Comparative analysis of accuracy of how to use the implant torque controllers

사용 방법에 따른 임플란트 토크 조절기의 정확성 비교 분석 연구

  • Cho, Lee-Ra (Department of Prosthodontics, College of Dentistry, Gangneung-Wonju National University) ;
  • Park, Chan-Jin (Department of Prosthodontics, College of Dentistry, Gangneung-Wonju National University) ;
  • Huh, Yoon-Hyuk (Department of Prosthodontics, Sanbon Dental Hospital, Wonkwang University) ;
  • Lee, Ji-Hye (Department of Dentistry, Uijeongbu St. Mary's Hospital, Catholic University of Korea) ;
  • Kim, Dae-Gon (Department of Prosthodontics, College of Dentistry, Gangneung-Wonju National University)
  • 조리라 (강릉원주대학교 치과대학 치과보철학교실) ;
  • 박찬진 (강릉원주대학교 치과대학 치과보철학교실) ;
  • 허윤혁 (원광대학교 산본치과병원 치과보철과) ;
  • 이지혜 (카톨릭대학교 의정부성모병원 치과학교실) ;
  • 김대곤 (강릉원주대학교 치과대학 치과보철학교실)
  • Received : 2013.09.13
  • Accepted : 2013.10.04
  • Published : 2013.10.31

Abstract

Purpose: This study was to evaluate the accuracy of how to use the Mechanical Torque-Limiting Devices. Materials and methods: Three different implant manufacturers (Dentium, Osstem, Shinhung) were prepared for each two implant torque controllers. Divided into two groups depending on the method used for removal torque was measured. Repeated measures of ANOVA test (${\alpha}=.05$) was used as statistics to evaluate the effect of repeated loading number on the removal torque. Independent t-test was used to evaluate the difference in removal torque of two groups. Results: The removal torque significantly decreased as the number of loading repetition increased (P<.05). There was significant difference between two groups. Conclusion: Loosen the handle part of the implant torque controllers spring resilience to recover one group compared to the group that did not, showed a little more closed to the reference value.

연구 목적: 임상에서 가장 보편적으로 사용되는 기계식 토크 조절기를 선택하여 토크 조절기를 사용하는 방법에 따라서 제조사에서 권장하는 토크를 적용하고 유지하는데 정확도의 차이가 있는지를 비교, 분석하고자 하였다. 연구 재료 및 방법: 36개월을 사용한 3개 제조회사 제품의 토크 최대값 제한 장치 형태의 기계식 토크 조절기를 대상으로 하였다. 토크 조절기를 사용할 때마다 핸들 부위를 조이고 사용 후에는 핸들 부위를 풀어서 토크 조절기 내부의 스프링이 압축력에 의해 변형되는 것을 막고자 한 그룹과 토크 조절기를 사용 후에도 핸들 부위를 풀어내지 않고 토크 조절기 내부의 스프링이 압축력 하에 유지되게 사용한 그룹으로 나누어, 총 6개의 토크 조절기를 이용하여 풀림 토크를 측정하였다. 반복 측정에 따른 풀림 토크값의 차이를 비교하기 위해 repeated measures of ANOVA test (${\alpha}=.05$)를 사용하였고, 사용 방법의 차이에 따른 두 그룹간의 비교를 위해 독립표본 t-검정(Independent sample t-test)을 시행하였다. 결과:모든 실험군에서 반복 측정에 따라 풀림 토크값이 유의성 있게 감소하는 것으로 나타났고(P<.05), 사용 방법에 따른 두 그룹 간에 풀림 토크값은 유의한 차이가 있었다. 결론: 토크 조절기를 사용한 뒤 다시 핸들부분을 풀어서 토크 조절기 내부의 스프링이 압축력에 의해 변형되는 것을 막고 스프링의 복원력을 회복한 그룹은 그렇게 사용하지 않은 그룹과 통계적으로 유의한 차이가 관찰됐고, 기준 토크값(30 Ncm)에 좀 더 가까운 풀림 토크값을 보였다.

