Electrical Compound Action Potential: Effects of different parameters
전기 자극 청신경 복합활동전위: 변인 영향
Abstract
Electrical compound action potential (ECAP) can be recorded on cochlear implant. This study will investigate stimulation and recording to enhance the efficacy of ECAP. 34 articles was used. We analyzed pulse and stimulating condition, artifact suppression, recording condition. The cathod-leading biphasic pulse was used with as short as possible pulse width and inter phase gap for the efficacy of neural firing, stable threshold and preventing neural degeneration. Around C-level was stimulated to apical, middle and basal turn of cochlea. Artifact was eliminated by forward-masking, template-subtraction technique. For clearer waveform, we need to change distance between stimulating and recording electrode, the gain of amplification, number of average.
전기 자극 청신경 복합활동전위(electrical compound action potential; ECAP)는 인공와우로 청신경을 자극하고 동시에 반응을 기록한다. 이 연구는 인공와우에서 ECAP의 효용성을 높이기 위하여 자극과 기록에 대하여 연구하고자 한다. 대상은 문헌 또는 사례 연구를 제외한 연구 논문 34 편으로 하였다. 연구는 펄스 및 자극 조건, 잡파 억제 및 기록 조건 등을 분석하였다. 효과적인 신경 발화, 안정적인 역치 확인, 신경 퇴행 방지 등을 위해서는 펄스폭과 위상 간격이 가급적 짧고, (-)위상이 앞선 양위상 펄스를 사용하였다. 자극은 펄스 간격, 자극비율, 자극 방식 등을 정하여 와우의 첨부, 중간, 기저회전으로 C-level 정도로 하였다. 잡파는 전방차폐법, 거푸집소거법으로 제거하였고, 분명한 파형을 획득하기 위해서는 전극 간격, 증폭 정도, 평균 가산 회수 등의 추가 조절이 필요하다.