. Ji-Koon Park, et al, "Performance evaluation of a selcenium based prototype digital radiation detector", J. Biomed. Eng, Res, pp. 300-305, 2007.
  5. Heo-Ye-Ji, et al, "A study of the photon counting sensor setup for the evaluation of the photon count efficiency, The Korean Society of Radiology Proceeding of 2012 Autumn conference. pp. 177-180, 2012.
  6. Fred P. Vaccaro, "Limitations of the Hecht equation encountered in measuring $\mu\tau$ products in Mercuric Iodine", Vol. 50, pp. 1-5, 2003.
  7. Safa Kasap, et al, "Amorphous and polycrystalline photoconductors for direct coversion flat panel x-ray image sensors", Sensor, Vol. 11, pp. 5112-5157, 2011. https://doi.org/10.3390/s110505112