Keywords

References

  1. Jaarda MJ, Razzoog ME, Gratton DG. Ultimate tensile strength of five interchangeable prosthetic retaining screws. Implant Dent 1996;5:16-9. https://doi.org/10.1097/00008505-199600510-00003
  2. Jaarda MJ, Razzoog ME, Gratton DG. Comparison of "look-alike" implant prosthetic retaining screws. J Prosthodont 1995;4:23-7. https://doi.org/10.1111/j.1532-849X.1995.tb00310.x
  3. Jaarda MJ, Razzoog ME, Gratton DG. Geometric comparison of five interchangeable implant prosthetic retaining screws. J Prosthet Dent 1995;74:373-9. https://doi.org/10.1016/S0022-3913(05)80377-4
  4. Mitrani R, Nicholls JI, Phillips KM, Ma T. Accuracy of electronic implant torque controllers following time in clinical service. Int J Oral Maxillofac Implants 2001;16:394-9.
  5. Standlee JP, Caputo AA, Chwu MY, Sun TT. Accuracy of mechanical torque-limiting devices for implants. Int J Oral Maxillofac Implants 2002;17:220-4.
  6. Vallee MC, Conrad HJ, Basu S, Seong WJ. Accuracy of frictionstyle and spring-style mechanical torque limiting devices for dental implants. J Prosthet Dent 2008;100:86-92. https://doi.org/10.1016/S0022-3913(08)60154-7
  7. Tan KB, Nicholls JI. The effect of 3 torque delivery systems on gold screw preload at the gold cylinder-abutment screw joint. Int J Oral Maxillofac Implants 2002;17:175-83.
  8. Dellinges MA, Tebrock OC. A measurement of torque values obtained with hand-held drivers in a simulated clinical setting. J Prosthodont 1993;2:212-4. https://doi.org/10.1111/j.1532-849X.1993.tb00411.x
  9. Gross M, Kozak D, Laufer BZ, Weiss EI. Manual closing torque in five implant abutment systems: an in vitro comparative study. J Prosthet Dent 1999;81:574-8. https://doi.org/10.1016/S0022-3913(99)70212-X
  10. Dellinges M, Curtis D. Effects of infection control procedures on the accuracy of a new mechanical torque wrench system for implant restorations. J Prosthet Dent 1996;75:93-8. https://doi.org/10.1016/S0022-3913(96)90424-2
  11. Saboury A, Sadr SJ, Fayaz A, Mahshid M. The effect of aging on the accuracy of new friction-style mechanical torque limiting devices for dental implants. J Dent 2013;10:41-50.
  12. McCracken MS, Mitchell L, Hegde R, Mavalli MD. Variability of mechanical torque-limiting devices in clinical service at a US dental school. J Prosthodont 2010;19:20-4. https://doi.org/10.1111/j.1532-849X.2009.00524.x
  13. Standlee JP, Caputo AA. Accuracy of an electric torque-limiting device for implants. Int J Oral Maxillofac Implants 1999;14:278-81.
  14. Gutierrez J, Nicholls JI, Libman WJ, Butson TJ. Accuracy of the implant torque wrench following time in clinical service. Int J Prosthodont 1997;10:562-7.
  15. Binon PP. Evaluation of the effectiveness of a technique to prevent screw loosening. J Prosthet Dent 1998;79:430-2. https://doi.org/10.1016/S0022-3913(98)70157-X
  16. Siamos G, Winkler S, Boberick KG. Relationship between implant preload and screw loosening on implant-supported prostheses. J Oral Implantol 2002;28:67-73. https://doi.org/10.1563/1548-1336(2002)028<0067:TRBIPA>2.3.CO;2

Cited by

  1. Accuracy of spring-style and friction-style mechanical implant torque wrench vol.54, pp.1, 2016, https://doi.org/10.4047/jkap.2016.54.1.8