DOI QR코드

DOI QR Code

Effect of a Bonding Layer between Electrodes on the Performance of a λ/4-Mode PVDF Ultrasound Transducer

λ/4 모드 PVDF 초음파 트랜스듀서에 있어서 전극 사이의 접합층이 성능에 미치는 영향

  • ;
  • 하강렬 (부경대학교 물리학과) ;
  • 김무준 (부경대학교 물리학과) ;
  • 김정순 (동명대학교 멀티미디어공학과)
  • Received : 2013.09.27
  • Accepted : 2014.01.04
  • Published : 2014.03.31

Abstract

The effect of a bonding layer on the performance of a quarter-wave (${\lambda}/4$) mode PVDF ultrasound transducer having not only a piezoelectric layer but also a non-piezoelectric layer between two electrodes was analyzed. The equivalent circuit of a transmission line model by Kikuchi et al.[Sound of IEICE, 55-A, 331-338 (1981)] was introduced for the analysis. The validity of the model was confirmed by comparison with a KLM model for three postulated adhesion cases of a $80{\mu}m$ thick piezoelectric PVDF film to a copper (Cu) backer. The pulse-echo responses of five PVDF transducers, each fabricated with a different thickness ($5{\mu}m{\sim}20{\mu}m$) of the bonding layer, were measured and the results were compared with those by simulation. The two results were in good agreement with each other and it was noted that the effect of the bonding layer on the performance of the transducer could be analyzed by the Kikuchi model. In detail, the $20{\mu}m$ bonding layer decreased the center frequency and the bandwidth by about 19.7 % and 25.0 %, respectively, and increased the insertion loss by 57.2 %.

양 전극 사이에 압전층 외에 비압전성의 접합층이 존재하는 ${\lambda}/4$ 모드 PVDF 초음파 트랜스듀서에 있어서 그 접합층이 트랜스듀서의 성능에 미치는 영향을 등가회로에 의해 해석하였다. 등가회로로서는 Kikuchi 등이 제안한 전송선로 모델[Sound of IEICE, 55-A, 331-338 (1981)]을 도입하였는데, 먼저 그 모델에 의한 해석의 타당성을 $80{\mu}m$ 두께의 PVDF 압전막이 동(Cu) 후면체에 부착되는 세 가지 경우를 가정한 KLM 모델과의 비교를 통해 검증하였다. 다음으로, 그 압전막과 더불어 $5{\mu}m{\sim}20{\mu}m$ 두께의 에폭시 접합층을 갖는 다섯 개의 트랜스듀서를 제작하여 펄스 에코 응답을 측정한 후 시뮬레이션 결과와 비교하였다. 두 결과는 서로 잘 일치하였는바, 도입한 Kikuchi 모델에 의해 접합층이 트랜스듀서의 성능에 미치는 영향을 파악할 수 있음을 알았는데, 접합층이 $20{\mu}m$일 때는 그 접합층이 없을 때에 비해 중심주파수와 대역폭은 각각 약 19.7 %, 25.0 % 감소하고, 삽입손실은 57.2 % 증가하는 것으로 나타났다.

Keywords

References

  1. P. Lum, M. Greenstein, C. Grossman, and T. L. Szabo, "High-frequency membrane hydrophone," IEEE Transactions on UFFC, 43, 536-544 (1996). https://doi.org/10.1109/58.503713
  2. F. S. Foster, C. J. Pavlin, K. A. Harasiewicz, D. A. Christopher, and D. H. Turnbull, "Advanced in ultrasound biomicroscopy," Ultrasound in Med. & Bilo. 26, 1-27 (2000). https://doi.org/10.1016/S0301-5629(99)00096-4
  3. M. Robert. G. Molingou, K. Snook, J. Cannata, and K. K. Shung, "Fabrication of focused poly (vinylidene fluoridetrifluoroethylene) P(VDF-TrFE) copolymer 40-50 MHz ultrasound transducers on curved surfaces," J. Appl. Phys. 96, 252-256 (2004). https://doi.org/10.1063/1.1760233
  4. J. A. Ketterling, F. L. Lizzi, O. Aristizabal, and D. H. Turbull. "Design and fabrication of a 40-MHz annular array transducer," IEEE Trans. on UFFC, 52, 672-681 (2005). https://doi.org/10.1109/TUFFC.2005.1428050
  5. H. Ohigashi and K. Koga, "Ferroelectric copolymers of vinylidenefluoride and trifluoroethylene with a large electromechanical coupling factor," Jpn. J. Appl. Phys. 21, L455-L457 (1982). https://doi.org/10.1143/JJAP.21.L455
  6. J. S. Jeong and K. K. Shung, "Improved fabrication of focused single element P (VDF-TrFE) transducer," Ultrasonics, 53, 455-458 (2013). https://doi.org/10.1016/j.ultras.2012.08.013
  7. C. H. Chung and Y. C. Lee, "Broadband poly(vinylidene fluoride-trifluoroethylene) ultrasound focusing transducers for determining elastic constants of coating materials," J. Nondestruct. Eval. 28, 101-110 (2009). https://doi.org/10.1007/s10921-009-0052-x
  8. L. F. Brown, "Design considerations for piezoelectric polymer ultrasound transducers," IEEE Trans. on UFFC, 47, 1377-1396 (2000). https://doi.org/10.1109/58.883527
  9. R. Krimholtz, D. A. Leedom, and G. L. Matthaei, "New equivalent circuit for elementary piezoelectric transducers," Electron. Lett. 6, 398-399 (1970). https://doi.org/10.1049/el:19700280
  10. Y. Kikuchi, N. Chiubachi, and S. Yamamizu, "Analusis of Ultrasonic Multi-layer Transducer for UHF and VHF Ranges"(in Japanese), Sound of IEICE, 55-A, 331-338 (1981)
  11. L. E. Kinsler, A. R. Frey, A. B. Coppens, and J. V. Sanders, Foundamentals of Acoustics (John Wiley & Sons, 2000), 390-406.
  12. J. H. Yoon, J. H. Oh, J. S. Kim, M. J. Kim, and K. L. Ha, "Fabrication and eivaluation of a VHF focusing ultrasonic transdvcer male of PVDF piezoelectric film" (in Korean), J. Acoust. Soc. Kr. 30, 215-222 (2011). https://doi.org/10.7776/ASK.2011.30.4.215

Cited by

  1. Highly Sensitive Impact Sensor Based on PVDF-TrFE/Nano-ZnO Composite Thin Film vol.19, pp.4, 2019, https://doi.org/10.3390/s19040830