Evaluation of the Spatial Resolution for Exposure Class in Computed Radiography by Using the Modulation Transfer Function

변조전달함수를 이용한 컴퓨터 방사선영상의 감도 노출 분류에 따른 공간분해능 평가

  • 성열훈 (청주대학교 방사선학과)
  • Received : 2013.06.19
  • Accepted : 2013.08.20
  • Published : 2013.08.28

Abstract

The purpose of the study was to present basic data to evaluation of the spatial resolution for exposure class(EC) in computed radiography(CR) by using the modulation transfer function(MTF). In this study, MTF was measured the edge method by using image plate(IP) of $100{\mu}mm$ pixels. A standard beam quality RQA5 based on an international electro-technical commission(IEC) standard was used to perform the X-ray imaging studies. Digital imaging began to set the sensitivity to EC 50, 100, 200, 300, 400, 600, 800, 1200 in X-ray irradiated to IP. The MTF 50% and 10% in the final images was analysis by using an authorized image analysis program the Origin 8.0 and the image J. As a results, the EC 200 was the best spatial resolution at MTF 50% ($1.979{\pm}0.114lp/mm$) and MTF 10% ($3.932{\pm}0.041$). Therefore, the EC 200 could be useful for the diagnosis of diseases that require high spatial resolution such as fractures.

본 연구에서는 변조전달함수(Modulation transfer function, MTF)를 이용한 컴퓨터 방사선영상의 감도 노출 분류에 따른 공간분해능을 평가하여 컴퓨터 방사선영상 획득의 기초자료로 제시하고자 하였다. 본 실험에서는 $100{\mu}mm$ 픽셀의 영상판을 이용하여 엣지법의 MTF를 측정하였다. 방사선 선질은 IEC 61267에서 권고하고 있는 RQA5를 이용하였다. X-선이 조사된 영상판은 감도 노출 분류를 50, 100, 200, 300, 400, 600, 800, 1200으로 각각 설정한 후 디지털 영상화하였다. 최종 획득된 영상들은 공인된 영상분석 프로그램인 image J와 Origin 8.0을 이용하여 MTF 50%와 10%를 구하여 평가하였다. 그 결과 감도 노출 분류 200에서 가장 우수한 MTF 50%($1.979{\pm}0.114lp/mm$)와 MTF 10%($3.932{\pm}0.041$)의 공간주파수를 구하였다. 따라서 골절 등과 같이 높은 공간분해능을 요구하는 질병진단에 유용할 것으로 사료된다.

Keywords